Implementation of Bioelectrical Impedance Measuring Instrument Based on Embedded System

被引:0
|
作者
Pawar C. [1 ,3 ]
Khan M. [2 ]
Saini J.P. [3 ]
Singh D. [4 ]
Bhardwaj M. [5 ]
Hu Y.-C. [6 ]
机构
[1] Department of Electronics and Communications Engineering, Netaji Subhash University of Technology, Delhi
[2] Electrical Department, Jamia Milia Islamia, New Delhi
[3] Netaji Subhash University of Technology, Delhi
[4] Department of Applied Mathematics, Allenhouse Institute of Technology, Kanpur
[5] Department of Computer Science and Engineering, KIET Group of Institutions, Delhi-NCR, Ghaziabad
[6] Department of Computer Science, Tunghai University, Taichung City
关键词
Statistics - User interfaces - Wireless local area networks (WLAN);
D O I
10.1155/2024/1024006
中图分类号
学科分类号
摘要
The present research aims to measure the segmental bioelectrical impedance (BI) of the human body at multifrequency, using a server user interface-based prototype, which provides subjects with measured data online anywhere accessed by their unique identifications. The present research measures the BI of the human leg and arm at a multifrequency range of 50-400 kHz with a developed and standard device. Recorded data can be transferred to the subject using Wi-Fi technology with their unique identifications and password. The system uses Wi-Fi interfacing for real-time data measurement and online data storage. The prototype can be used commercially. The compact size of the prototype becomes the demand of the common population. The pocket carry size makes it easy to carry anywhere for regular monitoring of the human body to prevent critical disease. The resultant data show that the mean and standard deviation of the left and right leg are 282.2 ± 14.29 and 274.80 ± 13.91 Ω, respectively. Similarly, in the case of the left arm and right arm, the mean and standard deviation are 325.41 ± 16.54 and 320.73 ± 16.07 Ω, respectively. The relative error between developed and standard devices is 3.53%. Results show that the left leg and arm impedance is always greater. However, the right leg and arm muscles are stronger than the left one, with less impedance at all frequencies. © 2024 Charu Pawar et al.
引用
收藏
相关论文
共 50 条
  • [31] An Algorithm and Implementation System for Measuring Impedance in the D-Q Domain
    Francis, Gerald
    Burgos, Rolando
    Boroyevich, Dushan
    Wang, Fred
    Karimi, Kamiar
    2011 IEEE ENERGY CONVERSION CONGRESS AND EXPOSITION (ECCE), 2011, : 3221 - 3228
  • [32] Implementation of a prototype embedded system for in-car multipoint temperature measuring
    Moise, Madalin Vasile
    Svasta, Paul Mugur
    Mazare, Alin Gheorghita
    2019 22ND EUROPEAN MICROELECTRONICS AND PACKAGING CONFERENCE & EXHIBITION (EMPC), 2019,
  • [33] A Simple Rapid Method for Measuring Liver Steatosis Using Bioelectrical Impedance
    Yoshimoto-Haramura, Tomoko
    Hara, Takanobu
    Soyama, Akihiko
    Kugiyama, Tota
    Matsushima, Hajime
    Matsuguma, Kunihito
    Imamura, Hajime
    Tanaka, Takayuki
    Adachi, Tomohiko
    Hidaka, Masaaki
    Okabe, Shuichi
    Murata, Masakazu
    Eguchi, Susumu
    IN VIVO, 2022, 36 (02): : 570 - 575
  • [34] Design, Implementation and Data Analysis of an Embedded System for Measuring Environmental Quantities
    Pies, Martin
    Hajovsky, Radovan
    Velicka, Jan
    SENSORS, 2020, 20 (08)
  • [35] Possibilities of use bioelectrical impedance analysis as measuring technique in prevention of osteoporosis
    Cvetko, Erna Davidovic
    Drenjancevic, Ines
    Nesic, Nebojsa
    Ahic, Jasminka Milas
    PERIODICUM BIOLOGORUM, 2014, 116 (01) : 65 - 70
  • [36] Bioelectrical impedance: A new method for measuring post-traumatic swelling
    King, Richard J.
    Clamp, Jonathan A.
    Hutchinson, James W.
    Moran, Christopher G.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (07) : 462 - 468
  • [37] Non-Invasive Blood Glucose Estimation System Based on a Neural Network with Dual-Wavelength Photoplethysmography and Bioelectrical Impedance Measuring
    Yen, Chih-Ta
    Chen, Un-Hung
    Wang, Guo-Chang
    Chen, Zong-Xian
    SENSORS, 2022, 22 (12)
  • [38] A Standardized Protocol for Measuring Bioelectrical Impedance in Green Turtles (Chelonia mydas)
    Kophamel, Sara
    Ward, Leigh C.
    Ariel, Ellen
    Mendez, Diana
    O'Brien, Lauren M.
    Burchell, Lauren
    Munns, Suzanne L.
    PHYSIOLOGICAL AND BIOCHEMICAL ZOOLOGY, 2023, 96 (02): : 87 - 99
  • [39] The effect of electrode placement in measuring ipsilateral/contralateral segmental bioelectrical impedance
    Cornish, BH
    Eles, PT
    Thomas, BJ
    Ward, LC
    IN VIVO BODY COMPOSITION STUDIES, 2000, 904 : 221 - 224
  • [40] Design of a Self-Measuring Device Based on Bioelectrical Impedance Analysis for Regular Monitoring of Rheumatoid Arthritis
    Jeong, Juyoung
    Park, Yun Soo
    Lee, Eunchae
    Choi, Seoyoun
    Lim, Dokshin
    Kim, Jiho
    SENSORS, 2024, 24 (08)