Performance evaluation of high-volume evacuation for removing droplets during dental treatment

被引:1
|
作者
Xing, Chaojie [1 ,2 ]
Ai, Zhengtao [1 ,2 ]
Mak, Cheuk Ming [3 ]
Wong, Hai Ming [4 ]
机构
[1] Hunan Univ, Dept Civil Engn, Changsha 410082, Peoples R China
[2] Hunan Univ, Natl Ctr Int Res Collaborat Bldg Safety & Environm, Changsha, Hunan, Peoples R China
[3] Hong Kong Polytech Univ, Dept Bldg Environm & Energy Engn, Hung Hom, Hong Kong, Peoples R China
[4] Univ Hong Kong, Fac Dent, Pok Fu Lam, Hong Kong, Peoples R China
关键词
high-volume evacuation (HVE); ultrasonic scaling instruments (USI); powder jet handpieces (PJH); high-speed air turbine handpiece (HATH); fallow time (FT); computational fluid dynamics (CFD); BACTERIAL AEROSOLS; REDUCTION; PARTICLES; EFFICACY; DYNAMICS; DEVICE; SALIVA;
D O I
10.1007/s12273-024-1187-5
中图分类号
O414.1 [热力学];
学科分类号
摘要
The high-volume evacuation (HVE) is commonly employed as a primary source control measure for removing splatter emitted from mouth during dental treatments, but there is still a lack of comprehensive understanding of its efficiency. Based on our previous experiments on the emission characteristics during dental treatments, this study employed computational fluid dynamics (CFD) simulations to investigate the impact of emission parameters (droplet size, emission velocity, emission angle), HVE usage methods (distance between HVE and the droplet release source), and HVE suction flow rates on its removal efficiency. The effect of HVE on fallow time (FT) was also examined. Cumulative removal efficiency that accurately reflected the HVE effect was proposed as an evaluation index. It was found that emission velocity and distance between HVE and the source were key factors determining cumulative removal efficiency. When the distance was 4 cm, the cumulative removal efficiencies for low-velocity (0.8 m/s), medium-velocity (3.4 m/s), and high-velocity droplets (6.0 m/s) were approximately 97.9%, 73.6%, and 58.0%, respectively. For high-velocity droplets at 6.0 m/s, decreasing the distance from 4 cm to 2 cm and 1 cm increased the cumulative removal efficiency from 58.0% to 76.7% and 100%. This study was expected to enhance the understanding of HVE performance and provide information on its usage method. It also indicated the need for developing advanced control measures that could have high efficiency in removing both low-velocity and high-velocity droplets.
引用
收藏
页码:2263 / 2280
页数:18
相关论文
共 50 条
  • [1] High-volume evacuation mitigates viral aerosol spread in dental procedures
    Malmgren, Rasmus
    Vaelimaa, Hanna
    Oksanen, Lotta
    Sanmark, Enni
    Nikuri, Petra
    Heikkilae, Paavo
    Hakala, Jani
    Ahola, Aleksi
    Yli-Urpo, Simeoni
    Palomaeki, Ville
    Asmi, Eija
    Sofieva, Svetlana
    Rostedt, Antti
    Laitinen, Sirpa
    Romantschuk, Martin
    Sironen, Tarja
    Atanasova, Nina
    Paju, Susanna
    Lahdentausta-Suomalainen, Laura
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [2] High-volume evacuation mitigates viral aerosol spread in dental procedures
    Rasmus Malmgren
    Hanna Välimaa
    Lotta Oksanen
    Enni Sanmark
    Petra Nikuri
    Paavo Heikkilä
    Jani Hakala
    Aleksi Ahola
    Simeoni Yli-Urpo
    Ville Palomäki
    Eija Asmi
    Svetlana Sofieva
    Antti Rostedt
    Sirpa Laitinen
    Martin Romantschuk
    Tarja Sironen
    Nina Atanasova
    Susanna Paju
    Laura Lahdentausta-Suomalainen
    Scientific Reports, 13
  • [3] Droplet Distribution in a University Dental Clinic Setting: The Importance of High-Volume Evacuation
    Gualtieri, Linda
    Yong, Ronald
    Ah-Kion, Jessley
    Jamil, Amanda L. A.
    Bazaei, Asmae
    Kotecha, Jhanvi
    Long, Sharron
    Silcock, Gloria
    Miller, Catherine M.
    HEALTHCARE, 2022, 10 (09)
  • [4] Visualization of airborne droplets generated with dental handpieces and verification of the efficacy of high-volume evacuators: an in vitro study
    Min Jung Kim
    Mana Kuroda
    Yoshikazu Kobayashi
    Takahisa Yamamoto
    Takako Aizawa
    Koji Satoh
    BMC Oral Health, 23
  • [5] Visualization of airborne droplets generated with dental handpieces and verification of the efficacy of high-volume evacuators: an in vitro study
    Kim, Min Jung
    Kuroda, Mana
    Kobayashi, Yoshikazu
    Yamamoto, Takahisa
    Aizawa, Takako
    Satoh, Koji
    BMC ORAL HEALTH, 2023, 23 (01)
  • [6] How the high-volume evacuation alters the flow-field and particle removal characteristics in the mock-up dental clinic
    Li, Xiujie
    Mak, Cheuk Ming
    Ma, Kuen Wai
    Wong, Hai Ming
    BUILDING AND ENVIRONMENT, 2021, 205 (205)
  • [7] The Performance of High-Volume Bronchoalveolar Lavage for the Evaluation and Diagnosis of Interstitial Lung Disease
    Singh, Karanvir
    Khan, Sadiq
    Agarwal, Sushil
    Walters, Gareth
    McGrath, Emmet E.
    ANNALS OF THORACIC SURGERY, 2014, 97 (02): : 720 - 722
  • [8] Comparing Isovac® to High-Volume Evacuation on Aerosols and Surface Contamination During Operative Dentistry in Children: A Randomized, Controlled Trial
    Dworkin, Sara D.
    Kotak, Nidhi
    Manchanda, Mega
    Lee, Susan
    PEDIATRIC DENTISTRY, 2024, 46 (06) : 395 - 399
  • [9] Performance evaluation of high-volume electret filter air samplers in aerosol microbiome research
    Kari Oline Bøifot
    Jostein Gohli
    Gunnar Skogan
    Marius Dybwad
    Environmental Microbiome, 15
  • [10] Performance evaluation of high-volume electret filter air samplers in aerosol microbiome research
    Boifot, Kari Oline
    Gohli, Jostein
    Skogan, Gunnar
    Dybwad, Marius
    ENVIRONMENTAL MICROBIOME, 2020, 15 (01)