Automated hyperpolarized 129 Xe gas generator for nuclear magnetic resonance spectroscopy and imaging applications

被引:1
|
作者
Hattori, Mineyuki [1 ]
Hiraga, Takashi [1 ]
Homma, Kazuhiro [1 ]
Ohtake, Norio [2 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Res Inst Mat & Chem Measurement, Natl Metrol Inst Japan NMIJ, Tsukuba, Ibaraki 3058565, Japan
[2] Toyoko Kagaku Co Ltd, Resaerch & Dev Div, Kawasaki, Kanagawa 2118502, Japan
来源
关键词
Hyperpolarized; Xenon; Nuclear magnetic resonance; Magnetic resonance imaging; SPIN-EXCHANGE; XE-129; GAS; NMR; XENON; MRI; POLARIZATION; BEHAVIOR; TIME; FLOW;
D O I
10.1016/j.jmro.2024.100154
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We describe a method and an apparatus for producing hyperpolarized (HP) Xe gas of high concentration without Xe condensation. The HP Xe generator works under atmospheric pressure and employs a quasi-continuous method to provide a continuous supply of HP Xe gas by adopting technologies for supplying a highly pure gas, a gas control system and precise pressure control. The apparatus has a glass cell containing solid Rb metal and solid Xe in vacuum at a low temperature and is heated so that the Xe becomes a gas and Rb exists in a gas- liquid mixture. Then a magnetic field is applied with laser beam irradiation to produce polarized Xe gas of high concentration. The device was evaluated using a 2T magnetic resonance imaging (MRI) scanner. Long-term experimental operation demonstrated the continuous collection of 30 ml syringes of HP Xe gas with a sufficient polarization rate for fast one scan acquisition. A practical device for the automated manufacture of HP 129 Xe gas was thus successfully developed.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Quantifying Regional Radiation-Induced Lung Injury in Patients Using Hyperpolarized 129Xe Gas Exchange Magnetic Resonance Imaging
    Rankine, Leith J.
    Lu, Junlan
    Wang, Ziyi
    Kelsey, Christopher R.
    Marks, Lawrence B.
    Das, Shiva K.
    Driehuys, Bastiaan
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (01): : 216 - 228
  • [42] Human pulmonary imaging and spectroscopy with hyperpolarized 129Xe at 0.2T
    Patz, Samuel
    Muradian, Iga
    Hrovat, Mirko I.
    Ruset, Iulian C.
    Topulos, George
    Covrig, Silviu D.
    Frederick, Eric
    Hatabu, Hiroto
    Hersman, F. W.
    Butler, James P.
    ACADEMIC RADIOLOGY, 2008, 15 (06) : 713 - 727
  • [43] NUCLEAR MAGNETIC RESONANCE OF XE129 IN NATURAL XENON
    STREEVER, RL
    CARR, HY
    PHYSICAL REVIEW, 1961, 121 (01): : 20 - &
  • [44] An intracellular diamine oxidase triggered hyperpolarized 129Xe magnetic resonance biosensor
    Zhang, Bin
    Guo, Qianni
    Luo, Qing
    Zhang, Xiaoxiao
    Zeng, QingBin
    Zhao, Longhui
    Yuan, Yaping
    Jiang, Weiping
    Yang, Yunhuang
    Liu, Maili
    Ye, Chaohui
    Zhou, Xin
    CHEMICAL COMMUNICATIONS, 2018, 54 (97) : 13654 - 13657
  • [45] NUCLEAR MAGNETIC RESONANCE OF XE129 IN NATURAL XENON
    HUNT, ER
    CARR, HY
    PHYSICAL REVIEW, 1963, 130 (06): : 2302 - &
  • [46] XE-129 PULSE FOURIER-TRANSFORM NUCLEAR MAGNETIC-RESONANCE SPECTROSCOPY
    SCHROBILGEN, GJ
    HOLLOWAY, JH
    GRANGER, P
    BREVARD, C
    INORGANIC CHEMISTRY, 1978, 17 (04) : 980 - 987
  • [47] Detection of radiation induced lung injury in rats using dynamic hyperpolarized 129Xe magnetic resonance spectroscopy
    Fox, Matthew S.
    Ouriadov, Alexei
    Thind, Kundan
    Hegarty, Elaine
    Wong, Eugene
    Hope, Andrew
    Santyr, Giles E.
    MEDICAL PHYSICS, 2014, 41 (07)
  • [48] Magnetic resonance imaging of hyperpolarized 129Xe produced by spin exchange with diode-laser pumped Cs
    Levron, D
    Walter, DK
    Appelt, S
    Fitzgerald, RJ
    Kahn, D
    Korbly, SE
    Sauer, KL
    Happer, W
    Earles, TL
    Mawst, LJ
    Botez, D
    Harvey, M
    DiMarco, L
    Connolly, JC
    Moller, HE
    Chen, XJ
    Cofer, GP
    Johnson, GA
    APPLIED PHYSICS LETTERS, 1998, 73 (18) : 2666 - 2668
  • [49] Magnetic resonance imaging of dissolved hyperpolarized 129Xe using a membrane-based continuous flow system
    Amor, N.
    Zaenker, P. P.
    Bluemler, P.
    Meise, F. M.
    Schreiber, L. M.
    Scholz, A.
    Schmiedeskamp, J.
    Spiess, H. W.
    Muennemann, K.
    JOURNAL OF MAGNETIC RESONANCE, 2009, 201 (01) : 93 - 99
  • [50] Correlation of Lung Clearance Index with Hyperpolarized 129Xe Magnetic Resonance Imaging in Pediatric Subjects with Cystic Fibrosis
    Kanhere, Nikhil
    Couch, Marcus J.
    Kowalik, Krzysztof
    Zanette, Brandon
    Rayment, Jonathan H.
    Manson, David
    Subbarao, Padma
    Ratjen, Felix
    Santyr, Giles
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2017, 196 (08) : 1073 - 1075