Lens thickness assessment : anterior segment optical coherence tomography versus A-scan ultrasonography

被引:0
|
作者
Nikoo Hamzeh [1 ]
Sasan Moghimi [1 ,2 ]
Golshan Latifi [1 ]
Massood Mohammadi [1 ]
Nassim Khatibi [1 ]
Shan C.Lin [2 ]
机构
[1] Tehran University of Medical Science, Farabi Eye Hospital
[2] Koret Vision Center, San Francisco Medical School,University of California
关键词
lens thickness; A-scan ultrasonography; anterior segment-optical coherence tomography; biometry; cataract;
D O I
暂无
中图分类号
R776.1 [白内障];
学科分类号
100212 ;
摘要
AIM: To assess lens thickness measurements with anterior segment-optical coherence tomography(AS-OCT)in comparison with A-scan ultrasonography(A-scan US).·METHODS: There were 218 adult subjects(218 eyes)aged 59.2 ±9.2y enrolled in this prospective cross-sectional study. Forty-three eyes had open angles and 175 eyes had narrow angles. Routine ophthalmic exam was performed and nuclear opacity was graded using the Lens Opacities Classification System Ⅲ(LOCS Ⅲ). Lens thickness was measured by AS-OCT(Visante OCT, Carl Zeiss Meditec, Dublin, CA, USA). The highest quality image was selected for each eye and lens thickness was calculated using Image J software. Lens thickness was also measured by A-scan US.·RESULTS: Interclass correlations showed a value of99.7% for intra-visit measurements and 95.3% for inter-visit measurements. The mean lens thickness measured by AS-OCT was not significantly different from that of A-scan US(4.861 ±0.404 vs 4.866 ±0.351 mm, P =0.74).Lens thickness values obtained from the two instruments were highly correlated overall(Pearson correlation coefficient =0.81,P <0.001), and in all LOCS Ⅲ specific subgroups except in grade 5 of nuclear opacity. Bland-Altman analysis revealed a 95% limit of agreement from-0.45 to 0.46 mm. Lens thickness difference between the two instruments became smaller as the lens thickness increased and AS-OCT yielded smaller values than A-scan US in thicker lens(β=-0.29, P <0.001)· CONCLUSION: AS-OCT-derived lens thickness measurement is valid and comparable to the results obtained by A-scan US. It can be used as a reliable noncontact method for measuring lens thickness in adults with or without significant cataract.
引用
收藏
页码:1151 / 1155
页数:5
相关论文
共 50 条
  • [11] Evaluation of Intraocular Lens Tilt With Anterior Segment Optical Coherence Tomography
    Kumar, Dhivya Ashok
    Agarwal, Amar
    Prakash, Gaurav
    Jacob, Soosan
    Saravanan, Yoga
    Agarwal, Athiya
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2011, 151 (03) : 406 - 412
  • [12] Anterior Segment Optical Coherence Tomography as a Screening Tool for the Assessment of the Anterior Segment Angle
    Pekmezci, Melike
    Porco, Travis C.
    Lin, Shan C.
    OPHTHALMIC SURGERY LASERS & IMAGING, 2009, 40 (04) : 389 - 398
  • [13] Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment
    Dada, Tanuj
    Sihota, Ramanjit
    Gadia, Ritu
    Aggarwal, Anand
    Mandal, Subrata
    Gupta, Viney
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05): : 837 - 840
  • [14] Measurement of LASIK Flap Thickness With Anterior Segment Optical Coherence Tomography
    Cheng, Arthur C. K.
    Ho, Thomas
    Lau, Silvania
    Wong, Amy L.
    Leung, Chris
    Lam, Dennis S. C.
    JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (09) : 879 - 884
  • [15] Optical coherence tomography of the anterior segment
    Simpson, Trefford
    Optom, Dip
    Fonn, Desmond
    Optom, Dip
    Optom, M.
    OCULAR SURFACE, 2008, 6 (03): : 117 - 127
  • [16] Anterior segment optical coherence tomography
    Ang, Marcus
    Baskaran, Mani
    Werkmeister, Rene M.
    Chua, Jacqueline
    Schmidl, Doreen
    dos Santos, Valentin Aranha
    Garhoefer, Gerhard
    Mehta, Jodhbir S.
    Schmetterer, Leopold
    PROGRESS IN RETINAL AND EYE RESEARCH, 2018, 66 : 132 - 156
  • [17] Anterior Segment Optical Coherence Tomography
    Stockman, Adam
    CORNEA, 2018, 37 : S9 - S10
  • [18] Anterior Segment Optical Coherence Tomography
    Jancevski, Maria
    Foster, C. Stephen
    SEMINARS IN OPHTHALMOLOGY, 2010, 25 (5-6) : 317 - 323
  • [19] Increase in Lens Thickness as Measured by A-Scan Ultrasonography in an Older African American Population
    Knighton, E. A.
    Zoltoski, R. K.
    Roberts, D. K.
    Wilensky, J. T.
    Kuszak, J. R.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [20] Anterior Segment Imaging: Optical Coherence Tomography Versus Ultrasound Biomicroscopy
    Garcia, Julian P. S., Jr.
    Rosen, Richard B.
    OPHTHALMIC SURGERY LASERS & IMAGING, 2008, 39 (06) : 476 - 484