A microcantilever-based viscometer for inline determination of liquid viscosity

被引:0
|
作者
Esmaeili, Sahar [1 ]
Shahrouzi, Javad Rahbar [1 ]
机构
[1] Sahand Univ Technol, Fac Chem Engn, Sahand New Town, Tabriz, Iran
来源
ENGINEERING RESEARCH EXPRESS | 2025年 / 7卷 / 01期
关键词
microcantilever; microfluidic; viscometer; viscosity; viscosity measurement; MICROFLUIDIC VISCOMETER; TECHNOLOGIES; RHEOMETER; RHEOLOGY; SYSTEM;
D O I
10.1088/2631-8695/adabba
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Inline viscosity monitoring is crucial for process control in many industrial plants. In this study, a viscometer was designed and fabricated based on microfluidic technology, and the influence of effective parameters on its performance was investigated. The construction of the viscometer involved the fabrication of two microcantilevers using polydimethylsiloxane. One microcantilever was designed with a circular cross-section featuring a diameter of 300 mu m, while the other was crafted with a triangular cross-section, characterized by a base of 270 mu m and a height of 250 mu m. Both microcantilevers measured 3 mm in height and were meticulously positioned within separate millichannels; each dimensioned at 2 x 5 x 30 mm3. The precision required for the millichannels and additional components was achieved using a mini-CNC milling machine. To evaluate the performance of the viscometer, a series of water-glycerol solutions comprising 50%, 60%, 65%, 70%, 75%, and 80% glycerol were injected into a microchannel. The flow rate was systematically varied between 3 and 18 ml min-1 in increments of 3 ml min-1, utilizing a syringe pump for precise control. The resultant displacement of the microcantilever, attributed to the fluid flow, was meticulously recorded using a digital camera. According to the experimental results, the displacement curves of circular and triangular microcantilevers are linear with viscosity; increasing the viscosity and flow rate of the fluid passing through the millichannel increases the microcantilever displacement rate. The response time of the microcantilever with a circular cross-section is 3 to 9 s, and that of the microcantilever with a triangular cross-section is 2 to 12 s. Also, the results of multiple tests in the same conditions showed that the system has a repeatability error of 4%. Finally, it can be concluded that the proposed device can be used to measure the viscosity of liquids within a specific range.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Estimating gas concentration using a microcantilever-based electronic nose
    Leis, John
    Zhao, Weichang
    Pinnaduwage, Lal A.
    Gehl, Anthony C.
    Allman, Steve L.
    Shepp, Allan
    Mahmud, Ken K.
    DIGITAL SIGNAL PROCESSING, 2010, 20 (04) : 1229 - 1237
  • [32] Microcantilever-based current balance for precise measurement of the photon force
    Pruchnik, Bartosz
    Orlowska, Karolina
    Swiadkowski, Bartosz
    Gacka, Ewelina
    Sierakowski, Andrzej
    Janus, Pawel
    Gotszalk, Teodor
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [33] Feedback-induced phase noise in microcantilever-based oscillators
    Malo, Javier
    Izpura, Jose-Ignacio
    SENSORS AND ACTUATORS A-PHYSICAL, 2009, 155 (01) : 188 - 194
  • [34] Inaccuracy in the detection of molecules using two microcantilever-based methods
    Luo, C
    JOURNAL OF APPLIED MECHANICS-TRANSACTIONS OF THE ASME, 2005, 72 (04): : 617 - 619
  • [35] Modifying Geometry to Enhance the Performance of Microcantilever-Based Acoustic Sensors
    Morshed, Shakib
    Baldwin, Kathryn E.
    Zhou, Bo
    Prorok, Barton C.
    SENSOR LETTERS, 2009, 7 (01) : 38 - 41
  • [36] DENSITY MEASUREMENT PERFORMANCE IN FLOWING LIQUID USING MICROCANTILEVER-BASED RESONATORS UNDER BENDING AND TORSION VIBRATIONS
    Huang, Linya
    Zhao, Libo
    Hu, Yingjie
    Wang, Hongyan
    Lu, Dejiang
    Fang, Xudong
    Luo, Guoxi
    Wang, Chenying
    Yang, Ping
    Jing, Weixuan
    Zhao, Yulong
    Jiang, Zhuangde
    2019 20TH INTERNATIONAL CONFERENCE ON SOLID-STATE SENSORS, ACTUATORS AND MICROSYSTEMS & EUROSENSORS XXXIII (TRANSDUCERS & EUROSENSORS XXXIII), 2019, : 1353 - 1356
  • [37] Design of Microcantilever-Based Biosensor with Digital Feedback Control Circuit
    Kalambe, Jayu P.
    Patrikar, Rajendra M.
    JOURNAL OF SENSORS, 2012, 2012
  • [38] Microcantilever-based current balance for precise measurement of the photon force
    Bartosz Pruchnik
    Karolina Orłowska
    Bartosz Świadkowski
    Ewelina Gacka
    Andrzej Sierakowski
    Paweł Janus
    Teodor Gotszalk
    Scientific Reports, 13
  • [39] Viscosity of starches - Determination in a rotating cylinder viscometer
    Barham, HN
    Wagoner, JA
    Reed, GN
    INDUSTRIAL AND ENGINEERING CHEMISTRY, 1942, 34 : 1490 - 1495
  • [40] Resonance Frequency Change in Microcantilever-based Sensor due to Humidity Variation
    Nuryadi, Ratno
    Djajadi, Arko
    Adiel, Reyhan
    Aprilia, Lia
    Aisah, Nuning
    NANOTECHNOLOGY APPLICATIONS IN ENERGY AND ENVIRONMENT, 2013, 737 : 176 - +