Contactless and continuous monitoring of respiratory rate in a hospital ward: a clinical validation study

被引:0
|
作者
Toften, Stale [1 ]
Kjellstadli, Jonas T. [1 ]
Kvaerness, Jorn [2 ]
Pedersen, Line [3 ,4 ]
Laugsand, Lars E. [4 ,5 ]
Thu, Ole K. F. [6 ,7 ]
机构
[1] Vitalthings, Dept Res & Data Sci, Trondheim, Norway
[2] God Klin Forskning, Oppdal, Norway
[3] St Olavs Univ Hosp, Dept Pain & Complex Disorders, Trondheim, Norway
[4] Norwegian Univ Sci & Technol NTNU, Fac Med & Hlth Sci, Dept Circulat & Med Imaging, Trondheim, Norway
[5] St Olavs Univ Hosp, Emergency Dept, Trondheim, Norway
[6] Vitalthings AS, Trondheim, Norway
[7] St Olavs Univ Hosp, Dept Anesthesia & Intens Care Med, Trondheim, Norway
关键词
contactless monitoring; continuous monitoring; monitoring; respiratory rate; hospital; clinical deterioration; validation; Vitalthings Guardian M10; ALARM FATIGUE; CARDIAC-ARREST; DETERIORATION; SYSTEMS; IMPACT;
D O I
10.3389/fphys.2024.1502413
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Introduction Continuous monitoring of respiratory rate in hospital wards can provide early detection of clinical deterioration, thereby reducing mortality, reducing transfers to intensive care units, and reducing the hospital length of stay. Despite the advantages of continuous monitoring, manually counting every 1-12 h remains the standard of care in most hospital wards. The objective of this study was to validate continuous respiratory rate measurements from a radar-based contactless patient monitor [Vitalthings Guardian M10 (Vitalthings AS, Norway)] in a hospital ward.Methods An observational study (clinicaltrials.gov: NCT06083272) was conducted at the emergency ward of a university hospital. Adult patients were monitored during rest with Vitalthings Guardian M10 in both a stationary and mobile configuration simultaneously with a reference device [Nox T3s (Nox Medical, Alpharetta, GA, United States)]. The agreement was assessed using Bland-Altman 95% limits of agreement. The sensitivity and specificity of clinical alarms were evaluated using a Clarke Error grid modified for continuous monitoring of respiratory rate. Clinical aspects were further evaluated in terms of trend analysis and examination of gaps between valid measurements.Results 32 patients were monitored for a median duration of 42 min [IQR (range) 35-46 (30-59 min)]. The bias was 0.1 and 0.0 breaths min-1 and the 95% limits of agreement ranged from -1.1 to 1.2 and -1.1 to 1.1 breaths min-1 for the stationary and mobile configuration, respectively. The concordances for trends were 96%. No clinical alarms were missed, and no false alarms or technical alarms were generated. No interval without a valid measurement was longer than 5 min.Conclusion Vitalthings Guardian M10 measured respiratory rate accurately and continuously in resting patients in a hospital ward.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Clinical evaluation of a wearable sensor for mobile monitoring of respiratory rate on hospital wards
    Jarvela, Kristiina
    Takala, Panu
    Michard, Frederic
    Vikatmaa, Leena
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (01) : 81 - 86
  • [22] Contactless Monitoring System Versus Gold Standard for Respiratory Rate Monitoring in Emergency Department Patients: Pilot Comparison Study
    Goldfine, Charlotte E.
    Oshim, Md Farhan Tasnim
    Chapman, Brittany P.
    Ganesan, Deepak
    Rahman, Tauhidur
    Carreiro, Stephanie P.
    JMIR FORMATIVE RESEARCH, 2024, 8
  • [23] Clinical feasibility of a contactless multiparameter continuous monitoring technology for neonates in a large public maternity hospital in Nairobi, Kenya
    Ginsburg, Amy Sarah
    Nia, Sahar Zandi
    Chomba, Dorothy
    Dunsmuir, Dustin
    Waiyego, Mary
    Coleman, Jesse
    Ochieng, Roseline
    Liu, Sichen
    Zhou, Guohai
    Ansermino, J. Mark
    Macharia, William M.
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [24] Clinical feasibility of a contactless multiparameter continuous monitoring technology for neonates in a large public maternity hospital in Nairobi, Kenya
    Amy Sarah Ginsburg
    Sahar Zandi Nia
    Dorothy Chomba
    Dustin Dunsmuir
    Mary Waiyego
    Jesse Coleman
    Roseline Ochieng
    Sichen Liu
    Guohai Zhou
    J. Mark Ansermino
    William M. Macharia
    Scientific Reports, 12
  • [25] Contactless Camera-Based Heart Rate and Respiratory Rate Monitoring Using AI on Hardware
    Kolosov, Dimitrios
    Kelefouras, Vasilios
    Kourtessis, Pandelis
    Mporas, Iosif
    SENSORS, 2023, 23 (09)
  • [26] SoC CMOS UWB Pulse Radar Sensor for Contactless Respiratory Rate Monitoring
    Zito, Domenico
    Pepe, Domenico
    Mincica, Martina
    Zito, Fabio
    Tognetti, Alessandro
    Lanata, Antonio
    De Rossi, Danilo
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2011, 5 (06) : 503 - 510
  • [27] Contactless monitoring of heart and respiratory rate in anesthetized pigs using infrared thermography
    Pereira, Carina Barbosa
    Dohmeier, Henriette
    Kunczik, Janosch
    Hochhausen, Nadine
    Tolba, Rene
    Czaplik, Michael
    PLOS ONE, 2019, 14 (11):
  • [28] CLINICAL APPLICATION OF CONTINUOUS MONITORING OF RESPIRATORY FAILURE
    TANTUM, KR
    DRIPPS, RD
    ANNUAL REVIEW OF MEDICINE, 1971, 22 : 293 - &
  • [29] Validation of a Wearable System for Respiratory Rate Monitoring in Dogs
    Angelucci, Alessandra
    Birettoni, Francesco
    Bufalari, Antonello
    Aliverti, Andrea
    IEEE ACCESS, 2024, 12 : 80308 - 80316
  • [30] Accelerometer Based System for Continuous Respiratory Rate Monitoring
    Preejith, S. P.
    JeeJani, Ahamed
    Maniyar, Paresh
    Joseph, Jayaraj
    Sivaprakasam, Mohanasankar
    2017 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2017, : 171 - 176