Development of a radioactive waste measurement system using arrayed NaI(Tl) detectors

被引:0
|
作者
Yang, Guang [1 ,2 ]
Shi, Rui [1 ]
Wang, Zhou [1 ]
Tuo, Xianguo [1 ,2 ]
机构
[1] Sichuan Univ Sci & Engn, Yibin 644005, Peoples R China
[2] Chengdu Univ Technol, Chengdu 610059, Peoples R China
基金
中国国家自然科学基金;
关键词
TGS; NaI(Tl) detector array; Detection efficiency; Minimum detectable activity; Image reconstruction; RECONSTRUCTION; IMAGE;
D O I
10.1007/s40042-024-01215-3
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The testing of radioactive waste drums is a fundamental aspect of nuclear waste disposal. To meet the need for rapid classification and measurement of radioactive wastes, we proposed a detection method based on array scintillation detectors to realize chromatographic gamma-scanning of casks of nuclear wastes. This method has been successfully developed and applied in the form of an array NaI(Tl) detector radioactive waste cask measurement system. The measurement system was divided into four main sections: (1) the radioactive source, (2) the mechanical and control system, (3) the detector system, and (4) the data processing system. The detector system comprises seven NaI(Tl) detectors, taking into account the detector layout, detection efficiency, and shielded room (to reduce the background of the environment). The results of the performance tests on the measurement system indicated that the detection limit was 6191.75 s(-1), the energy resolution was 7.698% at 661.7 keV (Cs-137), and the minimum detectable activity was 0.201 mu Ci for Cs-137 and 0.143 mu Ci for Co-60. Finally, transmission measurements and image reconstruction of the three different media in the drum using a Cs-137 radioactive source could effectively reconstruct the spatial location distribution of the media in the drum.
引用
收藏
页码:883 / 889
页数:7
相关论文
共 50 条
  • [21] Calibration of buried NaI(Tl) scintillator detectors for natural radionuclide measurement based on Monte Carlo modelling
    Chambon, A.
    Klinkby, E. B.
    Bu, M.
    Murray, A. S.
    Kook, M.
    Olesen, H.
    Nielsen, K. B.
    Lauritzen, B.
    RADIATION PHYSICS AND CHEMISTRY, 2024, 222
  • [22] Measurement of the sodium and iodine scintillation quenching factors across multiple NaI(Tl) detectors to identify systematics
    Cintas, D.
    Hedges, S.
    Thompson, W. G.
    An, P.
    Awe, C.
    Barbeau, P. S.
    de Souza, E. Barbosa
    Jo, J. H.
    Li, L.
    Martinez, M.
    Maruyama, R. H.
    Rich, G. C.
    Runge, J.
    Sarsa, M. L.
    PHYSICAL REVIEW C, 2024, 110 (01)
  • [23] A NaI(Tl)-based radioactive noble-gas monitoring system used for radiation monitoring
    Li, HuiBin
    Lei, WeiGuo
    Feng, TianCheng
    Huang, DingWei
    Tian, ZiNing
    APPLIED RADIATION AND ISOTOPES, 2020, 163
  • [24] A Bayesian Approach for Remote Depth Estimation of Buried Low-Level Radioactive Waste with a NaI(Tl) Detector
    Kim, Jinhwan
    Lim, Kyung Taek
    Park, Kyeongjin
    Cho, Gyuseong
    SENSORS, 2019, 19 (24)
  • [25] Spectrum-dose conversion operator of NaI(Tl) and CsI(Tl) scintillation detectors for air dose rate measurement in contaminated environments
    Tsuda, Shuichi
    Saito, Kimiaki
    JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2017, 166 : 419 - 426
  • [26] Temperature Peak-Drift Correction Method for NaI(Tl) Detectors Using the Background Peak
    Wen, Songlin
    Zhou, Rong
    SENSORS, 2024, 24 (08)
  • [27] NaI(Tl) Detector Efficiency Computation Using Radioactive Parallelepiped Sources Based on Efficiency Transfer Principle
    Badawi, Mohamed S.
    Gouda, Mona M.
    El-Khatib, Ahmed M.
    Thabet, Abouzeid A.
    Salim, Ahmed A.
    Abbas, Mahmoud I.
    SCIENCE AND TECHNOLOGY OF NUCLEAR INSTALLATIONS, 2015, 2015
  • [28] First results of a dedicated breast PET imager, BPET, using NaI(Tl) curve plate detectors
    Freifelder, R
    Cardi, C
    Grigoras, L
    Saffer, JR
    Karp, JS
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1241 - 1245
  • [29] Gamma-Ray Source Positioning Using Array NaI(Tl) Detectors in the Radiation Portal Monitors
    Qibiao Wang
    Chao Deng
    Jianbo Yang
    Rui Shi
    Songbai Zhang
    Xianguo Tuo
    Journal of the Korean Physical Society, 2019, 75 : 196 - 201
  • [30] Performance of a whole-body PET scanner using curve-plate NaI(Tl) detectors
    Adam, LE
    Karp, JS
    Daube-Witherspoon, ME
    Smith, RJ
    JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (12) : 1821 - 1830