Quantitative calcium-based assessment of osteoporosis in dual-layer spectral CT

被引:2
|
作者
Do, T. D. [1 ]
Rahn, S. [1 ]
Melzig, C. [1 ]
Heussel, C. P. [2 ]
Stiller, W. [1 ]
Kauczor, H. U. [1 ]
Weber, T. F. [1 ]
Skornitzke, S. [1 ]
机构
[1] Heidelberg Univ Hosp, Clin Diag & Intervent Radiol DIR, Im Neuenheimer Feld 130 3, D-69120 Heidelberg, Germany
[2] Heidelberg Univ, Thoraxklin, Diag & Intervent Radiol Nucl Med, Heidelberg, Germany
关键词
Computed tomography; Osteoporosis; Calcium imaging; Dual-layer detector; BONE-MINERAL DENSITY; X-RAY ABSORPTIOMETRY; COMPUTED-TOMOGRAPHY; POSTMENOPAUSAL WOMEN; ENERGY; ACCURACY; CONTRAST;
D O I
10.1016/j.ejrad.2024.111606
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To evaluate a novel calcium-only imaging technique (VCa) with subtracted bone marrow in osteoporosis in dual-layer CT (DLCT) compared to conventional CT images (CI) and dual-energy X-ray absorptiometry (DXA). Material and methods: Images of a multi-energy CT phantom with calcium inserts, quantitative CT calibration phantom, and of 55 patients (mean age: 64.6 +/- 11.5 years) were acquired on a DLCT to evaluate bone mineral density (BMD). CI, calcium-suppressed images, and VCa were calculated. For investigating the association of VCa and CI with DXA a subsample of 30 patients (<90 days between DXA and CT) was used. Multiple regression analysis was performed to identify further factors improving the prediction of DXA BMD. Results: The calcium concentrations of the CT phantom inserts were significantly associated with CT numbers from VCa (R-2 = 0.94) and from CI (R-2 = 0.89-0.92). VCa showed significantly higher CT numbers than CI in the phantom (p <= 0.001) and clinical setting (p < 0.001). CT numbers from VCa were significantly associated with CI (R-2 = 0.95, p < 0.001) and with DXA (R-2 = 0.31, p = 0.007), whereas no significant association between DXA and CI was found. Prediction of DXA BMD based on CT numbers derived from VCa yielded R-2 = 0.76 in multiple regression analysis. ROC for the differentiation of normal from pathologic BMD in VCa yielded an AUC of 0.7, and a cut-off value of 126HU (sensitivity: 0.90; specificity: 0.47). Conclusion: VCa images showed better agreement with DXA and known calcium concentrations than CI, and could be used to estimate BMD. A VCa cut-off of 126 HU could be used to identify abnormal bone mineral density.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Objective image characterization of a spectral CT scanner with dual-layer detector
    Ozguner, Orhan
    Dhanantwari, Amar
    Halliburton, Sandra
    Wen, Gezheng
    Utrup, Steven
    Jordan, David
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02):
  • [22] The Application of Dual-layer Spectral Detector CT in Abdominal Vascular Imaging
    Lai, Lu-Yao
    Jiang, Ying
    Shu, Jian
    CURRENT MEDICAL IMAGING, 2023, 19 (14) : 1609 - 1615
  • [23] Improvement of Breast Cancer Detection Using Dual-Layer Spectral CT
    Hasse, Felix Christian
    Giannakis, Athanasios
    Wehrse, Eckhard
    Stiller, Wolfram
    Wallwiener, Markus
    Kauczor, Hans-Ulrich
    Weber, Tim F.
    Heil, Jorg
    Mokry, Theresa
    DIAGNOSTICS, 2024, 14 (14)
  • [24] SNR performance comparison of dual-layer detector and dual-kVp spectral CT
    Heismann, B. J.
    Wirth, S.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3820 - 3822
  • [25] Effect of CT Acquisition Parameters on Iodine Density Measurement at Dual-Layer Spectral CT
    Kim, Hyungjin
    Park, Chang Min
    Kang, Chang Kyung
    Yoon, Jinyeong
    Chae, Kum Ju
    Goo, Jin Mo
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 211 (04) : 748 - 754
  • [26] Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system
    Ehn, Sebastian
    Sellerer, Thorsten
    Muenzel, Daniela
    Fingerle, Alexander A.
    Kopp, Felix
    Duda, Manuela
    Mei, Kai
    Renger, Bernhard
    Herzen, Julia
    Dangelmaier, Julia
    Schwaiger, Benedikt J.
    Sauter, Andreas
    Riederer, Isabelle
    Renz, Martin
    Braren, Rickmer
    Rummeny, Ernst J.
    Pfeiffer, Franz
    Noeel, Peter B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 204 - 217
  • [27] Phantom-based quantification of the spectral accuracy in dual-layer spectral CT for pediatric imaging at 100 kVp
    Meyer, Sebastian
    Liu, Leening P.
    Litt, Harold, I
    Halliburton, Sandra S.
    Shapira, Nadav
    Noel, Peter B.
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (02) : 924 - 934
  • [29] Early rectal neoplasm in dual-layer spectral detector computed tomography: dual-layer spectral computed tomography (CT) images improve tumor detection and staging
    Pan, Xuelin
    Wu, Zhihan
    Zhao, Jin
    Zhang, Xinyi
    Zhang, Xiaodi
    Tang, Li
    Yang, Jinlin
    Deng, Kai
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (12) : 8260 - 8271
  • [30] Material separation with dual-layer CT
    Carmi, Raz
    Naveh, Galit
    Altman, Ami
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1876 - 1878