Effect of Respirophasic Displacement of the Inferior Vena Cava on Size Measurement in 2-D Ultrasound Imaging

被引:0
|
作者
Policastro, Piero [1 ]
Ermini, Leonardo [2 ]
Civera, Stefania [3 ]
Albani, Stefano [3 ,4 ]
Musumeci, Giuseppe [3 ]
Roatta, Silvestro [2 ]
Mesin, Luca [1 ]
机构
[1] Politecn Torino, Dept Elect & Telecommun, Math Biol & Physiol, Turin, Italy
[2] Univ Torino, Dept Neurosci, Turin, Italy
[3] Osped Ordine Mauriziano Torino, Div Cardiol, Turin, Italy
[4] Umberto Parini Reg Hosp, Div Cardiol, Aosta, Italy
来源
ULTRASOUND IN MEDICINE AND BIOLOGY | 2024年 / 50卷 / 12期
关键词
Inferior vena cava; Ultrasound; Tracking;
D O I
10.1016/j.ultrasmedbio.2024.07.005
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective: Volume status assessment of a patient by ultrasound (US) imaging of the inferior vena cava (IVC) is important for the diagnosis and prognosis of various clinical conditions. In order to improve the clinical investigation of IVC, which is mainly based on unidirectional US (in M-mode), automated processing of 2-D US scans (in Bmode) has enabled tissue movement tracking on the visualized plane and can average this in various directions. However, IVC geometry outside of the visualized plane is not under control and could result in errors that have not yet been evaluated. Methods: We used a method that integrates information from long- and short-axis IVC views (simultaneously acquired in the X-plane) to assess challenges in IVC diameter estimations using 2-D US scans in eight healthy subjects. Results: Relative movements between the US probe and IVC induced the following problems when assessing IVC diameter via 2-D view: a median error (i.e., absolute difference with respect to diameter measured in the X-plane) of 17% using 2-D US scans in the long-axis view of the IVC affected by medio-lateral displacements (median: 4 mm); and a median error of 7% and 9% when measuring the IVC diameter from a short-axis view in the presence of pitch angle (median: 0.12 radians) and cranio-caudal movement (median: 15 mm), respectively. Conclusion: Relative movements in the IVC that are out of view of B-mode scans cannot be detected, which results in challenges in IVC diameter estimation.
引用
收藏
页码:1785 / 1792
页数:8
相关论文
共 50 条
  • [31] Expanding options for bedside placement of inferior vena cava filters with intravascular ultrasound when transabdominal duplex ultrasound imaging is inadequate
    Garrett, JV
    Passman, MA
    Guzman, RJ
    Dattilo, JB
    Naslund, TC
    ANNALS OF VASCULAR SURGERY, 2004, 18 (03) : 329 - 334
  • [32] INFERIOR VENA CAVA MEASUREMENT BY FOCUSED CARDIAC ULTRASOUND IN ACUTE DECOMPENSATED HEART FAILURE PREVENTS HOSPITAL READMISSIONS
    Laffin, Luke J.
    Patel, Amit
    Saha, Narayan
    Barbat, Julian
    Hall, James K.
    Cain, Matthew
    Parikh, Kishan
    Shah, Jay
    Spencer, Kirk
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 63 (12) : A542 - A542
  • [33] Comparative analysis between DSA and color Doppler ultrasound in measurement of kidney and inferior vena cava diameter in adults
    Yong, Jun
    Yang, Mu
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2018, 11 (12): : 13448 - 13452
  • [34] Intravascular Ultrasound and Multidetector Computed Tomography Imaging of Polymethylmethacrylate Cement Leakage Inducing Inferior Vena Cava Thrombosis
    Chang, Rei-Yeuh
    Chen, Tsung-Hsien
    Kuo, Chu-Jen
    Tsai, Han-Lin
    CIRCULATION-CARDIOVASCULAR IMAGING, 2023, 16 (02) : 220 - 223
  • [35] SEMI-AUTOMATED TRACKING AND CONTINUOUS MONITORING OF INFERIOR VENA CAVA DIAMETER IN SIMULATED AND EXPERIMENTAL ULTRASOUND IMAGING
    Mesin, Luca
    Pasquero, Paolo
    Albani, Stefano
    Porta, Massimo
    Roatta, Silvestro
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (03): : 845 - 857
  • [36] A TIME-SHARED ULTRASOUND DOPPLER MEASUREMENT AND 2-D IMAGING-SYSTEM
    KRISTOFFERSEN, K
    ANGELSEN, BAJ
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (05) : 285 - 295
  • [37] A critical evaluation of ultrasound measurement of inferior vena cava diameter in assessing dry weight in normotensive and hypertensive hemodialysis patients
    Katzarski, KS
    Nisell, J
    Randmaa, I
    Danielsson, A
    Freyschuss, U
    Bergstrom, J
    AMERICAN JOURNAL OF KIDNEY DISEASES, 1997, 30 (04) : 459 - 465
  • [38] The use of intravascular ultrasound imaging to improve use of inferior vena cava filters in a high-risk bariatric population
    Kardys, Clark M.
    Stoner, Michael C.
    Manwaring, Mark L.
    Bogey, William M.
    Parker, Frank M.
    Powell, Steve
    JOURNAL OF VASCULAR SURGERY, 2007, 46 (06) : 1248 - 1252
  • [39] MULTI-DIRECTIONAL ASSESSMENT OF RESPIRATORY AND CARDIAC PULSATILITY OF THE INFERIOR VENA CAVA FROM ULTRASOUND IMAGING IN SHORT AXIS
    Mesin, Luca
    Pasquero, Paolo
    Roatta, Silvestro
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2020, 46 (12): : 3475 - 3482
  • [40] 2-D displacement measurement system based on image processing
    Hou, Jie
    Qian, Jiaru
    Zhang, Weijing
    Zhao, Zuozhou
    Pan, Peng
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 2010, 50 (06): : 826 - 829