Validation of the Sysmex XN-V hematology analyzer for feline specimens

被引:2
|
作者
Guerlin, Maud [1 ,2 ]
Granat, Fanny [1 ,3 ]
Grebert, Margot [1 ,2 ]
Braun, Jean-Pierre [4 ]
Geffre, Anne [1 ,2 ]
Bourges-Abella, Nathalie [2 ]
Trumel, Catherine [1 ,2 ]
机构
[1] Univ Toulouse, Dept Sci Clin Anim Co & Sport, ENVT, Toulouse, France
[2] Univ Toulouse, CREFRE, INSERM, ENVT,UPS, Toulouse, France
[3] Univ Toulouse, CRCT, INSERM, UMR 1037,ENVT, Toulouse, France
[4] Anim & Comparat Clin Pathol, Toulouse, France
关键词
blood; cat; comparison study; method validation; platelet clumping; REFERENCE INTERVALS; ADVIA(R) 2120; XT-2000IV; CATS; DOGS; ACCURACY; CANINE; SYSTEM; COUNT;
D O I
10.1111/vcp.13377
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: The Sysmex XN- V is derived from the new Sysmex XN series of human hematology analyzers. The main changes from the previously validated XT- 2000iV analyzer include an optic- fluorescent analysis for platelets and a nucleated red blood cell (NRBC) count. Objective: We aimed to validate the Sysmex XN- V for feline blood following the American College for Veterinary Clinical Pathology and International Council for Standardization in Hematology recommendations. Methods: Feline EDTA blood specimens were analyzed on the Sysmex XN- V to evaluate repeatability, linearity, comparison with the XT- 2000iV analyzer and manual methods, stability, and to verify the previously established Sysmex XT- 2000iV RIs. Results: Repeatability was excellent for most variables. Visually determined linearity was excellent or good for most variables except eosinophils and platelet variables. The correlation between the XN- V and XT- 2000iV analyzers was good (>= 0.82) for all variables except reticulocyte indices. Correlations between the Sysmex XN- V and manual differential counts were good to excellent for most variables, acceptable for neutrophils, and fair for monocytes and NRBC. The previously established Sysmex XT- 2000iV RIs can be used to interpret results from the Sysmex XN- V analyzer for most variables except red cell distribution width and reticulocyte variables. The RI for platelet variables could not be evaluated because of platelet clumps. Changes in the Sysmex XN- V measurements after storage at 4 and 24 degrees C were similar to those described for the Sysmex XT- 2000iV analyzer. Conclusions: The performance of the Sysmex XN- V analyzer was good and compared favorably with the Sysmex XT- 2000iV analyzer.
引用
收藏
页码:294 / 308
页数:15
相关论文
共 50 条
  • [41] EVALUATION OF AUTOMATED HEMATOLOGY ANALYZER DYMIND DH76 COMPARED TO SYSMEX XN 1000 SYSTEM
    Velizarova, Milena
    Yacheva, Teodora
    Genova, Mariana
    Svinarov, Dobrin
    JOURNAL OF MEDICAL BIOCHEMISTRY, 2021, 40 (04) : 367 - 377
  • [42] METHOD COMPARISON OF COBAS M 511 INTEGRATED HEMATOLOGY ANALYZER AND SYSMEX® XN-10 AUTOMATED HEMATOLOGY ANALYZER USING SAMPLES WITH TARGETED MEDICAL CONDITIONS
    Russcher, Henk
    Khartabil, Tania
    Bracco, Dan
    Allen, Tim
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 123 - 124
  • [43] The fully automated Sysmex XN-31 hematology analyzer can detect bloodstream form Trypanosoma brucei
    Khartabil, Tania
    van Schaik, Ron H. N.
    Haanstra, Jurgen R.
    Koelewijn, Rob
    Russcher, Henk
    Hellemond, Jaap J. van
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2024, 108 (04)
  • [44] COLD AGGLUTININ SAMPLE RESULTS ON THE SYSMEX XN-1000 AND COBAS M 511 INTEGRATED HEMATOLOGY ANALYZER
    Khartabil, Tania
    de Rijke, Yolanda
    Russcher, Henk
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 : 85 - 85
  • [45] Diagnostic Values of Immature Granulocytes Detected by the Sysmex XN 9000 Hematology Analyzer in Children with Urinary Tract Infections
    Cetin, Nuran
    Kocaturk, Evin
    Tufan, Asli Kavaz
    Kiraz, Zeynep Kusku
    Alatas, Ozkan
    SAUDI JOURNAL OF KIDNEY DISEASES AND TRANSPLANTATION, 2023, 34 (SUPPL 1) : S133 - S141
  • [46] Evaluation the diagnostic performance of immature granulocyte cells flag on automated hematology analyzer sysmex XN-series
    Chieh-Hsi, W.
    Fang-Yu, W.
    Hsiang-Ling, H.
    CLINICA CHIMICA ACTA, 2022, 530 : S187 - S187
  • [47] Does mixing improve the comparability between hemoglobin measured on the GEM Premier 4000 blood gas analyzer and the Sysmex XN hematology analyzer?
    Ismail, Ola
    Balion, Cynthia
    Hill, Stephen
    Martin, Janet
    CLINICAL BIOCHEMISTRY, 2019, 74 : 97 - 97
  • [48] Application of the Fluorescence Method on Sysmex XN9000 Hematology Analyzer for Correcting Platelet Count in Individuals with Microcytosis
    Deng, Jiankai
    Xie, Shuhua
    Chen, Yaoming
    Ma, Qinghua
    He, Yuting
    Liu, Min
    Wang, Dong
    Yu, Xuegao
    LABORATORY MEDICINE, 2023, 54 (01) : E10 - E13
  • [49] Performance evaluation of the digital cell imaging analyzer DI-60 integrated into the fully automated Sysmex XN hematology analyzer system
    Tabe, Yoko
    Yamamoto, Takemasa
    Maenou, Imiko
    Nakai, Rie
    Idei, Mayumi
    Horii, Takashi
    Miida, Takashi
    Ohsaka, Akimichi
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (02) : 281 - 289
  • [50] EVALUATION OF A NEW METHOD FOR THE ENUMERATION OF NUCLEATED RED BLOOD CELLS ON THE NEW SYSMEX XN™ AUTOMATED HEMATOLOGY ANALYZER
    Rosenthal, Nancy
    Brown, Bonnie
    Connell, Barbara
    Kruger, Julie
    Capper, Mary
    Blaine, Kimberly
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 131 - 131