An 8-Channel Transceiver Coil for Carotid Artery Imaging at 7T Using an Optimized Shield Design

被引:0
|
作者
Yazdanbakhsh, Pedram [1 ,2 ]
Couch, Marcus [1 ,2 ,3 ]
Gilbert, Kyle M. [4 ]
Der Hovagimian, Johnny [1 ,5 ,6 ]
Rudko, David A. [1 ,2 ,5 ]
Hoge, Richard D. [1 ,2 ,5 ]
机构
[1] McGill Univ, Montreal Neurol Inst & Hosp, McConnell Brain Imaging Ctr, Montreal, PQ H3A 0G4, Canada
[2] McGill Univ, Dept Neurol & Neurosurg, Montreal, PQ H3A 0G4, Canada
[3] Siemens Healthcare Ltd, Ottawa, ON, Canada
[4] Western Univ, Ctr Funct & Metab Mapping, London, ON, Canada
[5] McGill Univ, Dept Biomed Engn, Montreal, PQ, Canada
[6] AEM Antennas Inc, Montreal, PQ, Canada
基金
加拿大健康研究院;
关键词
Coils; Radio frequency; Switches; Capacitors; Carotid arteries; Preamplifiers; Phase shifters; 7; tesla; carotid imaging; high-field MRI; parallel transmission; phased array RF coil; ARRAYS;
D O I
10.1109/TBME.2024.3379980
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: To design and fabricate a transmit/receive (T/R) radiofrequency (RF) coil array for MRI of the carotid arteries at 7T with optimal shielding to improve transmit performance in parallel transmit (pTx) mode. Methods: The carotid coil included 8 total RF elements, with left and right subarrays, each consisting of 4 overlapping loops with RF shields. Electromagnetic (EM) simulations were performed to optimize and improve the transmit performance of the array by determining the optimal distance between the RF shield and each subarray. EM simulations were further used to calculate local specific absorption rate (SAR) matrices. Based on the SAR matrices, virtual observation points (VOPs) were applied to ensure safety during parallel transmission. The efficacy of the coil design was evaluated by measuring coil performance metrics when imaging a phantom and by acquiring in-vivo images. Results: The optimal distance between the RF shield and each subarray was determined to be 45 mm. This resulted in a maximum B-1(+) efficiency of 1.23 mu T/$\surd W$ in the carotid arteries and a peak, 10-g-average SAR per Watt of 0.86 kg(-1) when transmitting in the nominal CP+ mode. Optimizing the RF shield resulted in up to 37% improvement in B-1(+) efficiency and 14% improvement in SAR efficiency compared to an unshielded design. Conclusion and significance: Optimizing the distance between the RF shield and coil array provided significant improvement in the transmit characteristics of the bilateral carotid coil. The bilateral coil topology provides a compelling platform for imaging the carotid arteries with high field MRI.
引用
收藏
页码:2537 / 2544
页数:8
相关论文
共 50 条
  • [21] From Complex B1 Mapping to Local SAR Estimation for Human Brain MR Imaging Using Multi-Channel Transceiver Coil at 7T
    Zhang, Xiaotong
    Schmitter, Sebastian
    Van de Moortele, Pierre-Francois
    Liu, Jiaen
    He, Bin
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (06) : 1058 - 1067
  • [22] A 32-channel receive array coil for bilateral breast imaging and spectroscopy at 7T
    Del Bosque, Romina
    Cui, Jiaming
    Ogier, Stephen
    Cheshkov, Sergey
    Dimitrov, Ivan E.
    Malloy, Craig
    Wright, Steven M.
    McDougall, Mary
    MAGNETIC RESONANCE IN MEDICINE, 2021, 85 (01) : 551 - 559
  • [23] Parallel imaging performance investigation of an 8-channel common-mode differential-mode (CMDM) planar array for 7T MRI
    Hu, Xiaoqing
    Chen, Xiao
    Liu, Xin
    Zheng, Hairong
    Li, Ye
    Zhang, Xiaoliang
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2014, 4 (01) : 33 - 42
  • [24] Sodium imaging of the heart at 7T: design, evaluation and application of a four-channel transmit/receive surface coil array
    A Graessl
    A Ruehle
    W Renz
    L Winter
    H Pfeiffer
    J Ruff
    J Rieger
    T Niendorf
    Journal of Cardiovascular Magnetic Resonance, 15 (Suppl 1)
  • [25] Design and Simulation of a Birdcage Coil using CST Studio Suite for Application at 7T
    Tomas, Bernat Palau
    Li, Houmin
    Anjum, M. R.
    1ST INTERNATIONAL CONFERENCE ON SENSING FOR INDUSTRY, CONTROL, COMMUNICATIONS, & SECURITY TECHNOLOGIES (ICSICCST 2013), 2013, 51
  • [26] A flexible five-channel shielded-coaxial-cable (SCC) transceive neck coil for high-resolution carotid imaging at 7T
    Ruytenberg, Thomas
    Webb, Andrew
    Zivkovic, Irena
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (03) : 1672 - 1677
  • [27] Optimization of an 8-Channel Loop-Array Coil for a 7 T MRI System with the Guidance of a Co-Simulation Approach
    Yan, Xinqiang
    Ma, ChuangXin
    Shi, Lei
    Zhuo, Yan
    Zhou, Xiaohong Joe
    Wei, Long
    Xue, Rong
    APPLIED MAGNETIC RESONANCE, 2014, 45 (05) : 437 - 449
  • [28] Optimization of an 8-Channel Loop-Array Coil for a 7 T MRI System with the Guidance of a Co-Simulation Approach
    Xinqiang Yan
    ChuangXin Ma
    Lei Shi
    Yan Zhuo
    Xiaohong Joe Zhou
    Long Wei
    Rong Xue
    Applied Magnetic Resonance, 2014, 45 : 437 - 449
  • [29] Nineteen-Channel Receive Array and Four-Channel Transmit Array Coil for Cervical Spinal Cord Imaging at 7T
    Zhao, Wei
    Cohen-Adad, Julien
    Polimeni, Jonathan R.
    Keil, Boris
    Guerin, Bastien
    Setsompop, Kawin
    Serano, Peter
    Mareyam, Azma
    Hoecht, Philipp
    Wald, Lawrence L.
    MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (01) : 291 - 300
  • [30] High-resolution volumetric dynamic magnetic resonance imaging of the wrist using an 8-channel flexible receive coil
    Zi, Ruoxun
    Abbas, Batool
    Wang, Bili
    Walczyk, Jerzy
    Brown, Ryan
    Petchprapa, Catherine
    Fishbaugh, James
    Gerig, Guido
    Block, Kai Tobias
    Lattanzi, Riccardo
    SKELETAL RADIOLOGY, 2024, : 1291 - 1299