Dynamic anthropomorphic thorax phantom for quality assurance of motion management in radiotherapy

被引:1
|
作者
Abdollahi, Sara [1 ,2 ,9 ]
Mowlavi, Ali Asghar [3 ,8 ]
Yazdi, Mohammad Hadi Hadizadeh [1 ]
Ceberg, Sofie [4 ]
Aznar, Marianne Camille [5 ]
Tabrizi, Fatemeh Varshoee [6 ]
Salek, Roham [7 ]
Guckenberger, Matthias [2 ]
Tanadini-Lang, Stephanie [2 ]
机构
[1] Ferdowsi Univ Mashhad, Fac Sci, Dept Phys, Mashhad, Iran
[2] Univ Hosp Zurich, Dept Radiat Oncol, CH-8091 Zurich, Switzerland
[3] Hakim Sabzevari Univ, Dept Phys, Sabzevar, Iran
[4] Lund Univ, Dept Med Radiat Phys, Lund, Sweden
[5] Univ Manchester, Fac Biol Med & Hlth, Div Canc Sci, Manchester, England
[6] Reza Radiotherapy & Oncol Ctr, Dept Radiat Oncol, Mashhad, Iran
[7] Mashhad Univ Med Sci, Dept Radiat Oncol, Mashhad, Iran
[8] Univ Montreal, Dept Phys, Montreal, PQ, Canada
[9] Univ Hosp Zurich, CH-8091 Zurich, Switzerland
关键词
Lung SBRT; Dynamic anthropomorphic phantom; End-to-end test; LUNG-TUMOR MOTION; MITIGATION; DOSIMETRY;
D O I
10.1016/j.phro.2024.100587
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Motion management techniques are important to spare the healthy tissue adequately. However, they are complex and need dedicated quality assurance. The aim of this study was to create a dynamic phantom designed for quality assurance and to replicate a patient's size, anatomy, and tissue density. Materials and methods: A computed tomography (CT) scan of a cancer patient was used to create molds for the lungs, heart, ribs, and vertebral column via additive manufacturing. A pump system and software were developed to simulate respiratory dynamics. The extent of respiratory motion was quantified using a 4DCT scan. End- to-end tests were conducted to evaluate two motion management techniques for lung stereotactic body radiotherapy (SBRT). Results: The chest wall moved between 4 mm and 13 mm anteriorly and 2 mm to 7 mm laterally during the breathing. The diaphragm exhibited superior-inferior movement ranging from 5 mm to 16 mm in the left lung and 10 mm to 36 mm in the right lung. The left lung tumor displaced f 7 mm superior-inferiorly and anterior- posteriorly. The CT numbers were for lung:-716 f 108 HU (phantom) and-713 f 70 HU (patient); bone: 460 f 20 HU (phantom) and 458 f 206 HU (patient); soft tissue: 92 f 9 HU (phantom) and 60 f 25 HU (patient). The end-to-end testing showed an excellent agreement between the measured and the calculated dose for ion chamber and film dosimetry. Conclusions: The phantom is recommended for quality assurance, evaluating the institution's specific planning and motion management strategies either through end-to-end testing or as an external audit phantom.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] A moving liver phantom in an anthropomorphic thorax for SPECT MP imaging
    S. Panagi
    Α. Hadjiconstanti
    G. Charitou
    D. Kaolis
    I. Petrou
    C. Kyriacou
    Y. Parpottas
    Physical and Engineering Sciences in Medicine, 2022, 45 : 63 - 72
  • [32] Design and CT imaging of casper, an anthropomorphic breathing thorax phantom
    Laidlaw, Josie
    Earl, Nicolas
    Shavdia, Nihal
    Davis, Rayna
    Mayer, Sarah
    Karaman, Dmitri
    Richtsmeier, Devon
    Rodesch, Pierre-Antoine
    Bazalova-Carter, Magdalena
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (02):
  • [33] A deformable phantom for quality assurance in 4D radiotherapy
    Margeanu, M.
    Heath, E.
    Stroian, G.
    Seuntjens, J.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S80 - S81
  • [34] The Radiological Physics Center's anthropomorphic quality assurance phantom family: New developments
    Followill, D
    Lowenstein, J
    Jhingran, A
    Roll, J
    Hernandez, N
    Ibbott, G
    MEDICAL PHYSICS, 2005, 32 (06) : 2129 - 2129
  • [35] Quality assurance evaluation of spot scanning beam proton therapy with an anthropomorphic prostate phantom
    Iqbal, K.
    Gillin, M.
    Summers, P. A.
    Dhanesar, S.
    Gifford, K. A.
    Buzdar, S. A.
    BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1031):
  • [36] DESIGN, FABRICATION AND IMPLEMENTATION OF PELVIC ANTHROPOMORPHIC PHANTOM FOR QUALITY ASSURANCE BY USING GEL DOSIMETRY
    Nedaie, H. A.
    Vaezzadeh, S. A.
    Sadoughi, H.
    Allahverdi, M.
    Aghili, M.
    Esfahani, M.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S509 - S509
  • [37] A proposal for a standard electronic anthropomorphic phantom for radiotherapy
    Aldridge, JS
    Reckwerdt, PJ
    Mackie, TR
    MEDICAL PHYSICS, 1999, 26 (09) : 1901 - 1903
  • [38] An anthropomorphic breathing phantom of the thorax for testing new motion mitigation techniques for pencil beam scanning proton therapy
    Perrin, R. L.
    Zakova, M.
    Peroni, M.
    Bernatowicz, K.
    Bikis, C.
    Knopf, A. K.
    Safai, S.
    Fernandez-Carmona, P.
    Tscharner, N.
    Weber, D. C.
    Parkel, T. C.
    Lomax, A. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (06): : 2486 - 2504
  • [39] Development of an anthropomorphic lung phantom for imaging and radiotherapy
    Weidner, A.
    Runz, A.
    Johnen, W.
    Echner, G.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S475 - S475
  • [40] Design, development, and implementation of the Radiological Physics Center's pelvis and thorax anthropomorphic quality assurance phantoms
    Followill, David S.
    Evans, DeeAnn Radford
    Cherry, Christopher
    Molineu, Andrea
    Fisher, Gary
    Hanson, William F.
    Lbbott, Geoffrey S.
    MEDICAL PHYSICS, 2007, 34 (06) : 2070 - 2076