Prospective head motion correction at 3 Tesla with wireless NMR markers and ultrashort echo navigators

被引:0
|
作者
Sengupta, Saikat [1 ,2 ]
Glenn, Antonio [3 ]
Rogers, Baxter P. [1 ,2 ]
机构
[1] Vanderbilt Univ, Inst Imaging Sci, Med Ctr, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Med Ctr, Dept Radiol & Radiol Sci, Nashville, TN 37235 USA
[3] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
关键词
Prospective motion correction; Wireless NMR markers; Navigator; FAT IMAGE NAVIGATORS; MRI; RESOLUTION;
D O I
10.1016/j.mri.2024.110238
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Prospective motion correction (PMC) with inductively-coupled wireless NMR markers has been shown to be an effective plug-and-play method for dealing with head motion at 7 Tesla [29,30]. However, technical challenges such as one-to-one identification of three wireless markers, generation of hyper-intense marker artifacts and low marker peak SNR in the navigators has limited the adoption of this technique. The goal of this work is to introduce solutions to overcome these issues and extend this technique to PMC for brain imaging at 3 Tesla. Methods: PMC with 6 degrees of freedom (DOF) was implemented using a novel similar to 8 ms, ultrashort echo time (UTE) navigator in concert with optimally chosen MnCl2 marker samples to minimize marker artifacts. Distinct head coil sensitivities were leveraged to enable identification and tracking of individual markers and a variable flip angle (VFA) scheme and real time filtering were used to boost marker SNR. PMC was performed in 3D T-1 weighted brain imaging at 3 Tesla with voluntary head motions in adult volunteers. Results: PMC with wireless markers improved image quality in 3D T-1 weighted images in all subjects compared to non-motion corrected images for similar motions with no noticeable marker artifacts. Precision of motion tracking was found to be in the range of 0.01-0.06 mm/degrees. Navigator execution had minimal impact on sequence duration. Conclusions: Wireless NMR markers provide an accurate, calibration-free and economical option for 6 DOF PMC in brain imaging across field strengths. Challenges in this technique can be addressed by combining navigator design, sample selection and real time data processing strategies.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Comparison of Turbo Spin Echo and Echo Planar Imaging for intravoxel incoherent motion and diffusion tensor imaging of the kidney at 3 Tesla
    Hilbert, Fabian
    Wech, Tobias
    Neubauer, Henning
    Veldhoen, Simon
    Bley, Thorsten Alexander
    Koestler, Herbert
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (03): : 193 - 201
  • [22] Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T
    Serger, Matthias
    Stirnberg, Ruediger
    Ehses, Philipp
    Stoecker, Tony
    MAGNETIC RESONANCE IN MEDICINE, 2025,
  • [23] Optical tracking with two markers for robust prospective motion correction for brain imaging
    Aditya Singh
    Benjamin Zahneisen
    Brian Keating
    Michael Herbst
    Linda Chang
    Maxim Zaitsev
    Thomas Ernst
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2015, 28 : 523 - 534
  • [24] Optical tracking with two markers for robust prospective motion correction for brain imaging
    Singh, Aditya
    Zahneisen, Benjamin
    Keating, Brian
    Herbst, Michael
    Chang, Linda
    Zaitsev, Maxim
    Ernst, Thomas
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2015, 28 (06) : 523 - 534
  • [25] Self-encoded Marker for Optical Prospective Head Motion Correction in MRI
    Forman, Christoph
    Aksoy, Murat
    Hornegger, Joachim
    Bammer, Roland
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2010, PT I, 2010, 6361 : 259 - +
  • [26] Self-encoded marker for optical prospective head motion correction in MRI
    Forman, Christoph
    Aksoy, Murat
    Hornegger, Joachim
    Bammer, Roland
    MEDICAL IMAGE ANALYSIS, 2011, 15 (05) : 708 - 719
  • [27] Prospective motion correction of fMRI: Improving the quality of resting state data affected by large head motion
    Maziero, Danilo
    Rondinoni, Carlo
    Marins, Theo
    Stenger, Victor Andrew
    Ernst, Thomas
    NEUROIMAGE, 2020, 212
  • [28] Prospective motion correction of 3D echo-planar imaging data for functional MRI using optical tracking
    Todd, Nick
    Josephs, Oliver
    Callaghan, Martina F.
    Lutti, Antoine
    Weiskopf, Nikolaus
    NEUROIMAGE, 2015, 113 : 1 - 12
  • [29] Prospective motion correction in 2D multishot MRI using EPI navigators and multislice-to-volume image registration
    Hoinkiss, Daniel Christopher
    Porter, David Andrew
    MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (06) : 2127 - 2135
  • [30] Prospective Motion Correction and Selective Reacquisition Using Volumetric Navigators for Vessel-Encoded Arterial Spin Labeling Dynamic Angiography
    Frost, Robert
    Hess, Aaron T.
    Okell, Thomas W.
    Chappell, Michael A.
    Tisdall, M. Dylan
    van der Kouwe, Andre J. W.
    Jezzard, Peter
    MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (05) : 1420 - 1430