Research on Calibration Method of Flexible Blood Pressure Sensor Based on Photoplethysmography

被引:0
|
作者
Shi, Wei [1 ,2 ]
Dong, Lin-Xi [1 ,2 ]
Weng, Bin-Hui [1 ,2 ]
Cheng, Jia-Gen [1 ,2 ]
Yang, Wei-Huang [1 ,2 ]
Liu, Chao-Ran [1 ,2 ]
机构
[1] School of Electronic Information, Hangzhou Dianzi University, Zhejiang, Hangzhou,310018, China
[2] Smart Microsensor and Microsystem Engineering Research Center, Ministry of Education, Zhejiang, Hangzhou,310018, China
来源
Tien Tzu Hsueh Pao/Acta Electronica Sinica | 2024年 / 52卷 / 08期
关键词
Blood pressure;
D O I
10.12263/DZXB.20230191
中图分类号
学科分类号
摘要
The flexible sensor based on photo plethysmo graphy (PPG) can detect heart rate (HR) and blood pressure (BP), but the calibration of their detection results is rarely reported. Therefore, this paper proposes a reflective PPG heart rate detection and blood pressure calibration system based on simulated blood circulation. The peristaltic pump is used to generate pulsating flow,and the frequency and pressure of simulated blood delivery are controlled by adjusting its rotational speed,thus causing the change of the volume of simulated blood in the elastic latex tube and changing the signal period and intensity of reflected light, which is closer to match the actual scenario of human pulse measurement process. The mean value of heart rate detection error of the system is 0.277 78, and the 95% consistency limit is (-2.595 62, 3.151 17). The goodness of fit of measured systolic blood pressure (SBP) and diastolic blood pressure (DBP) are 0.971 85 and 0.981 11, respectively. The mean value of mean deviation (MD) ± standard deviation (SD) of SBP and DBP detected by the calibrated flexible PPG sensor on four volunteers is (1.21±2.16) mmHg and (0.76±2.02) mmHg, respectively, which are in line with and far less than the standard index of 5±8 mmHg for measuring the accuracy of blood pressure monitors set by the association for the advancement of medical instrumentation (AAMI). The results show that this system can calibrate the flexible PPG sensor accurately and efficiently, which provides the calibration basis for the accurate blood pressure detection of por- table wearable devices. © 2024 Chinese Institute of Electronics. All rights reserved.
引用
收藏
页码:2648 / 2658
相关论文
共 50 条
  • [31] Blood Pressure Evaluation Based on Photoplethysmography Using Deep Learning
    Sun Xiaoxiao
    Zhou Liang
    Liu Zhaohui
    Yu Jiangjun
    Qiao Wenlong
    AOPC 2020: OPTICAL SPECTROSCOPY AND IMAGING; AND BIOMEDICAL OPTICS, 2020, 11566
  • [32] A Novel Flexible Blood Pressure Sensor Based on Ultrathin Silicon for Continuous Monitoring
    Zeng, Yi
    Liu, Hao
    Yamamoto, Michitaka
    Takamatsu, Seiichi
    Itoh, Toshihiro
    2024 IEEE SENSORS, 2024,
  • [33] The Calibration Method for Blood Pressure Pulse Wave Measurement Based on Arterial Tonometry Method
    Shimura, Koichi
    Hori, Masataka
    Dohi, Tctsuji
    Takao, Hiroyuki
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 3642 - 3645
  • [34] Research on Creep Resistance of Flexible Pressure Sensor Based on Anhydride Curing Agent
    Xu, Fenlan
    Wang, Zhenguo
    Li, Xiuyan
    Liang, Mengmeng
    Liu, Ruping
    Wang, Wei
    Li, Luhai
    APPLIED SCIENCES IN GRAPHIC COMMUNICATION AND PACKAGING, 2018, 477 : 901 - 908
  • [35] On Evaluating Blood Pressure Through Photoplethysmography
    Sannino, Giovanna
    De Falco, Ivanoe
    De Pietro, Giuseppe
    INTERNET OF THINGS: IOT INFRASTRUCTURES, PT I, 2016, 169 : 523 - 529
  • [36] The Research of Optical Fiber Sensor Calibration based on Acoustic Sensor Calibration System
    Zhang, Tao
    Chen, Yongxia
    Cui, Yaoyao
    Du, Yinfei
    Fu, BeiBei
    Jia, Qishen
    Gao, Shen
    Li, Yuanyao
    Wang, Peng
    Zheng, Hui
    2019 INTERNATIONAL CONFERENCE ON OPTICAL INSTRUMENTS AND TECHNOLOGY: OPTOELECTRONIC MEASUREMENT TECHNOLOGY AND SYSTEMS, 2020, 11439
  • [37] Research of Grain Moisture Measurement Method Based on Pressure Sensor
    Li, Junxing
    Zhang, Yaqiu
    Wu, Wenfu
    Liu, Zhe
    Han, Feng
    ADVANCED MANUFACTURING TECHNOLOGY, PTS 1-4, 2012, 472-475 : 1518 - +
  • [38] Research on Sleeping Posture Recognition Method Based on Pressure Sensor
    Wang, Huabing
    Wan, Changyuan
    ADVANCES IN HUMAN FACTORS AND ERGONOMICS IN HEALTHCARE AND MEDICAL DEVICES, 2020, 957 : 235 - 244
  • [39] Research of Altitude Measurement Method Based on INS and Pressure Sensor
    Jiao, Zhun
    Hu, Jin
    Gong, Xun
    PROCEEDINGS OF 2017 IEEE 2ND INFORMATION TECHNOLOGY, NETWORKING, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (ITNEC), 2017, : 1684 - 1687
  • [40] Accurate and Flexible Calibration Method for a Class of Visual Sensor Networks
    Deng, Heng
    Yang, Kun
    Quan, Quan
    Cai, Kai-Yuan
    IEEE SENSORS JOURNAL, 2020, 20 (06) : 3257 - 3269