The influence of miscentering on radiation dose during computed tomography head examinations and the role of localiser orientation: A phantom study

被引:0
|
作者
Ravenscroft, L. [1 ]
Baker, L. [2 ]
机构
[1] Univ Hosp Derby & Burton NHS Fdn Trust, Derby, Staffs, England
[2] Nottingham Univ Hosp NHS Trust, Res & Dev Imaging Support Unit, Nottingham, England
关键词
Computed tomography; Miscentering; Radiation dose; Brain; Radiographic positioning; TUBE CURRENT MODULATION; IMAGE NOISE; PATIENT; MDCT; POSTEROANTERIOR; EXPOSURE; SYSTEMS; IMPACT;
D O I
10.1016/j.radi.2024.09.051
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Computed Tomography (CT) chest, abdomen and pelvis research demonstrates a relationship between vertical phantom positioning and radiation dose. Moving the phantom closer or further from the x-ray source results in magnification or minimisation of the localiser. As automatic tube current modulation (ATCM) algorithms use localisers to estimate patient size and calculate required tube current, magnification or minimisation results in the incorrect provision of radiation dose. Radiation dose changes also depend on localiser orientation, changes with anteroposterior (AP) and posteroanterior (PA) localisers demonstrating an inverse relationship. However, within CT head literature often attributes radiation dose changes on impaired function of the bow-tie filter instead. The current study investigated the role of miscentering on ATCM function within CT head, paying particular attention to localiser orientation.<br /> Methods: Head scanning was performed with an anthropomorphic phantom at the isocentre, alongside ten vertically miscentered positions. This was performed three times, with an AP, PA and lateral localiser. CT dose index values at each miscentered level were compared across conditions. Results: Vertical miscentering altered radiation dose in both AP and PA conditions, radiation dose linearly increasing (up to 17.05%) when positioning the phantom closer to the x-ray source and decreasing when positioning away (up to -13.13%). Changes across AP and PA conditions demonstrated an inverse relationship. Radiation dose was unaffected in the lateral condition.<br /> Conclusions: Miscentering during CT head alters ATCM function due to magnification/minimisation of the localiser image, causing ATCM algorithms to misinterpret patient size and miscalculate required tube current.<br /> Implications for practice: Radiographers should be accurate when centering for CT head, avoiding any potential radiation dose changes. Further research into vertical miscentering and image quality during CT head is recommended. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those
引用
收藏
页码:1517 / 1523
页数:7
相关论文
共 50 条
  • [21] Changing Radiation Dose From Diagnostic Computed Tomography Examinations in Saskatchewan
    Dumaine, Chance S.
    Leswick, David A.
    Fladeland, Derek A.
    Lim, Hyun J.
    Toews, Lori J.
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2012, 63 (03): : 183 - 191
  • [22] Development of a computed tomography phantom to evaluate radiation dose and image quality
    Gomez Cardona, Daniel
    Estrada Leon, Edgar Santiago
    Zuluaga Santamaria, Alejandro
    Benitez, Luis
    Ramirez Giraldo, Juan Carlos
    2013 PAN AMERICAN HEALTH CARE EXCHANGES (PAHCE), 2013,
  • [23] How Localizer Orders Impact Image Quality and Radiation Dose in Computed Tomography - A Phantom Study
    Jimenez, J.
    Rong, J.
    Jensen, C.
    Jones, A.
    Liu, X.
    MEDICAL PHYSICS, 2020, 47 (06) : E538 - E538
  • [24] Effective radiation dose and radiogenic cancer risk during contrast enhanced abdominal computed tomography examinations
    Sulieman, A.
    Tamam, N.
    Elnour, Abdelrahman
    Alkhorayef, M.
    Babikir, E.
    Al-Mohammed, Huda I.
    Abuhadi, N.
    Khandaker, Mayeen Uddin
    Bradley, David A.
    Radiation Physics and Chemistry, 2022, 200
  • [25] Radiation Dose Reduction in Head and Neck Computed Tomography Angiography
    Zhao, Ying-Jia
    Ma, Xi
    Zheng, Ya-Ting
    Zuo, Zi-Wei
    Xu, Ying-Jin
    Wang, Jia-Ning
    Yin, Xiao-Ping
    Gao, Bu-Lang
    CURRENT MEDICAL IMAGING, 2022, 18 (07) : 757 - 763
  • [26] Effective radiation dose and radiogenic cancer risk during contrast enhanced abdominal computed tomography examinations
    Sulieman, A.
    Tamam, N.
    Elnour, Abdelrahman
    Alkhorayef, M.
    Babikir, E.
    Al-Mohammed, Huda I.
    Abuhadi, N.
    Khandaker, Mayeen Uddin
    Bradley, David A.
    RADIATION PHYSICS AND CHEMISTRY, 2022, 200
  • [27] Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates
    Abuzaid, Mohamed M.
    Elshami, Wiam
    Tekin, Huseyin Ozan
    Ghonim, Hatem
    Shawki, Mona
    Salama, Dina H.
    INSIGHTS INTO IMAGING, 2020, 11 (01)
  • [28] Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates
    Mohamed M. Abuzaid
    Wiam Elshami
    Huseyin Ozan Tekin
    Hatem Ghonim
    Mona Shawki
    Dina H. Salama
    Insights into Imaging, 11
  • [29] Low radiation dose in computed tomography: the role of iodine
    Aschoff, Andrik J.
    Catalano, Carlo
    Kirchin, Miles A.
    Krix, Martin
    Albrecht, Thomas
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1076):
  • [30] Radiation doses to neonates during X ray computed tomography examinations
    Akahane, K
    Hashimoto, M
    Hada, M
    Takano, Y
    Ono, K
    Kai, M
    Kusama, T
    RADIATION PROTECTION DOSIMETRY, 2003, 103 (01) : 41 - 45