Dynamic anthropomorphic thorax phantom for quality assurance of motion management in radiotherapy

被引:1
|
作者
Abdollahi, Sara [1 ,2 ,9 ]
Mowlavi, Ali Asghar [3 ,8 ]
Yazdi, Mohammad Hadi Hadizadeh [1 ]
Ceberg, Sofie [4 ]
Aznar, Marianne Camille [5 ]
Tabrizi, Fatemeh Varshoee [6 ]
Salek, Roham [7 ]
Guckenberger, Matthias [2 ]
Tanadini-Lang, Stephanie [2 ]
机构
[1] Ferdowsi Univ Mashhad, Fac Sci, Dept Phys, Mashhad, Iran
[2] Univ Hosp Zurich, Dept Radiat Oncol, CH-8091 Zurich, Switzerland
[3] Hakim Sabzevari Univ, Dept Phys, Sabzevar, Iran
[4] Lund Univ, Dept Med Radiat Phys, Lund, Sweden
[5] Univ Manchester, Fac Biol Med & Hlth, Div Canc Sci, Manchester, England
[6] Reza Radiotherapy & Oncol Ctr, Dept Radiat Oncol, Mashhad, Iran
[7] Mashhad Univ Med Sci, Dept Radiat Oncol, Mashhad, Iran
[8] Univ Montreal, Dept Phys, Montreal, PQ, Canada
[9] Univ Hosp Zurich, CH-8091 Zurich, Switzerland
关键词
Lung SBRT; Dynamic anthropomorphic phantom; End-to-end test; LUNG-TUMOR MOTION; MITIGATION; DOSIMETRY;
D O I
10.1016/j.phro.2024.100587
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Motion management techniques are important to spare the healthy tissue adequately. However, they are complex and need dedicated quality assurance. The aim of this study was to create a dynamic phantom designed for quality assurance and to replicate a patient's size, anatomy, and tissue density. Materials and methods: A computed tomography (CT) scan of a cancer patient was used to create molds for the lungs, heart, ribs, and vertebral column via additive manufacturing. A pump system and software were developed to simulate respiratory dynamics. The extent of respiratory motion was quantified using a 4DCT scan. End- to-end tests were conducted to evaluate two motion management techniques for lung stereotactic body radiotherapy (SBRT). Results: The chest wall moved between 4 mm and 13 mm anteriorly and 2 mm to 7 mm laterally during the breathing. The diaphragm exhibited superior-inferior movement ranging from 5 mm to 16 mm in the left lung and 10 mm to 36 mm in the right lung. The left lung tumor displaced f 7 mm superior-inferiorly and anterior- posteriorly. The CT numbers were for lung:-716 f 108 HU (phantom) and-713 f 70 HU (patient); bone: 460 f 20 HU (phantom) and 458 f 206 HU (patient); soft tissue: 92 f 9 HU (phantom) and 60 f 25 HU (patient). The end-to-end testing showed an excellent agreement between the measured and the calculated dose for ion chamber and film dosimetry. Conclusions: The phantom is recommended for quality assurance, evaluating the institution's specific planning and motion management strategies either through end-to-end testing or as an external audit phantom.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] dynamic anthropomorphic thorax phantom for quality assurance of motion management in radiotherapy
    Abdollahi, Sara
    Mowlavi, Ali Asghar
    Yazdi, Mohammad Hadi Hadizadeh
    Ceberg, Sofie
    Aznar, Marianne Camille
    Tabrizi, Fatemeh Varshoee
    Salek, Roham
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3281 - S3285
  • [2] A programmable motion phantom for quality assurance of motion management in radiotherapy
    L. Dunn
    T. Kron
    P. N. Johnston
    L. N. McDermott
    M. L. Taylor
    J. Callahan
    R. D. Franich
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 93 - 100
  • [3] A programmable motion phantom for quality assurance of motion management in radiotherapy
    Dunn, L.
    Kron, T.
    Johnston, P. N.
    McDermott, L. N.
    Taylor, M. L.
    Callahan, J.
    Franich, R. D.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 93 - 100
  • [4] Fabrication of a Patient-Derived Anthropomorphic Thorax Phantom for End-to-End Quality Assurance for Online Adaptive Radiotherapy
    Stanley, D. N.
    Harms, J.
    Pogue, J. A.
    Cardan, R. A.
    Viscariello, N. N.
    Popple, R. A.
    Cardenas, C. E.
    MEDICAL PHYSICS, 2024, 51 (10) : 7716 - 7716
  • [5] The Design and Fabrication of a Dynamic Anthropomorphic Thorax Phantom
    Xu, Z.
    Sun, Z.
    Shi, X.
    Lin, C.
    Lu, W.
    Lu, W.
    Qiu, J.
    Shi, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E556 - E557
  • [6] ADAM a Breathing Anthropomorphic Phantom for Quality Assurance of Respiratory Motion-Management Techniques in Radiation Therapy
    Pallotta, S.
    Calusi, S.
    Foggi, L.
    Lisci, R.
    Marrazzo, L.
    Masi, L.
    Talamonti, C.
    Livi, L.
    Simontacchi, G.
    MEDICAL PHYSICS, 2017, 44 (06) : 3223 - 3224
  • [7] An anthropomorphic 3D printed inhomogeneity thorax phantom slab for SBRT commissioning and quality assurance
    How, Stephen
    Banjade, Dilli
    Crowe, Scott
    Dillon, Greg
    Skimmings, Andrew
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 575 - 583
  • [8] An anthropomorphic 3D printed inhomogeneity thorax phantom slab for SBRT commissioning and quality assurance
    Stephen How
    Dilli Banjade
    Scott Crowe
    Greg Dillon
    Andrew Skimmings
    Physical and Engineering Sciences in Medicine, 2023, 46 : 575 - 583
  • [9] A novel anthropomorphic multimodality phantom for MRI-based radiotherapy quality assurance testing
    Singhrao, Kamal
    Fu, Jie
    Wu, Holden H.
    Hu, Peng
    Kishan, Amar U.
    Chin, Robert K.
    Lewis, John H.
    MEDICAL PHYSICS, 2020, 47 (04) : 1443 - 1451
  • [10] An anthropomorphic phantom for quality assurance and training in gynaecological brachytherapy
    de Almeida, CE
    Rodriguez, M
    Vianello, E
    Ferreira, IH
    Sibata, C
    RADIOTHERAPY AND ONCOLOGY, 2002, 63 (01) : 75 - 81