Design and characterisation of a minibeam collimator utilising Monte Carlo simulation and a clinical linear accelerator

被引:0
|
作者
Carver, Antony [1 ]
Baker, Sam [2 ]
Dumbill, Andrew [1 ]
Horton, Steven [1 ]
Green, Stuart [1 ]
机构
[1] Univ Hosp Birmingham NHS Fdn Trust, Birmingham, England
[2] Royal Marsden NHS Fdn Trust, London, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 13期
关键词
minibeam radiotherapy; spatial fractionation; dosimetry; Monte Carlo simulation; RADIATION-THERAPY; BEARING RATS; RADIOTHERAPY; BEAM; CODE;
D O I
10.1088/1361-6560/ad4d52
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Spatially fractionated radiotherapy is showing promise as a treatment modality. Initial focus was on beams of photons at low energy produced from a synchrotron but more recently research has expanded to include applications in proton therapy. Interest in photon beams remains and this is the focus of this paper Approach. This study presents a 3D printed tungsten minibeam collimator intended to produce peak-to-valley dose ratios (PVDR) of between seven and ten with a 1 MV, bremsstrahlung generated, photon beam. The design of the collimator is motivated by a Monte Carlo study estimating the PVDR for different collimator designs at different energies. This collimator was characterised on a clinical linear accelerator (Elekta VersaHD) as well as an orthovoltage unit. Main results. The performance of the fabricated collimator was measured on Elekta VersaHD running in unflattened mode with a 6 MV beam. On the Elekta VersaHD units the PVDR was measured to be between approximately 1.5 and 2.0 at 3 cm deep. For measurements with the orthovoltage unit PVDRs of greater than 10 were observed at a depth of 4 cm. Significance. The results confirmed that the predictions from simulation could be reproduced on linear accelerators currently in clinical usage, producing PVDRs between 2-2.5. Using the model to predict PVDRs using 1 MV photon beams, the threshold considered to produce enhanced normal tissue dose tolerance ( > 7) was surpassed. This suggests the possibility of using such techniques with versions of existing Linac technology which have been modified to operate at low energy and high beam currents.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Monte Carlo simulation of a clinical linear accelerator
    Lin, SY
    Chu, TC
    Lin, JP
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (06) : 759 - 765
  • [2] Concept of a clinical linear accelerator optimized for IMRT with Monte Carlo simulation
    Yasunaga, M.
    Yagi, M.
    Mukumoto, N.
    Tsujii, K.
    Uehara, K.
    Saito, S.
    Numasaki, H.
    Teshima, T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S670 - S670
  • [3] Monte Carlo Simulation of Linear Accelerator for Dosimetry Analysis
    Francis, Seenia
    Suresh, Darshana
    Nath, Shalini
    Lakshmi, Saai D. R.
    Jayaraj, P. B.
    Puzhakkal, Niyas
    Pournami, P. N.
    2021 6TH INTERNATIONAL CONFERENCE FOR CONVERGENCE IN TECHNOLOGY (I2CT), 2021,
  • [4] Characterisation of a megavoltage linear accelerator Bremsstrahlung by means of Monte Carlo simulations
    Valente, M.
    Malano, F.
    Perez, P.
    Castro, N.
    Carrasco, F.
    X-RAY SPECTROMETRY, 2010, 39 (06) : 384 - 390
  • [5] Design of a spherical applicator for intraoperative radiotherapy with a linear accelerator-a Monte Carlo simulation
    Ma, P.
    Li, Y.
    Tian, Y.
    Liu, B.
    Zhou, F.
    Dai, J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (01):
  • [6] Application of variance reduction techniques in Monte Carlo simulation of clinical electron linear accelerator
    Zoubair, M.
    El Bardouni, T.
    El Gonnouni, L.
    Boulaich, Y.
    El Bakkari, B.
    El Younoussi, C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 661 (01): : 93 - 97
  • [7] Monte Carlo simulation of a medical linear accelerator for radiotherapy use
    Serrano, B.
    Hachem, A.
    Franchisseur, E.
    Herault, J.
    Marcie, S.
    Costa, A.
    Bensadoun, R. J.
    Barthe, J.
    Gerard, J. P.
    RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 506 - 509
  • [8] Monte Carlo characterisation of a fan beam collimator (WIP)
    Martínez-Davalos, A
    Rodriguez-Villafuert, M
    MEDICAL PHYSICS, 2003, 30 (06) : 1462 - 1462
  • [9] Determining PGAA collimator plug design using Monte Carlo simulation
    Jalil, A.
    Chetaine, A.
    Amsil, H.
    Embarch, K.
    Benchrif, A.
    Laraki, K.
    Marah, H.
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2021, 53 (03) : 942 - 948
  • [10] Collimator design for neutron radiography station using Monte Carlo simulation
    Wonglee, S.
    Liamsuwan, T.
    Pornroongruengchok, W.
    Channuie, J.
    Picha, R.
    Khaweerat, S.
    SIAM PHYSICS CONGRESS 2018 (SPC2018): A CREATIVE PATH TO SUSTAINABLE INNOVATION, 2018, 1144