Experiment and design improvement of an axial flow blood pump

被引:0
|
作者
School of Jet Propulsion, Beijing University of Aeronautics and Astronautics, Beijing 100083, China [1 ]
不详 [2 ]
机构
来源
Hangkong Dongli Xuebao | 2008年 / 6卷 / 1061-1066期
关键词
Flow separation - Shear stress - Shear flow - Diseases - Pumps - Blood - Blood vessels - Efficiency;
D O I
暂无
中图分类号
学科分类号
摘要
The numerical investigation and experimental results of the proto-pump LAP-00 show that at design rotational speed of 9000 r/min, the flow rate corresponding to the highest efficiency point is 11 L/min, which is much higher than clinical use (3-8 L/min), causing the pump to work at off design point in clinic use with a low efficiency. Besides, in the diffuser region, flow separation and recirculation may cause blood stagnation and thrombosis. The improved design for pump LAP-23 decreases the design rotational speed from 9000 r/min to 8000 r/min, thus reduces the high shear stress on blood cells at impeller tip; meanwhile, it decreases the design flow rate to 6 L/min and eliminates flow separation and recirculation in the diffuser region.
引用
收藏
相关论文
共 50 条
  • [21] Flow Visualization in the Outflow Cannula of an Axial Blood Pump
    Liu, Guangmao
    Zhang, Yan
    Chen, Haibo
    Sun, Hansong
    Zhou, Jianye
    Hu, Shengshou
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 117 - 122
  • [22] Miniaturization of a Magnetically Levitated Axial Flow Blood Pump
    Cheng, Shanbao
    Olles, Mark W.
    Olsen, Don B.
    Joyce, Lyle D.
    Day, Steven W.
    ARTIFICIAL ORGANS, 2010, 34 (10) : 807 - 815
  • [23] Induction of ventricular collapse by an axial flow blood pump
    Amin, Devin V.
    Antaki, James F.
    Litwak, Philip
    Thomas, Douglas
    Wu, Zhongjun J.
    Watach, Mary
    ASAIO Journal, 44 (05):
  • [24] Influence of impeller design on hemolysis of an axial blood pump
    Liu, Lei
    Wang, Fangqun
    Wu, Qinlin
    Wu, Wenjue
    Qian, Kunxi
    Ji, Jinghua
    Zhao, Wenxiang
    He, Weixing
    Li, Tianbo
    MEDICAL MATERIALS AND ENGINEERING, 2012, 140 : 162 - 166
  • [25] In vitro performance of a centrifugal, a mixed flow, and an axial flow blood pump
    Medical Informatics, Miyazaki Medical College Hospital, Miyazaki, Japan
    不详
    不详
    不详
    Artif. Organs, 5 (366-370):
  • [26] In vitro performance of a centrifugal, a mixed flow, and an axial flow blood pump
    Araki, K
    Anai, H
    Oshikawa, M
    Nakamura, K
    Onitsuka, T
    ARTIFICIAL ORGANS, 1998, 22 (05) : 366 - 370
  • [27] Development of miniaturized mass flow meter for an axial flow blood pump
    Kosaka, Ryo
    Maruyama, Osamu
    Nishida, Masahiro
    Yamane, Takashi
    ARTIFICIAL ORGANS, 2007, 31 (05) : 412 - 419
  • [28] A HYDRAULIC DESIGN PROCEDURE OF AXIAL-FLOW PUMP
    Shen, Zhenhua
    Rasmussen, Svend
    Ma, Xiaofen
    Jacobsen, Christian Brix
    PROCEEDINGS OF THE ASME/JSME/KSME JOINT FLUIDS ENGINEERING CONFERENCE, 2015, VOL 1A, SYMPOSIA, PT 2, 2016,
  • [29] Design and Simulation of Axial Piston Pump with Constant Flow
    Yin, Jun
    Wang, Xinyu
    Wang, Lu
    Wang, Liang
    2012 10TH IEEE INTERNATIONAL CONFERENCE ON INDUSTRIAL INFORMATICS (INDIN), 2012, : 591 - 595
  • [30] Structural improvement study of streamline design method, conical hub, and auxiliary blades for axial blood pump
    Yu, Zheqin
    Tan, Jianping
    Wang, Shuai
    Guo, Bin
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2021, 44 (04): : 251 - 261