Strength of titanium-zirconium alloy implants with a conical connection after implantoplasty

被引:2
|
作者
Shah, Sweety D. [1 ]
Zheng, Fengyuan [2 ]
Seghi, Robert R. [3 ]
Lee, Damian J. [1 ]
机构
[1] Ohio State Univ, Coll Dent, Adv Prosthodont Program, Div Restorat & Prosthet Dent, Columbus, OH USA
[2] Univ Minnesota, Sch Dent, Dept Restorat Sci, Adv Educ Prosthodont Program, Minneapolis, MN USA
[3] Ohio State Univ, Coll Dent, Div Restorat & Prosthet Dent, Columbus, OH USA
来源
JOURNAL OF PROSTHETIC DENTISTRY | 2024年 / 132卷 / 03期
关键词
3-YEAR CLINICAL-TRIAL; SHAPED ORAL IMPLANTS; PERI-IMPLANTITIS; SURFACE-ROUGHNESS; MECHANICAL-PROPERTIES; RESECTIVE SURGERY; TACTILE FUNCTION; DENTAL IMPLANTS; IN-VIVO; THERAPY;
D O I
10.1016/j.prosdent.2022.08.015
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem.Peri-implantitis occurs around dental implants, and implantoplasty hasbeen used to address this ongoing disease; however, the changes to the physical properties ofan implant after implantoplasty have not been well documented. Purpose.The purpose of this in vitro study was to determine the effect of implantoplasty onfracture strength and the load required for plastic deformation after cyclic fatigue on dentalimplants. Material and methods.Twenty-six titanium/zirconium (TiZr) alloy implants (Roxolid Bone LevelImplant; 4.1x10 mm) were embedded with 50% thread exposure and divided into 4 groupsbased on whether they had implantoplasty treatment by using different diamond rotaryinstruments and/or cyclic loading at 250 N for 2 million cycles: C0 (control, no cyclic loading), T0(test, no cyclic loading), CM (control, cyclic loading), and TM (test, cyclic loading). Afterimplantoplasty and/or cyclic loading, all implants underwent a load-to-failure test. The maximumfracture strength (FS) and load required for the onset of plastic deformation (PD) were recordedin Newtons. One-way ANOVA and nonparametric comparisons with control by using the Dunnand Wilcoxon method for joint ranking were used for statistical analysis. Results.The mean +/- standard deviation FS for C0, CM, T0, and TM was 1465.2 +/- 86.4 N, 1480.7 +/- 64.1N, 1299.3 +/- 123.8 N, and 1252.1 +/- 85.7 N, respectively. The mean +/- standard deviation load for onsetof PD for C0, CM, T0, and TM was 860.2 +/- 88.1 N, 797.0 +/- 130.5 N, 776.5 +/- 181.8 N, and 631.3 +/- 84.5 N,respectively. The TM group had a significantly lower FS and PD than the C0, CM, and T0 groups(P<.05) Conclusions.Both fracture strength (FS) and the onset of plastic deformation (PD) weresignificantly reduced after a TiZr alloy implant received implantoplasty and cyclic loading. (JProsthet Dent 2024;132:593-9)
引用
收藏
页码:593 / 599
页数:7
相关论文
共 50 条
  • [31] Cell biological responses of osteoblasts on anodized nanotubular surface of a titanium-zirconium alloy
    Sista, Subhash
    Nouri, Alireza
    Li, Yuncang
    Wen, Cuie
    Hodgson, Peter D.
    Pande, Gopal
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2013, 101 (12) : 3416 - 3430
  • [32] A Prospective Case-Control Clinical Study of Titanium-Zirconium Alloy Implants with a Hydrophilic Surface in Patients with Type 2 Diabetes Mellitus
    Cabrera-Dominguez, Jose J.
    Castellanos-Cosano, Lizett
    Torres-Lagares, Daniel
    Machuca-Portillo, Guillermo
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (05) : 1135 - 1144
  • [33] Hydroxyapatite/titania sol-gel coatings on titanium-zirconium alloy for biomedical applications
    Wen, C. E.
    Xu, W.
    Hu, W. Y.
    Hodgson, P. D.
    ACTA BIOMATERIALIA, 2007, 3 (03) : 403 - 410
  • [34] Survival rate of titanium-zirconium narrow diameter dental implants versus commercially pure titanium narrow diameter dental implants: A systematic review
    Iegami, Carolina Mayumi
    Uehara, Priscila Nakasone
    Sesma, Newton
    Pannuti, Claudio Mendes
    Neto, Pedro Tortamano
    Mukai, Marcio Katsuyoshi
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2017, 19 (06) : 1015 - 1022
  • [35] Mechanical Property Assessment of Bone Healing around a Titanium-Zirconium Alloy Dental Implant
    Anchieta, Rodolfo B.
    Baldassarri, Marta
    Guastaldi, Fernando
    Tovar, Nick
    Janal, Malvin N.
    Gottlow, Jan
    Dard, Michel
    Jimbo, Ryo
    Coelho, Paulo G.
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2014, 16 (06) : 913 - 919
  • [36] Influence of Tantalum Addition on the Corrosion Passivation of Titanium-Zirconium Alloy in Simulated Body Fluid
    Sherif, El-Sayed M.
    Bahri, Yassir A. A.
    Alharbi, Hamad F. F.
    Ijaz, Muhammad Farzik
    Alnaser, Ibrahim A. A.
    MATERIALS, 2022, 15 (24)
  • [37] Dose-response effect of chlorhexidine on a multispecies oral biofilm formed on pure titanium and on a titanium-zirconium alloy
    Silva Souza, Joao Gabriel
    Lima, Carolina Veloso
    Costa Oliveira, Barbara Emanoele
    Ricomini-Filho, Antonio Pedro
    Faveri, Marcelo
    Sukotjo, Cortino
    Feres, Magda
    Del Bel Cury, Altair Antoninha
    Ricardo Barao, Valentim Adelino
    BIOFOULING, 2018, 34 (10) : 1175 - 1184
  • [38] Influence of implant location on titanium-zirconium alloy narrow-diameter implants: A 1-year prospective study in smoking and nonsmoking populations
    Xiao, Wenlan
    Chen, Yaqian
    Chu, Chenyu
    Dard, Michel M.
    Man, Yi
    JOURNAL OF PROSTHETIC DENTISTRY, 2022, 128 (02): : 159 - 166
  • [39] Clinical outcomes of titanium-zirconium alloy narrow-diameter implants for single-crown restorations: a systematic review and meta-analysis
    Cao, Rongkai
    Chen, Beibei
    Xu, Hui
    Fan, Zhen
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2023, 61 (06): : 403 - 410
  • [40] EFFECT OF THE BINDER METAL ON THE STRENGTH AND WEAR-RESISTANCE OF PLASMA COATINGS BASED ON TITANIUM AND TITANIUM-ZIRCONIUM CARBONITRIDES
    TOLSTOBROV, AK
    MITROFANOV, BV
    ZASHLYAPIN, MY
    SOVIET POWDER METALLURGY AND METAL CERAMICS, 1992, 31 (11): : 948 - 952