The influence of miscentering on radiation dose during computed tomography head examinations and the role of localiser orientation: A phantom study

被引:0
|
作者
Ravenscroft, L. [1 ]
Baker, L. [2 ]
机构
[1] Univ Hosp Derby & Burton NHS Fdn Trust, Derby, Staffs, England
[2] Nottingham Univ Hosp NHS Trust, Res & Dev Imaging Support Unit, Nottingham, England
关键词
Computed tomography; Miscentering; Radiation dose; Brain; Radiographic positioning; TUBE CURRENT MODULATION; IMAGE NOISE; PATIENT; MDCT; POSTEROANTERIOR; EXPOSURE; SYSTEMS; IMPACT;
D O I
10.1016/j.radi.2024.09.051
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Computed Tomography (CT) chest, abdomen and pelvis research demonstrates a relationship between vertical phantom positioning and radiation dose. Moving the phantom closer or further from the x-ray source results in magnification or minimisation of the localiser. As automatic tube current modulation (ATCM) algorithms use localisers to estimate patient size and calculate required tube current, magnification or minimisation results in the incorrect provision of radiation dose. Radiation dose changes also depend on localiser orientation, changes with anteroposterior (AP) and posteroanterior (PA) localisers demonstrating an inverse relationship. However, within CT head literature often attributes radiation dose changes on impaired function of the bow-tie filter instead. The current study investigated the role of miscentering on ATCM function within CT head, paying particular attention to localiser orientation.<br /> Methods: Head scanning was performed with an anthropomorphic phantom at the isocentre, alongside ten vertically miscentered positions. This was performed three times, with an AP, PA and lateral localiser. CT dose index values at each miscentered level were compared across conditions. Results: Vertical miscentering altered radiation dose in both AP and PA conditions, radiation dose linearly increasing (up to 17.05%) when positioning the phantom closer to the x-ray source and decreasing when positioning away (up to -13.13%). Changes across AP and PA conditions demonstrated an inverse relationship. Radiation dose was unaffected in the lateral condition.<br /> Conclusions: Miscentering during CT head alters ATCM function due to magnification/minimisation of the localiser image, causing ATCM algorithms to misinterpret patient size and miscalculate required tube current.<br /> Implications for practice: Radiographers should be accurate when centering for CT head, avoiding any potential radiation dose changes. Further research into vertical miscentering and image quality during CT head is recommended. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those
引用
收藏
页码:1517 / 1523
页数:7
相关论文
共 50 条
  • [31] Comparison of selected photon shield and organ-based tube current modulation for radiation dose reduction in head computed tomography: A phantom study
    Yuan, Zilong
    Liu, Tao
    Zhang, Biao
    Wu, Jiaxing
    He, Yaoyao
    Chen, Tiao
    Zhang, Zhaoxi
    Li, Cuiling
    Liu, Yulin
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2023, 31 (04) : 745 - 756
  • [32] Fetal radiation dose of four tube voltages in abdominal CT examinations during pregnancy: A phantom study
    Matsunaga, Yuta
    Haba, Tomonobu
    Kobayashi, Masanao
    Suzuki, Shoichi
    Asada, Yasuki
    Chida, Koichi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (02): : 178 - 184
  • [33] Comparison of estimated and calculated fetal radiation dose for a pregnant woman who underwent computed tomography and conventional X-ray examinations based on a phantom study
    Saeed, M. K.
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (01) : 25 - 33
  • [34] Comparison of estimated and calculated fetal radiation dose for a pregnant woman who underwent computed tomography and conventional X-ray examinations based on a phantom study
    M. K. Saeed
    Radiological Physics and Technology, 2021, 14 : 25 - 33
  • [35] Evaluation of Radiation Dose in Coronary Computed Tomography-Angiography Examinations: Using Organ dose and Effective dose
    Shibata, Hideki
    Kondo, Yuya
    Fukada, Shinji
    Asada, Yasuki
    Kozawa, Isao
    RADIATION PHYSICS AND CHEMISTRY, 2019, 158 : 218 - 221
  • [36] Regional survey of image quality and radiation dose in computed tomography examinations in Saudi Arabia
    M. K. Saeed
    A. S. Alzoubi
    J. M. Al-Qahtani
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 279 - 283
  • [37] Regional survey of image quality and radiation dose in computed tomography examinations in Saudi Arabia
    Saeed, M. K.
    Alzoubi, A. S.
    Al-Qahtani, J. M.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 279 - 283
  • [38] Radiation doses in cerebral perfusion computed tomography: patient and phantom study
    Arandjic, D.
    Bonutti, F.
    Biasizzo, E.
    Ciraj-Bjelac, O.
    Floreani, M.
    Giustizieri, M.
    Iaiza, F.
    Inkoom, S.
    Tommasini, G.
    Padovani, R.
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (04) : 459 - 464
  • [39] Determination of radiation dose during computed tomography examinations in southern Saudi Arabian hospitals using size-specific dose estimates
    Saeed, Mohammed K.
    Asiri, Abdullah Ali M.
    MAEJO INTERNATIONAL JOURNAL OF SCIENCE AND TECHNOLOGY, 2021, 15 (02) : 147 - 158
  • [40] Radiation dose reduction in pediatric great vessel stent computed tomography using iterative reconstruction: A phantom study
    den Harder, Annemarie M.
    Sucha, Dominika
    van Doormaal, Pieter J.
    Budde, Ricardo P. J.
    de Jong, Pim A.
    Schilham, Arnold M. R.
    Breur, Johannes M. P. J.
    Leiner, Tim
    PLOS ONE, 2017, 12 (04):