Lens thickness assessment : anterior segment optical coherence tomography versus A-scan ultrasonography

被引:0
|
作者
Nikoo Hamzeh [1 ]
Sasan Moghimi [1 ,2 ]
Golshan Latifi [1 ]
Massood Mohammadi [1 ]
Nassim Khatibi [1 ]
Shan C.Lin [2 ]
机构
[1] Tehran University of Medical Science, Farabi Eye Hospital
[2] Koret Vision Center, San Francisco Medical School,University of California
关键词
lens thickness; A-scan ultrasonography; anterior segment-optical coherence tomography; biometry; cataract;
D O I
暂无
中图分类号
R776.1 [白内障];
学科分类号
100212 ;
摘要
AIM: To assess lens thickness measurements with anterior segment-optical coherence tomography(AS-OCT)in comparison with A-scan ultrasonography(A-scan US).·METHODS: There were 218 adult subjects(218 eyes)aged 59.2 ±9.2y enrolled in this prospective cross-sectional study. Forty-three eyes had open angles and 175 eyes had narrow angles. Routine ophthalmic exam was performed and nuclear opacity was graded using the Lens Opacities Classification System Ⅲ(LOCS Ⅲ). Lens thickness was measured by AS-OCT(Visante OCT, Carl Zeiss Meditec, Dublin, CA, USA). The highest quality image was selected for each eye and lens thickness was calculated using Image J software. Lens thickness was also measured by A-scan US.·RESULTS: Interclass correlations showed a value of99.7% for intra-visit measurements and 95.3% for inter-visit measurements. The mean lens thickness measured by AS-OCT was not significantly different from that of A-scan US(4.861 ±0.404 vs 4.866 ±0.351 mm, P =0.74).Lens thickness values obtained from the two instruments were highly correlated overall(Pearson correlation coefficient =0.81,P <0.001), and in all LOCS Ⅲ specific subgroups except in grade 5 of nuclear opacity. Bland-Altman analysis revealed a 95% limit of agreement from-0.45 to 0.46 mm. Lens thickness difference between the two instruments became smaller as the lens thickness increased and AS-OCT yielded smaller values than A-scan US in thicker lens(β=-0.29, P <0.001)· CONCLUSION: AS-OCT-derived lens thickness measurement is valid and comparable to the results obtained by A-scan US. It can be used as a reliable noncontact method for measuring lens thickness in adults with or without significant cataract.
引用
收藏
页码:1151 / 1155
页数:5
相关论文
共 50 条
  • [21] Accuracy of cornea and lens biometry using anterior segment optical coherence tomography
    Dunne, Mark C. M.
    Davies, Leon N.
    Wolffsohn, James S.
    JOURNAL OF BIOMEDICAL OPTICS, 2007, 12 (06)
  • [22] Imaging of crystalline lens using a new anterior segment optical coherence tomography
    Mori, Haruhiro
    Ueno, Yuta
    Terauchi, Akari
    Hiraoka, Takahiro
    Oshika, Tetsuro
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [23] Assessment of conjunctival, episcleral and scleral thickness in healthy individuals using anterior segment optical coherence tomography
    Teeuw, G. J.
    Vergouwen, D. P. C.
    Ramdas, W. D.
    Sanchez-Brea, L.
    Andrade De Jesus, D.
    Rothova, A.
    Vingerling, J. R.
    Ten Berge, J. C.
    ACTA OPHTHALMOLOGICA, 2024, 102 (05) : 573 - 580
  • [24] Assessment of anterior scleral thickness in Turkish open angle glaucoma patients: an anterior segment optical coherence tomography study
    Torun, Isil Merve
    Saridogan, Melike
    BMC OPHTHALMOLOGY, 2025, 25 (01)
  • [25] Measurement of the Lower Lid Margin Thickness by Anterior Segment Optical Coherence Tomography
    Wang, Da-Hu
    Li, Zhao-Yan
    Liu, Xin-Quan
    OPHTHALMIC RESEARCH, 2021, 64 (01) : 22 - 27
  • [26] Anterior segment optical coherence tomography scanning protocols and corneal thickness repeatability
    Fisher, Damien
    Collins, Michael J.
    Vincent, Stephen J.
    CONTACT LENS & ANTERIOR EYE, 2020, 43 (05): : 433 - 440
  • [27] Imaging of Anterior Segment Tumours: A Comparison of Ultrasound Biomicroscopy Versus Anterior Segment Optical Coherence Tomography
    Kottaridou, Eleni
    Hatoum, Adam
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (01)
  • [28] Comparison of Ultrasound Biomicroscopy versus Anterior Segment Optical Coherence Tomography for Imaging of Anterior Segment Tumors
    Bianciotto, C. G.
    Romanelli, M.
    Guzman, M.
    Mazzuca, D.
    Green, R.
    Shields, C. L.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [29] Anterior Segment Tomography with the Cirrus Optical Coherence Tomography
    Rodrigues, Eduardo B.
    Johanson, Margara
    Penha, Fernando M.
    JOURNAL OF OPHTHALMOLOGY, 2012, 2012
  • [30] Agreement between Scheimpflug photography and A-scan ultrasonography in anterior segment ocular measurements in children
    Tong, L
    Wong, EM
    Chan, YH
    Balakrishnan, V
    OPTOMETRY AND VISION SCIENCE, 2003, 80 (07) : 529 - 534