Calibration of buried NaI(Tl) scintillator detectors for natural radionuclide measurement based on Monte Carlo modelling

被引:0
|
作者
Chambon, A. [1 ]
Klinkby, E. B. [1 ]
Bu, M. [1 ]
Murray, A. S. [2 ,3 ]
Kook, M. [1 ]
Olesen, H. [1 ]
Nielsen, K. B. [1 ]
Lauritzen, B. [1 ]
机构
[1] Tech Univ Denmark, Dept Phys, DTU Riso Campus, DK-4000 Roskilde, Denmark
[2] Aarhus Univ, Dept Geosci, Nord Lab Luminescence Dating, DTU Riso Campus, DK-4000 Roskilde, Denmark
[3] DTU Phys, DTU Riso Campus, DK-4000 Roskilde, Denmark
关键词
NaI(Tl) detector; Field measurement; Calibration; MCNP; RESPONSE FUNCTIONS; GAMMA;
D O I
10.1016/j.radphyschem.2024.111803
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Measurements of naturally occurring concentrations of 40K, and of the decay series of 238U and 232Th, are of interest in the earth sciences in general, and in particular, scintillator-based gamma spectrometers can be used for the low cost determination of burial dose rates in natural geological samples. We are currently developing a robust, portable, wireless detector specifically intended for field measurement of natural radionuclide concentrations and hence, the calculation of dose rates. One of the challenges in developing and applying such an instrument is reliable calibration. Most calibrations of field instruments depend on access to non-finite matrices of known K, U, Th activity concentrations, in either a 4 pi or 2 pi geometry; these are only available at a few facilities around the world. Here we investigate an alternative approach, based on the measurement of small samples containing well-known activity concentrations of only K or U or Th, and Monte-Carlo radiation transport modelling to convert the observed spectra into those expected from specific activity concentrations in a nonfinite 4 pi geometry. We first validate our modelling procedure by simulating these observed spectra. The nonfinite matrix calibration spectra are then predicted, and least-squares fitted to the spectrum observed at the centre of a 1 m3 of granite chips; the resulting predicted U, Th and K activity concentrations are compared with independently known values.
引用
收藏
页数:8
相关论文
共 46 条
  • [1] Monte-Carlo modeling of a NaI(Tl) scintillator
    S. Ashrafi
    S. Anvarian
    S. Sobhanian
    Journal of Radioanalytical and Nuclear Chemistry, 2006, 269 : 95 - 98
  • [2] Monte-Carlo modeling of a NaI(Tl) scintillator
    Ashrafi, S.
    Anvarian, S.
    Sobhanian, S.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2006, 269 (01) : 95 - 98
  • [3] Monte Carlo calibration of whole-body counters with NaI(Tl) detectors in stretcher geometry
    Breustedt, B.
    Eschner, W.
    RADIATION PROTECTION DOSIMETRY, 2010, 139 (04) : 510 - 518
  • [4] MONTE CARLO SIMULATION APPLIED TO HEXAGONAL NaI(Tl) DETECTORS.
    Jerrestam, Dan
    Nuclear instruments and methods in physics research, 1986, 244 (03): : 542 - 546
  • [5] Sensitivity calibration of NaI(Tl) scintillator based on single photon counting
    He, Xiaoan
    Li, Chaoguang
    Yi, Rongqing
    Su, Ming
    Zhang, Qiwei
    Ruan, Xichao
    Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2015, 27 (03):
  • [6] Modelling of scintillator based flat-panel detectors with Monte-Carlo simulations
    Reims, N.
    Sukowski, F.
    Uhlmann, N.
    JOURNAL OF INSTRUMENTATION, 2011, 6
  • [7] MONTE CARLO CALCULATED EFFICIENCY OF NAI(TL) DETECTORS IMBEDDED INTO (OR EXTERNAL TO) CYLINDRICAL SOURCES
    DEFRANCESCHI, L
    PAGNI, F
    NUCLEAR INSTRUMENTS & METHODS, 1969, 70 (03): : 325 - +
  • [8] Characterization of CsI(Tl) and LYSO(Ce) scintillator detectors by measurements and Monte Carlo simulations
    Mouhti, I.
    Elanique, A.
    Messous, M. Y.
    Benahmed, A.
    McFee, J. E.
    Elgoub, Y.
    Griffith, P.
    APPLIED RADIATION AND ISOTOPES, 2019, 154
  • [9] MONTE-CARLO CALCULATION OF INTRINSIC GAMMA-RAY EFFICIENCIES OF CYLINDRICAL NAI(TL) DETECTORS
    GROSSWENDT, B
    WAIBEL, E
    NUCLEAR INSTRUMENTS & METHODS, 1976, 133 (01): : 25 - 28
  • [10] Monte Carlo modelling provides accurate calibration factor for radionuclide activity meters
    Zagni, F.
    Cicoria, G.
    Pancaldi, D.
    Sabba, N.
    Moretti, E.
    Lodi, F.
    Marengo, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S394 - S394