Completely noninvasive multi-analyte monitoring system for cell culture processes

被引:1
|
作者
Rahmatnejad, Vida [1 ]
Tolosa, Michael [1 ]
Ge, Xudong [1 ]
Rao, Govind [1 ]
机构
[1] Univ Maryland Baltimore Cty, Ctr Adv Sensor Technol, Dept Chem Biochem & Environm Engn, Baltimore, MD 21250 USA
关键词
Dissolved carbon dioxide; Dissolved oxygen; Noninvasive; Sensor; Mammalian cell culture; Flow cell; DISSOLVED-OXYGEN; PH; DIFFUSION; CHIP; O-2;
D O I
10.1007/s10529-024-03521-z
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Although online monitoring of dissolved O2, pH, and dissolved CO2 is critical in bioprocesses, nearly all existing technologies require some level of direct contact with the cell culture environment, posing risks of contamination. This study addresses the need for an accurate, and completely noninvasive technique for simultaneous measurement of these analytes. A "non-contact" technique for simultaneous monitoring of dissolved O2, pH, and dissolved CO2 was developed. Instead of direct contact with the culture media, the measurements were made through permeable membranes via either a sampling port in the culture vessel wall or a flow cell. The efficacy of the "non-contact" technique was validated in Escherichia coli (E.coli), Chinese hamster ovary (CHO) culture processes, and dynamic environments created by sparging gases in cell culture medium. The measurements obtained through the developed techniques were comparable to those obtained through control methods. The noninvasive monitoring system can offer accurate, and contamination-minimized monitoring of critical process parameters including dissolved O2, pH, and dissolved CO2. These advancements will enhance the control and optimization of cell culture processes, promising improved cell culture performance.
引用
收藏
页码:983 / 996
页数:14
相关论文
共 50 条
  • [11] Line-Monitoring, Hyperspectral Fluorescence Setup for Simultaneous Multi-Analyte Biosensing
    Liu, Zhiyi
    Shi, Heng
    Liu, Le
    Deng, Sunan
    Ji, Yanhong
    Ma, Suihua
    Ma, Hui
    He, Yonghong
    SENSORS, 2011, 11 (11) : 10038 - 10047
  • [12] Microdevice with integrated dialysis probe and biosensor array for continuous multi-analyte monitoring
    Petrou, PS
    Moser, I
    Jobst, G
    BIOSENSORS & BIOELECTRONICS, 2003, 18 (5-6): : 613 - 619
  • [13] Multi-analyte analysis system using an antibody-based biochip
    Maria Moreno-Bondi
    Jean Alarie
    Tuan Vo-Dinh
    Analytical and Bioanalytical Chemistry, 2003, 375 : 120 - 124
  • [14] Multi-analyte analysis system using an antibody-based biochip
    Moreno-Bondi, MC
    Alarie, JP
    Vo-Dinh, T
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 375 (01) : 120 - 124
  • [15] FABRICATION PROCESS OF INTEGRATED MULTI-ANALYTE BIOCHIP SYSTEM FOR IMPLANTABLE APPLICATION
    Tsai, Y. -C.
    Chiu, N. -F.
    Liu, P. -C.
    Ou, Y-C
    Liao, H. -H.
    Yang, Y-J.
    Yang, L. -J.
    Lei, U.
    Chao, F. -S.
    Lu, S. -S.
    Lin, C. -W.
    Chang, P. -Z
    Shih, W. -P.
    IEEE 22ND INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS 2009), 2009, : 204 - 207
  • [16] Development and calibration of a multi-analyte piezo-optical dosimeter system for environmental monitoring of nitrogen dioxide and ozone
    Schillinger, EFJ
    Wright, JD
    SENSORS AND ACTUATORS B-CHEMICAL, 2004, 98 (2-3): : 262 - 268
  • [17] Advanced multi-modal, multi-analyte optochemical sensing platform for cell analysis
    V. Zhdanov, Alexander
    Li, Liang
    Yang, Panpan
    Shkirdova, Alena O.
    Tang, Shuze
    V. Yashunsky, Dmitri
    V. Ponomarev, Gelii
    Zamilatskov, Ilya A.
    Papkovsky, Dmitri B.
    SENSORS AND ACTUATORS B-CHEMICAL, 2022, 355
  • [18] A noninvasive multi-analyte diagnostic assay: combining protein and DNA markers to stratify bladder cancer patients
    Fernandez, Cecilia A.
    Millholland, John M.
    Zwarthoff, Ellen C.
    Feldman, Adam S.
    Karnes, R. Jeffrey
    Shuber, Anthony P.
    RESEARCH AND REPORTS IN UROLOGY, 2012, 4 : 17 - 26
  • [19] Development of a novel, multi-analyte biosensor system for assaying cell division: Identification of cell proliferation/death precursor events
    Kintzios, S
    Marinopoulou, I
    Moschopoulou, G
    Manganab, O
    Nomikou, K
    Endo, K
    Papanastasiou, I
    Simonian, A
    BIOSENSORS & BIOELECTRONICS, 2006, 21 (07): : 1365 - 1373
  • [20] Clinical Implementation of a Noninvasive, Multi-Analyte Droplet Digital PCR Test to Screen for Androgen Receptor Alterations
    Stitz, Regina
    Stoiber, Franz
    Silye, Rene
    Vlachos, Georgios
    Andaloro, Silvia
    Rebhan, Elisabeth
    Dunzinger, Michael
    Puehringer, Franz
    Gallo, Caroline
    El-Heliebi, Amin
    Heitzer, Ellen
    Hauser-Kronberger, Cornelia
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2024, 26 (06): : 467 - 478