Performance assessment of the surface-guided radiation therapy system: Varian Identify

被引:0
|
作者
Ayrancioglu, Oguzhan [1 ]
Ayrancioglu, Ceren [2 ]
Arikan, Serife Ceren [1 ]
Alicikus, Lutfiye Zumre Arican [1 ]
机构
[1] Izmir Tinaztepe Univ, Galen Hosp, Dept Radiat Oncol, Izmir, Turkiye
[2] Izmir Tinaztepe Univ, Vocat Sch, Izmir, Turkiye
关键词
Surface-guided radiation therapy; Performance assessment; Offsets; Setup accuracy; CONCOMITANT EXPOSURES; TARGET VOLUME; HEAD; VERIFICATION; ERRORS; IMRT; RISK;
D O I
10.1016/j.meddos.2024.01.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different com puted tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1 degrees. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0 degrees gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose. (c) 2024 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:222 / 228
页数:7
相关论文
共 50 条
  • [31] Real-time liver tracking algorithm based on LSTM and SVR networks for use in surface-guided radiation therapy
    Wang, Guangyu
    Li, Zhibin
    Li, Guangjun
    Dai, Guyu
    Xiao, Qing
    Bai, Long
    He, Yisong
    Liu, Yaxin
    Bai, Sen
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [32] Intrafraction Motion in Surface-Guided Breast Radiation Therapy and its Implications on a Single Planning Target Volume Margin Strategy
    Malone, Ciaran
    Ryan, Samantha
    Nicholson, Jill
    Mcardle, Orla
    Brennan, Sinead
    Mccavana, Pat
    Mcclean, Brendan
    Duane, Frances
    PRACTICAL RADIATION ONCOLOGY, 2025, 15 (01) : e63 - e71
  • [33] Can surface-guided radiation therapy (SGRT) be used to eliminate tattoos and reduce setup time for SBRT lung treatments?
    Day, J.
    Moreira, A.
    Konig, B.
    Andratschke, N.
    Guckenberger, M.
    Zamburlini, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (06) : 603 - 603
  • [34] Evaluation of initial patient setup methods for breast cancer between surface-guided radiation therapy and laser alignment based on skin marking in the Halcyon system
    Seonghee Kang
    Hyeongmin Jin
    Ji Hyun Chang
    Bum‑Sup Jang
    Kyung Hwan Shin
    Chang Heon Choi
    Jung-in Kim
    Radiation Oncology, 18
  • [35] Dosimetric Analysis of Intra-Fraction Motion Detected by Surface-Guided Radiation Therapy During Linac Stereotactic Radiosurgery
    Foster, Ryan D.
    Moeller, Benjamin J.
    Robinson, Myra
    Bright, Megan
    Ruiz, Justin L.
    Hampton, Carnell J.
    Heinzerling, John H.
    ADVANCES IN RADIATION ONCOLOGY, 2023, 8 (03)
  • [36] Evaluation of initial patient setup methods for breast cancer between surface-guided radiation therapy and laser alignment based on skin marking in the Halcyon system
    Kang, Seonghee
    Jin, Hyeongmin
    Chang, Ji Hyun
    Jang, Bum-Sup
    Shin, Kyung Hwan
    Choi, Chang Heon
    Kim, Jung-in
    RADIATION ONCOLOGY, 2023, 18 (01)
  • [37] Assessment of Surface Guided Radiation therapy in Breast cancer: An Indian Experience
    Barthwal, M.
    Pareek, V.
    Yadav, J.
    Ms, S.
    Bisht, R. K.
    Patil, P.
    Sharma, A.
    Sharma, D. N.
    Sharma, S., V
    Mallick, S.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S874 - S874
  • [38] Implementation of Linac-based VMAT total body irradiation technique on Elekta platform using surface-guided radiation therapy
    Sandt, Magali
    Marcet, Stephanie
    Guesnel, Nathalie
    Claude, Line
    Martel, Isabelle
    Biston, Marie-Claude
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2025, 131
  • [39] A Smartphone-Based Surface Guided Radiation Therapy Tracking System for Respiratory Monitoring in Radiation Therapy
    Capaldi, D.
    Axente, M.
    Yu, A.
    Skinner, L.
    Prionas, N.
    Hirata, E.
    Nano, T.
    MEDICAL PHYSICS, 2022, 49 (06) : E333 - E334
  • [40] Using Varian Eclipse as An Independent MU Check for MRI-Guided Radiation Therapy
    Yang, J.
    Zhang, P.
    Tyagi, N.
    Scripes, P. Godoy
    Mechalakos, J.
    Li, A.
    Lovelock, D.
    Lim, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E874 - E874