The influence of miscentering on radiation dose during computed tomography head examinations and the role of localiser orientation: A phantom study

被引:0
|
作者
Ravenscroft, L. [1 ]
Baker, L. [2 ]
机构
[1] Univ Hosp Derby & Burton NHS Fdn Trust, Derby, Staffs, England
[2] Nottingham Univ Hosp NHS Trust, Res & Dev Imaging Support Unit, Nottingham, England
关键词
Computed tomography; Miscentering; Radiation dose; Brain; Radiographic positioning; TUBE CURRENT MODULATION; IMAGE NOISE; PATIENT; MDCT; POSTEROANTERIOR; EXPOSURE; SYSTEMS; IMPACT;
D O I
10.1016/j.radi.2024.09.051
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Computed Tomography (CT) chest, abdomen and pelvis research demonstrates a relationship between vertical phantom positioning and radiation dose. Moving the phantom closer or further from the x-ray source results in magnification or minimisation of the localiser. As automatic tube current modulation (ATCM) algorithms use localisers to estimate patient size and calculate required tube current, magnification or minimisation results in the incorrect provision of radiation dose. Radiation dose changes also depend on localiser orientation, changes with anteroposterior (AP) and posteroanterior (PA) localisers demonstrating an inverse relationship. However, within CT head literature often attributes radiation dose changes on impaired function of the bow-tie filter instead. The current study investigated the role of miscentering on ATCM function within CT head, paying particular attention to localiser orientation.<br /> Methods: Head scanning was performed with an anthropomorphic phantom at the isocentre, alongside ten vertically miscentered positions. This was performed three times, with an AP, PA and lateral localiser. CT dose index values at each miscentered level were compared across conditions. Results: Vertical miscentering altered radiation dose in both AP and PA conditions, radiation dose linearly increasing (up to 17.05%) when positioning the phantom closer to the x-ray source and decreasing when positioning away (up to -13.13%). Changes across AP and PA conditions demonstrated an inverse relationship. Radiation dose was unaffected in the lateral condition.<br /> Conclusions: Miscentering during CT head alters ATCM function due to magnification/minimisation of the localiser image, causing ATCM algorithms to misinterpret patient size and miscalculate required tube current.<br /> Implications for practice: Radiographers should be accurate when centering for CT head, avoiding any potential radiation dose changes. Further research into vertical miscentering and image quality during CT head is recommended. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those
引用
收藏
页码:1517 / 1523
页数:7
相关论文
共 50 条
  • [1] Radiation dose optimization for computed tomography of the head in pediatric population - An experimental phantom study
    Priyanka
    Kadavigere, R.
    Sukumar, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (04): : 747 - 751
  • [2] Thyroid, Parathyroid and Eye Dose Evaluation in Head and Neck Computed Tomography Examinations, Phantom and Clinical Study
    Asgari, A.
    Parach, A. A.
    Sharafi, A. A.
    Nazarparvar, B.
    Parvizi, S.
    WEST INDIAN MEDICAL JOURNAL, 2016, 65 (03): : 538 - 542
  • [3] Increased Computed Tomography Dose Due to Miscentering With Use of Automated Tube Voltage Selection: Phantom and Patient Study
    Filev, Peter D.
    Mittal, Pardeep K.
    Tang, Xiangyang
    Phuong-Anh Duong
    Wang, Xiaojing
    Small, William C.
    Applegate, Kimberly
    Moreno, Courtney C.
    CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2016, 45 (04) : 265 - 270
  • [4] Analysis of Radiation Dose to Pediatric Patients During Computed Tomography Examinations
    Kharbanda, Anupam B.
    Krause, Ernest
    Lu, Yi
    Blumberg, Karen
    ACADEMIC EMERGENCY MEDICINE, 2015, 22 (06) : 670 - 675
  • [5] Radiation dose of chaperones during common pediatric computed tomography examinations
    Overhoff, Daniel
    Weis, Meike
    Riffel, Philipp
    Sudarski, Sonja
    Froelich, Matthias F.
    Fries, Peter
    Schoenberg, Stefan
    Gawlitza, Joshua
    PEDIATRIC RADIOLOGY, 2020, 50 (08) : 1078 - 1082
  • [6] Radiation dose of chaperones during common pediatric computed tomography examinations
    Daniel Overhoff
    Meike Weis
    Philipp Riffel
    Sonja Sudarski
    Matthias F. Froelich
    Peter Fries
    Stefan Schönberg
    Joshua Gawlitza
    Pediatric Radiology, 2020, 50 : 1078 - 1082
  • [7] Fetal dose conversion factor for fetal computed tomography examinations: A mathematical phantom study
    Matsunaga, Yuta
    Kawaguchi, Ai
    Kobayashi, Masanao
    Suzuki, Shoichi
    Asada, Yasuki
    Ito, Kiyoshi
    Chida, Koichi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05): : 330 - 335
  • [8] Effect of miscentering and low-dose protocols on contrast resolution in computed tomography head examination
    Rahmat, Said Mohd Shaffiq Said
    Karim, Muhammad Khalis Abdul
    Isa, Iza Nurzawani Che
    Abd Rahman, Mohd Amiruddin
    Noor, Noramaliza Mohd
    Hoong, Ng Kwan
    COMPUTERS IN BIOLOGY AND MEDICINE, 2020, 123
  • [9] International variation in radiation dose for computed tomography examinations: prospective cohort study
    Smith-Bindman, Rebecca
    Wang, Yifei
    Chu, Philip
    Chung, Robert
    Einstein, Andrew J.
    Balcombe, Jonathan
    Cocker, Mary
    Das, Marcos
    Delman, Bradley N.
    Flynn, Michael
    Gould, Robert
    Lee, Ryan K.
    Yellen-Nelson, Thomas
    Schindera, Sebastian
    Seibert, Anthony
    Starkey, Jay
    Suntharalingam, Saravanabavaan
    Wetter, Axel
    Wildberger, Joachim E.
    Miglioretti, Diana L.
    BMJ-BRITISH MEDICAL JOURNAL, 2019, 364
  • [10] Influence of tube potential on quantitative coronary plaque analyses by low radiation dose computed tomography: a phantom study
    Chunhong Wang
    Yuliang Liao
    Haibin Chen
    Xin Zhen
    Jianhong Li
    Yikai Xu
    Linghong Zhou
    The International Journal of Cardiovascular Imaging, 2018, 34 : 1315 - 1322