Research on Calibration Method of Flexible Blood Pressure Sensor Based on Photoplethysmography

被引:0
|
作者
Shi, Wei [1 ,2 ]
Dong, Lin-Xi [1 ,2 ]
Weng, Bin-Hui [1 ,2 ]
Cheng, Jia-Gen [1 ,2 ]
Yang, Wei-Huang [1 ,2 ]
Liu, Chao-Ran [1 ,2 ]
机构
[1] School of Electronic Information, Hangzhou Dianzi University, Zhejiang, Hangzhou,310018, China
[2] Smart Microsensor and Microsystem Engineering Research Center, Ministry of Education, Zhejiang, Hangzhou,310018, China
来源
Tien Tzu Hsueh Pao/Acta Electronica Sinica | 2024年 / 52卷 / 08期
关键词
Blood pressure;
D O I
10.12263/DZXB.20230191
中图分类号
学科分类号
摘要
The flexible sensor based on photo plethysmo graphy (PPG) can detect heart rate (HR) and blood pressure (BP), but the calibration of their detection results is rarely reported. Therefore, this paper proposes a reflective PPG heart rate detection and blood pressure calibration system based on simulated blood circulation. The peristaltic pump is used to generate pulsating flow,and the frequency and pressure of simulated blood delivery are controlled by adjusting its rotational speed,thus causing the change of the volume of simulated blood in the elastic latex tube and changing the signal period and intensity of reflected light, which is closer to match the actual scenario of human pulse measurement process. The mean value of heart rate detection error of the system is 0.277 78, and the 95% consistency limit is (-2.595 62, 3.151 17). The goodness of fit of measured systolic blood pressure (SBP) and diastolic blood pressure (DBP) are 0.971 85 and 0.981 11, respectively. The mean value of mean deviation (MD) ± standard deviation (SD) of SBP and DBP detected by the calibrated flexible PPG sensor on four volunteers is (1.21±2.16) mmHg and (0.76±2.02) mmHg, respectively, which are in line with and far less than the standard index of 5±8 mmHg for measuring the accuracy of blood pressure monitors set by the association for the advancement of medical instrumentation (AAMI). The results show that this system can calibrate the flexible PPG sensor accurately and efficiently, which provides the calibration basis for the accurate blood pressure detection of por- table wearable devices. © 2024 Chinese Institute of Electronics. All rights reserved.
引用
收藏
页码:2648 / 2658
相关论文
共 50 条
  • [21] Live Demonstration: A Novel Cuffless Photoplethysmography Sensor for Continuous Blood Pressure measurement
    Kao, Yung-Hua
    Chao, Paul C. -P.
    Hung, Yueh
    Wey, Chin-Long
    2017 IEEE SENSORS, 2017, : 470 - 470
  • [22] Flexible and wearable capacitive pressure sensor for blood pressure monitoring
    Bijender
    Kumar, Ashok
    SENSING AND BIO-SENSING RESEARCH, 2021, 33
  • [23] Flexible mesostructured capacitive pressure sensor for blood pressure monitoring
    Kumar, Shubham
    Yadav, Sanjay
    Kumar, Ashok
    2023 IEEE SENSORS, 2023,
  • [24] Parameters for Mean Blood Pressure Estimation Based on Electrocardiography and Photoplethysmography
    Spulak, Daniel
    Cmejla, Roman
    Fabian, Vratislav
    2011 INTERNATIONAL CONFERENCE ON APPLIED ELECTRONICS (AE), 2011,
  • [25] Development of a Device Based on Oscillometry and Photoplethysmography for Measuring Blood Pressure
    Alicia Yanez-Ocampo, Haydee
    Becerra-Luna, Brayans
    Cartas-Rosado, Raul
    Martinez-Memije, Raul
    Infante-Vazquez, Oscar
    VIII LATIN AMERICAN CONFERENCE ON BIOMEDICAL ENGINEERING AND XLII NATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, 2020, 75 : 560 - 566
  • [26] A NOVEL CONTINUOUS AND NONINVASIVE MEASUREMENT FOR BLOOD PRESSURE BASED ON PHOTOPLETHYSMOGRAPHY
    Li, Xiru
    Li, Xiaofeng
    Xu, Jinlin
    Tan, Haibo
    Yuan, Munan
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2017, 17 (03)
  • [27] Blood Pressure Estimation Based on Photoplethysmography: Finger Versus Wrist
    Paliakaite, Birute
    Charlton, Peter H.
    Rapalis, Andrius
    Plusciauskaite, Vilma
    Piartli, Povilas
    Kaniusas, Eugenijus
    Marozas, Vaidotas
    2021 COMPUTING IN CARDIOLOGY (CINC), 2021,
  • [28] Blood Pressure Estimation Algorithm Based on Photoplethysmography Pulse Analyses
    Kim, Seon-Chil
    Cho, Sung-Hyoun
    APPLIED SCIENCES-BASEL, 2020, 10 (12):
  • [29] Photoplethysmography Based Blood Pressure Monitoring Using the Senbiosys Ring
    Haddad, Serj
    Boukhayma, Assim
    Di Pietrantonio, Gilles
    Barison, Anthony
    de Preux, Gilles
    Caizzone, Antonino
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 1609 - 1612
  • [30] Novel Blood Pressure Estimation Method Using Single Photoplethysmography Feature
    Chen, Yang
    Cheng, Shuo
    Wang, Tong
    Ma, Ting
    2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2017, : 1712 - 1715