Effect of Respirophasic Displacement of the Inferior Vena Cava on Size Measurement in 2-D Ultrasound Imaging

被引:0
|
作者
Policastro, Piero [1 ]
Ermini, Leonardo [2 ]
Civera, Stefania [3 ]
Albani, Stefano [3 ,4 ]
Musumeci, Giuseppe [3 ]
Roatta, Silvestro [2 ]
Mesin, Luca [1 ]
机构
[1] Politecn Torino, Dept Elect & Telecommun, Math Biol & Physiol, Turin, Italy
[2] Univ Torino, Dept Neurosci, Turin, Italy
[3] Osped Ordine Mauriziano Torino, Div Cardiol, Turin, Italy
[4] Umberto Parini Reg Hosp, Div Cardiol, Aosta, Italy
来源
ULTRASOUND IN MEDICINE AND BIOLOGY | 2024年 / 50卷 / 12期
关键词
Inferior vena cava; Ultrasound; Tracking;
D O I
10.1016/j.ultrasmedbio.2024.07.005
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective: Volume status assessment of a patient by ultrasound (US) imaging of the inferior vena cava (IVC) is important for the diagnosis and prognosis of various clinical conditions. In order to improve the clinical investigation of IVC, which is mainly based on unidirectional US (in M-mode), automated processing of 2-D US scans (in Bmode) has enabled tissue movement tracking on the visualized plane and can average this in various directions. However, IVC geometry outside of the visualized plane is not under control and could result in errors that have not yet been evaluated. Methods: We used a method that integrates information from long- and short-axis IVC views (simultaneously acquired in the X-plane) to assess challenges in IVC diameter estimations using 2-D US scans in eight healthy subjects. Results: Relative movements between the US probe and IVC induced the following problems when assessing IVC diameter via 2-D view: a median error (i.e., absolute difference with respect to diameter measured in the X-plane) of 17% using 2-D US scans in the long-axis view of the IVC affected by medio-lateral displacements (median: 4 mm); and a median error of 7% and 9% when measuring the IVC diameter from a short-axis view in the presence of pitch angle (median: 0.12 radians) and cranio-caudal movement (median: 15 mm), respectively. Conclusion: Relative movements in the IVC that are out of view of B-mode scans cannot be detected, which results in challenges in IVC diameter estimation.
引用
收藏
页码:1785 / 1792
页数:8
相关论文
共 50 条
  • [41] SENSOR SYSTEM FOR MEASUREMENT OF 2-D DISPLACEMENT AND FORCE VECTORS
    JANOCHA, H
    SENSORS AND ACTUATORS A-PHYSICAL, 1992, 35 (02) : 143 - 149
  • [42] 2-D NIR imaging reconstruction with ultrasound guidance
    Huang, MM
    Xie, TQ
    Chen, NG
    Piao, DQ
    Zhu, Q
    2002 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING, PROCEEDINGS, 2002, : 1031 - 1034
  • [43] SPECTRAL ESTIMATION FOR 2-D DOPPLER ULTRASOUND IMAGING
    COSTA, A
    DEFRANCISCIS, A
    DEGLORIA, A
    FARABOSCHI, P
    OLIVIERI, M
    ELECTRONICS LETTERS, 1992, 28 (23) : 2177 - 2179
  • [44] The value of the inferior vena cava ultrasound in the decision to hospitalise in patients with acute decompensated heart failure; the best sonographic measurement method?
    Yamanoglu, Adnan
    Celebi Yamanoglu, Nalan Gokce
    Ozturk, Semi
    Cakmak, Sumeyye
    Akay, Serhat
    Akyol, Pinar Yesim
    Sogut, Ozgur
    ACTA CARDIOLOGICA, 2021, 76 (03) : 245 - 257
  • [45] Three-Dimensional Ultrasound Imaging of Inferior Vena Cava Outperforms Standard Two-Dimensional Imaging as a Diagnostic Tool in Heart Failure
    Struyk, Griffin
    Essa, Mohammed
    Flueckiger, Peter
    Bregasi, Alda
    McNamara, Robert
    Sugeng, Lissa
    Testani, Jeffrey
    Lin, Ben
    JOURNAL OF CARDIAC FAILURE, 2018, 24 (08) : S41 - S41
  • [46] Central Venous Pressure Estimation by Ultrasound Measurement of Inferior Vena Cava and Aorta Diameters in Pediatric Critical Patients: An Observational Study
    Mercolini, Federico
    Di Leo, Valentina
    Bordin, Giulia
    Melotti, Roberto
    Sperotto, Francesca
    Pettenazzo, Andrea
    Amigoni, Angela
    Tosoni, Alvise
    PEDIATRIC CRITICAL CARE MEDICINE, 2021, 22 (01) : E1 - E9
  • [47] Feasibility and accuracy of nurse performed pocket-size ultrasound imaging of the pleura and vena cava inferior to assess volume state in patients with heart failure in an outpatient clinic
    Gundersen, G. H.
    Graven, T.
    Haug, H. H.
    Norekvaal, T. M.
    Skjetne, K.
    Kleinau, O.
    Dalen, H.
    EUROPEAN HEART JOURNAL, 2014, 35 : 837 - 837
  • [48] Ultrasound imaging and central venous pressure in spontaneously breathing patients: a comparison of ultrasound-based measures of internal jugular vein and inferior vena cava
    Parenti, Nicola
    Bastiani, Luca
    Tripolino, Cesare
    Bacchilega, Igor
    ANAESTHESIOLOGY INTENSIVE THERAPY, 2022, 54 (02) : 150 - 155
  • [49] TRADITIONAL M-MODE ASSESSMENT OF INFERIOR VENA CAVA COLLAPSIBILITY IS NOT RELIABLE - A ROLE FOR 3D IMAGING?
    Gopalakrishnan, Prabhakaran
    Potakamuri, Lakshmi
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (18) : 1395 - 1395
  • [50] RIGHT ATRIAL PRESSURE ESTIMATES BY 2D AND 3D INFERIOR VENA CAVA MEASUREMENTS ARE DISCORDANT
    Essa, Mohammed
    Flueckiger, Peter
    McNamara, Robert
    Sugeng, Lissa
    Testani, Jeffrey
    Lin, Ben