Completely noninvasive multi-analyte monitoring system for cell culture processes

被引:1
|
作者
Rahmatnejad, Vida [1 ]
Tolosa, Michael [1 ]
Ge, Xudong [1 ]
Rao, Govind [1 ]
机构
[1] Univ Maryland Baltimore Cty, Ctr Adv Sensor Technol, Dept Chem Biochem & Environm Engn, Baltimore, MD 21250 USA
关键词
Dissolved carbon dioxide; Dissolved oxygen; Noninvasive; Sensor; Mammalian cell culture; Flow cell; DISSOLVED-OXYGEN; PH; DIFFUSION; CHIP; O-2;
D O I
10.1007/s10529-024-03521-z
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Although online monitoring of dissolved O2, pH, and dissolved CO2 is critical in bioprocesses, nearly all existing technologies require some level of direct contact with the cell culture environment, posing risks of contamination. This study addresses the need for an accurate, and completely noninvasive technique for simultaneous measurement of these analytes. A "non-contact" technique for simultaneous monitoring of dissolved O2, pH, and dissolved CO2 was developed. Instead of direct contact with the culture media, the measurements were made through permeable membranes via either a sampling port in the culture vessel wall or a flow cell. The efficacy of the "non-contact" technique was validated in Escherichia coli (E.coli), Chinese hamster ovary (CHO) culture processes, and dynamic environments created by sparging gases in cell culture medium. The measurements obtained through the developed techniques were comparable to those obtained through control methods. The noninvasive monitoring system can offer accurate, and contamination-minimized monitoring of critical process parameters including dissolved O2, pH, and dissolved CO2. These advancements will enhance the control and optimization of cell culture processes, promising improved cell culture performance.
引用
收藏
页码:983 / 996
页数:14
相关论文
共 50 条
  • [1] A noninvasive multi-analyte assay for triaging patients
    Annette Fenner
    Nature Reviews Urology, 2012, 9 (4) : 176 - 176
  • [2] A multi-analyte assay for the noninvasive detection of bladder cancer
    Miyake, Makito
    Goodison, Steve
    Chang, Myron
    Dai, Yunfeng
    Urquidi, Virginia
    Rosser, Charles Joel
    JOURNAL OF CLINICAL ONCOLOGY, 2013, 31 (06)
  • [3] Evaluation of the Paratrend Multi-Analyte Sensor for Potential Utilization in Long-Duration Automated Cell Culture Monitoring
    Emma Y. Hwang
    Dimitri Pappas
    Antony S. Jeevarajan
    Melody M. Anderson
    Biomedical Microdevices, 2004, 6 : 241 - 249
  • [4] Evaluation of the paratrend multi-analyte sensor for potential utilization in long-duration automated cell culture monitoring
    Hwang, EY
    Pappas, D
    Jeevarajan, AS
    Anderson, MM
    BIOMEDICAL MICRODEVICES, 2004, 6 (03) : 241 - 249
  • [5] Microfluidic organ-on-chip system for multi-analyte monitoring of metabolites in 3D cell cultures
    Dornhof, Johannes
    Kieninger, Jochen
    Muralidharan, Harshini
    Maurer, Jochen
    Urban, Gerald A.
    Weltin, Andreas
    LAB ON A CHIP, 2022, 22 (02) : 225 - 239
  • [6] The application of multiple reaction monitoring and multi-analyte profiling to HDL proteins
    Hussein N Yassine
    Angela M Jackson
    Chad R Borges
    Dean Billheimer
    Hyunwook Koh
    Derek Smith
    Peter Reaven
    Serrine S Lau
    Christoph H Borchers
    Lipids in Health and Disease, 13
  • [7] The application of multiple reaction monitoring and multi-analyte profiling to HDL proteins
    Yassine, Hussein N.
    Jackson, Angela M.
    Borges, Chad R.
    Billheimer, Dean
    Koh, Hyunwook
    Smith, Derek
    Reaven, Peter
    Lau, Serrine S.
    Borchers, Christoph H.
    LIPIDS IN HEALTH AND DISEASE, 2014, 13
  • [8] Continuous Metabolic Monitoring Based on Multi-Analyte Biomarkers to Predict Exhaustion
    Michail Kastellorizios
    Diane J. Burgess
    Scientific Reports, 5
  • [9] Continuous Metabolic Monitoring Based on Multi-Analyte Biomarkers to Predict Exhaustion
    Kastellorizios, Michail
    Burgess, Diane J.
    SCIENTIFIC REPORTS, 2015, 5
  • [10] Prototype of micro flow cell for multi-analyte sandwich immunoassay
    Honda, N
    Shoji, S
    Isomura, M
    Ashihara, Y
    Akahori, K
    Sato, H