Performance assessment of the surface-guided radiation therapy system: Varian Identify

被引:0
|
作者
Ayrancioglu, Oguzhan [1 ]
Ayrancioglu, Ceren [2 ]
Arikan, Serife Ceren [1 ]
Alicikus, Lutfiye Zumre Arican [1 ]
机构
[1] Izmir Tinaztepe Univ, Galen Hosp, Dept Radiat Oncol, Izmir, Turkiye
[2] Izmir Tinaztepe Univ, Vocat Sch, Izmir, Turkiye
关键词
Surface-guided radiation therapy; Performance assessment; Offsets; Setup accuracy; CONCOMITANT EXPOSURES; TARGET VOLUME; HEAD; VERIFICATION; ERRORS; IMRT; RISK;
D O I
10.1016/j.meddos.2024.01.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different com puted tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1 degrees. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0 degrees gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose. (c) 2024 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:222 / 228
页数:7
相关论文
共 50 条
  • [1] Surface-Guided Radiation Therapy Error Assessment: Impact of Surface Discrepancies in Breast Cancer Patients
    Tehrani, J. Nasehi
    MEDICAL PHYSICS, 2024, 51 (09) : 6639 - 6640
  • [2] The Role of Surface-Guided Radiation Therapy to Improve Patient Safety
    Viscariello, N.
    Al-Hallaq, H.
    Batista, V.
    Kugele, M.
    Ford, E.
    Meyer, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [3] Review of clinical applications and challenges with surface-guided radiation therapy
    Bellala, Ravishankar
    Kuppusamy, Anandakrishnan
    Bellala, Venkata Madhavi
    Tyagi, Tulika
    Manoharan, Surendhiran
    Gangarapu, Gunasekhar
    Bellala, Rishik
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 (05) : 1160 - 1169
  • [4] The role of surface-guided radiation therapy for improving patient safety
    Al-Hallaq, Hania
    Batista, Vania
    Kugele, Malin
    Ford, Eric
    Viscariello, Natalie
    Meyer, Juergen
    RADIOTHERAPY AND ONCOLOGY, 2021, 163 : 229 - 236
  • [5] Surface-guided radiation therapy for breast cancer: more precise positioning
    A. González-Sanchis
    L. Brualla-González
    C. Fuster-Diana
    J. C. Gordo-Partearroyo
    T. Piñeiro-Vidal
    T. García-Hernandez
    J. L. López-Torrecilla
    Clinical and Translational Oncology, 2021, 23 : 2120 - 2126
  • [6] Commissioning of a 3D optical imaging system for surface-guided breast radiation therapy
    Song, Yulin
    Hsu, Hao-Yun
    Mueller, Boris A.
    Braunstein, Lior Z.
    TRANSLATIONAL BIOPHOTONICS: DIAGNOSTICS AND THERAPEUTICS, 2021, 11919
  • [7] Impact of Deformations and Posture Variations On Surface-Guided Radiation Therapy of the Breast
    Meyer, J.
    Wootton, L.
    MEDICAL PHYSICS, 2019, 46 (06) : E422 - E423
  • [8] Surface-guided radiation therapy for breast cancer: more precise positioning
    Gonzalez-Sanchis, A.
    Brualla-Gonzalez, L.
    Fuster-Diana, C.
    Gordo-Partearroyo, J. C.
    Pineiro-Vidal, T.
    Garcia-Hernandez, T.
    Lopez-Torrecilla, J. L.
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2021, 23 (10): : 2120 - 2126
  • [9] Initial experience with Surface-guided radiation therapy (SGRT) in SRS treatments
    Sanchez-Rubio, P.
    Rodriguez-Romero, R.
    Pinto-Monedero, M.
    Martinez-Ortega, J.
    Torres-Lopez, M.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1473 - S1474
  • [10] Quality Assurance Program for Surface-guided Radiation Therapy: A Review of Guidelines
    Biltekin, Fatih
    Ozyigit, Gokhan
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2023, 38 (02): : 246 - 253