Start-to-end modeling and transmission efficiency optimization for a cyclotron-based proton therapy beamline

被引:2
|
作者
Chen, Yu [1 ]
Qin, Bin [1 ]
Liu, Xu [1 ]
Wang, Wei [1 ]
Liao, Yicheng [1 ]
机构
[1] Huazhong Univ Sci & Technol, Sch Elect & Elect Engn, State Key Lab Adv Electromagnet Technol, Wuhan 430074, Peoples R China
基金
中国国家自然科学基金;
关键词
Start-to-end modeling; Monte Carlo simulation; Transmission efficiency optimization; Multi-objective Bayesian optimization; SIMULATION;
D O I
10.1016/j.net.2024.05.041
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Utilizing first-order beam dynamics models is adequate for studying the beam properties during the conceptual design of a cyclotron-based proton therapy beamline. After finishing lattice design, particle-matter interaction simulations for passive elements (e.g., degrader, collimators, energy slit) are required. The cascade simulation is used for lattice updates in each iteration, which is complicated. In addition, when the models involve particle tracking and particle-matter interaction, their optimization process is time-consuming. Therefore, this study proposes a start-to-end modeling method using Monte Carlo Beam Delivery Simulation (BDSIM) software that considers more realistic factors, such as particle-matter interaction and the realistic vacuum chamber, to precisely evaluate working parameters, along with an efficient optimization method that utilizes multi-objective Bayesian optimization (MOBO) to improve transmission efficiency. Taking the Huazhong University of Science and Technology proton therapy facility (HUST-PTF) as an example, beam loss along the beamline is located, quantified, and subsequently reduced by tuning the quadrupole strengths based on MOBO. The results show that: (i) By considering the particle-matter interaction and the realistic vacuum chamber, the precision in the prediction of the beam properties is improved; (ii) After optimization, the transmission efficiency of the entire beamline is relatively increased by an average of 6.52 % under different energy settings, especially 11.39 % at 70 MeV.
引用
收藏
页码:4365 / 4374
页数:10
相关论文
共 16 条
  • [1] A data-efficient surrogate modeling method for a cyclotron-based proton therapy beamline based on active learning
    Chen, Yu
    Qin, Bin
    Liu, Xu
    Wang, Wei
    Liao, Yicheng
    INTERNATIONAL JOURNAL OF MODERN PHYSICS A, 2025, 40 (01):
  • [2] Increase of the transmission and emittance acceptance through a cyclotron-based proton therapy gantry
    Maradia, Vivek
    Giovannelli, Anna Chiara
    Meer, David
    Weber, Damien Charles
    Lomax, Antony John
    Schippers, Jacobus Maarten
    Psoroulas, Serena
    MEDICAL PHYSICS, 2022, 49 (04) : 2183 - 2192
  • [3] Emulating the Delivery of Sawtooth Proton Arc Therapy Plans on a Cyclotron-Based Proton Beam Therapy System
    Burford-Eyre, Samuel
    Aitkenhead, Adam
    Aylward, Jack D.
    Henthorn, Nicholas T.
    Ingram, Samuel P.
    Mackay, Ranald
    Manger, Samuel
    Merchant, Michael J.
    Sitch, Peter
    Warmenhoven, John-William
    Appleby, Robert B.
    CANCERS, 2024, 16 (19)
  • [4] Optimization of epithermal neutron field for cyclotron-based boron neutron capture therapy
    Yonai, S
    Aoki, T
    Nakamura, T
    Yashima, H
    Baba, M
    Yokohori, H
    Tahara, Y
    RESEARCH AND DEVELOPMENT IN NEUTRON CAPTURE THERAPY, 2002, : 299 - 303
  • [5] The Use of a Mini Ridge Filter with Cyclotron-Based Pencil Beam Scanning Proton Therapy
    O'grady, F.
    Janson, M.
    Rao, A.
    Chawla, A.
    Fan, J.
    Wang, P.
    MEDICAL PHYSICS, 2022, 49 (06) : E344 - E344
  • [6] The use of a mini-ridge filter with cyclotron-based pencil beam scanning proton therapy
    O'Grady, Fionnbarr
    Janson, Martin
    Rao, Avani D.
    Chawla, Ashish K.
    Stephenson, Lisa
    Fan, Jiajin
    Wang, Peng
    MEDICAL PHYSICS, 2023, 50 (04) : 1999 - 2008
  • [7] New cyclotron-based proton therapy facility development to achieve conformal FLASH treatment plans
    Maradia, Vivek
    van de Water, Steven
    Weber, Damien Charles
    Lomax, Antony John
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4636 - S4638
  • [8] A new emittance selection system to maximize beam transmission for low-energy beams in cyclotron-based proton therapy facilities with gantry
    Maradia, Vivek
    Meer, David
    Weber, Damien Charles
    Lomax, Antony John
    Schippers, Jacobus Maarten
    Psoroulas, Serena
    MEDICAL PHYSICS, 2021, 48 (12) : 7613 - 7622
  • [9] The Beam Commissioning of SC Cyclotron-Based Proton Therapy System With Fast Energy Selection and Precise Gantry
    Zhang, Tianjue
    Wang, Chuan
    Yin, Zhiguo
    Jia, Xianlu
    Wang, Yang
    Fu, Wei
    Liu, Jingyuan
    Jin, Zhaojun
    Zhou, Hongji
    Zhang, Suping
    Cai, Hongru
    Pan, Gaofeng
    Zhu, Xiaofeng
    Lin, Jun
    Zhao, Bohan
    Guo, Qiankun
    Liu, Chuanye
    Xiong, Rui
    He, Aolai
    Jiang, Tianyi
    Hu, Mingzhi
    Song, Qiqi
    Ding, Aoxuan
    Jiang, Shangmeng
    Liu, Ping
    Yang, Fangdi
    Huang, Xiesi
    Chu, Zhichao
    Xia, Xiaoxue
    Zhang, Zheyuan
    Qin, Ling
    Ma, Yongjun
    Cheng, Ziyi
    Xu, Zheng
    Le, Hao
    Zhang, Youwen
    Wang, Xi
    Liu, Zhan
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2025, 72 (03) : 438 - 445
  • [10] Commissioning a clinical proton pencil beam scanning beamline for pre-clinical ultra-high dose rate irradiations on a cyclotron-based system
    Saini, Jatinder
    Erickson, Danielle P. Johnson
    Stappen, Francois Vander
    Ruth, Matt
    Cui, Sunan
    Gorman, Vanessa
    Rossomme, Severine
    Cao, Ning
    Ford, Eric C.
    Meyer, Juergen
    Bloch, Charles
    Wong, Tony
    Grassberger, Clemens
    Rengan, Ramesh
    Zeng, Jing
    Schwarz, Marco
    FRONTIERS IN ONCOLOGY, 2024, 14