Automated hyperpolarized 129 Xe gas generator for nuclear magnetic resonance spectroscopy and imaging applications

被引:1
|
作者
Hattori, Mineyuki [1 ]
Hiraga, Takashi [1 ]
Homma, Kazuhiro [1 ]
Ohtake, Norio [2 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Res Inst Mat & Chem Measurement, Natl Metrol Inst Japan NMIJ, Tsukuba, Ibaraki 3058565, Japan
[2] Toyoko Kagaku Co Ltd, Resaerch & Dev Div, Kawasaki, Kanagawa 2118502, Japan
来源
关键词
Hyperpolarized; Xenon; Nuclear magnetic resonance; Magnetic resonance imaging; SPIN-EXCHANGE; XE-129; GAS; NMR; XENON; MRI; POLARIZATION; BEHAVIOR; TIME; FLOW;
D O I
10.1016/j.jmro.2024.100154
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We describe a method and an apparatus for producing hyperpolarized (HP) Xe gas of high concentration without Xe condensation. The HP Xe generator works under atmospheric pressure and employs a quasi-continuous method to provide a continuous supply of HP Xe gas by adopting technologies for supplying a highly pure gas, a gas control system and precise pressure control. The apparatus has a glass cell containing solid Rb metal and solid Xe in vacuum at a low temperature and is heated so that the Xe becomes a gas and Rb exists in a gas- liquid mixture. Then a magnetic field is applied with laser beam irradiation to produce polarized Xe gas of high concentration. The device was evaluated using a 2T magnetic resonance imaging (MRI) scanner. Long-term experimental operation demonstrated the continuous collection of 30 ml syringes of HP Xe gas with a sufficient polarization rate for fast one scan acquisition. A practical device for the automated manufacture of HP 129 Xe gas was thus successfully developed.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Hyperpolarized 129Xe Magnetic Resonance Imaging Measures of Gas Exchange in Non-Specific Interstitial Pneumonia
    Mummy, D.
    Wang, Z.
    Bier, E.
    Tighe, R. M.
    Driehuys, B.
    Mammarappallil, J.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [22] MR imaging and spectroscopy using hyperpolarized Xe-129 gas: Preliminary human results
    Mugler, JP
    Driehuys, B
    Brookeman, JR
    Cates, GD
    Berr, SS
    Bryant, RG
    Daniel, TM
    deLange, EE
    Downs, JH
    Erickson, CJ
    Happer, W
    Hinton, DP
    Kassel, NF
    Maier, T
    Phillips, CD
    Saam, BT
    Sauer, KL
    Wagshul, ME
    MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (06) : 809 - 815
  • [23] Diffusion Pore Imaging by Hyperpolarized Xenon-129 Nuclear Magnetic Resonance
    Kuder, Tristan Anselm
    Bachert, Peter
    Windschuh, Johannes
    Laun, Frederik Bernd
    PHYSICAL REVIEW LETTERS, 2013, 111 (02)
  • [24] Hyperpolarized 129Xe Magnetic Resonance Imaging Sensor for H2S
    Yang, Shengjun
    Yuan, Yaping
    Jiang, Weiping
    Ren, Lili
    Deng, He
    Bouchard, Louis S.
    Zhou, Xin
    Liu, Maili
    CHEMISTRY-A EUROPEAN JOURNAL, 2017, 23 (32) : 7648 - 7652
  • [25] In vivo MR imaging and spectroscopy using hyperpolarized Xe-129
    Wagshul, ME
    Button, TM
    Li, HFF
    Liang, ZR
    Springer, CS
    Zhong, K
    Wishnia, A
    MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (02) : 183 - 191
  • [26] Hyperpolarized 129Xe nuclear magnetic resonance studies of isoreticular metal-organic frameworks
    Pawsey, Shane
    Moudrakovski, Igor
    Ripmeester, John
    Wang, Li-Qiong
    Exarhos, Gregory J.
    Rowsell, Jesse L. C.
    Yaghi, Omar M.
    JOURNAL OF PHYSICAL CHEMISTRY C, 2007, 111 (16): : 6060 - 6067
  • [27] Reproducibility Of Hyperpolarized 129xe Mri: Alveolar Capillary Gas-Transfer Spectroscopy And Imaging
    Kaushik, S. S.
    Liljeroth, M.
    Virgincar, R. S.
    Robertson, S. H.
    Davies, J.
    Stiles, J.
    Kelly, K. T.
    Foster, W. M.
    Morrison, L.
    McAdams, H. P.
    Driehuys, B.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2013, 187
  • [28] Regional ventilation mapping of the rat lung using hyperpolarized 129Xe magnetic resonance imaging
    Couch, Marcus J.
    Ouriadov, Alexei
    Santyr, Giles E.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (05) : 1623 - 1631
  • [29] Hyperpolarized 129Xe Magnetic Resonance Imaging Is Sensitive to Therapy Response in Idiopathic Pulmonary Fibrosis
    Mummy, D.
    Rankine, L.
    Wang, Z.
    Bier, E.
    Rackley, C.
    Morrison, L. D.
    Mammarappallil, J.
    Driehuys, B.
    Tighe, R. M.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [30] Optimized production of hyperpolarized 129Xe at 2 bars for in vivo lung magnetic resonance imaging
    Norquay, Graham
    Parnell, Steven R.
    Xu, Xiaojun
    Parra-Robles, Juan
    Wild, Jim M.
    JOURNAL OF APPLIED PHYSICS, 2013, 113 (04)