A comparison of the relative biological effectiveness of low energy electronic brachytherapy sources in breast tissue: a Monte Carlo study

被引:21
|
作者
White, Shane A. [1 ]
Reniers, Brigitte [1 ]
de Jong, Evelyn E. C. [1 ]
Rusch, Thomas [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] Maastricht Univ, Med Ctr, Dept Radiat Oncol MAASTRO, GROW Sch Oncol & Dev Biol, NL-6200 MD Maastricht, Netherlands
[2] Xoft, San Jose, CA USA
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 01期
关键词
relative biological effectiveness; low energy brachytherapy; electronic brachytherapy; Monte Carlo; INTRAOPERATIVE RADIOTHERAPY TARGIT; CLUSTERED DNA LESIONS; X-RAYS; DAMAGE; RBE; IRRADIATION; SIMULATION; INDUCTION; DOSIMETRY; OXYGEN;
D O I
10.1088/0031-9155/61/1/383
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electronic brachytherapy sources use low energy photons to treat the tumor bed during or after breast-conserving surgery. The relative biological effectiveness of two electronic brachytherapy sources was explored to determine if spectral differences due to source design influenced radiation quality and if radiation quality decreased with distance in the breast. The RBE was calculated through the number of DNA double strand breaks (RBEDSB) using the Monte Carlo damage simulator (MCDS) in combination with other Monte Carlo electron/photon spectrum calculations. 50kVp photons from the Intrabeam (Carl Zeiss Surgical) and Axxent (Xoft) through 40-mm spherical applicators were simulated to account for applicator and tissue attenuation in a variety of breast tissue compositions. 40kVp Axxent photons were also simulated. Secondary electrons (known to be responsible for most DNA damage) spectra at different distance were inputted into MCDS to calculate the RBEDSB. All RBEDSB used a cobalt-60 reference. RBEDSB data was combined with corresponding average photon spectrum energy for the Axxent and applied to model-based average photon energy distributions to produce an RBEDSB map of an accelerated partial breast irradiation (APBI) patient. Both Axxent and Intrabeam 50kVp spectra were shown to have a comparable RBEDSB of between 1.4 and 1.6 at all distances in spite of progressive beam hardening. The Axxent 40kVp also demonstrated a similar RBEDSB at distances. Most RBEDSB variability was dependent on the tissue type as was seen in rib (RBEDSB approximate to 1.4), gland (approximate to 1.55), adipose (approximate to 1.59), skin (approximate to 1.52) and lung (approximate to 1.50). RBEDSB variability between both sources was within 2%. A correlation was shown between RBEDSB and average photon energy and used to produce an RBEDSB map of a dose distribution in an APBI patient dataset. Radiation quality is very similar between electronic brachytherapy sources studied. No significant reductions in RBEDSB were observed with increasing distance from the source.
引用
收藏
页码:383 / 399
页数:17
相关论文
共 50 条
  • [21] Monte Carlo aided design of an improved well-type ionization chamber for low energy brachytherapy sources
    Bohm, Tim D.
    Micka, John A.
    DeWerd, Larry A.
    MEDICAL PHYSICS, 2007, 34 (04) : 1274 - 1285
  • [22] The Effects of Tissue Inhomogeneities and Biological Effectiveness on the Outcome Prediction of Electronic Brachytherapy for Accelerator Partial Breast Irradiation
    Guo, B.
    Cheng, C.
    Shi, C.
    Papanikolaou, N.
    MEDICAL PHYSICS, 2011, 38 (06) : 3736 - +
  • [23] The difference of scoring dose to water or tissues in Monte Carlo dose calculations for low energy brachytherapy photon sources
    Landry, Guillaume
    Reniers, Brigitte
    Pignol, Jean-Philippe
    Beaulieu, Luc
    Verhaegen, Frank
    MEDICAL PHYSICS, 2011, 38 (03) : 1526 - 1533
  • [24] BREAST TISSUE MODELING IN MONTE CARLO POST-IMPLANT EVALUATION OF BREAST LDR BRACHYTHERAPY
    Afsharpour, H.
    Landry, G.
    Reniers, B.
    Pignol, J. P.
    Beaulieu, L.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S30 - S31
  • [25] A Sensitivity Study of Tissue Characterization for Brachytherapy Monte Carlo Dose Calculation
    Bedwani, S.
    Carrier, J.
    Bouchard, H.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [26] Monte Carlo Dose Comparison Assessing Material Inhomogeneity Effects in Breast Brachytherapy
    Graf, M.
    Scanderbeg, D.
    Yashar, C.
    Jiang, S.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [27] MONTE CARLO-BASED INVESTIGATION OF MICRODOSIMETRIC DISTRIBUTION OF HIGH ENERGY BRACHYTHERAPY SOURCES
    Chattaraj, Arghya
    Selvam, T. Palani
    Datta, D.
    RADIATION PROTECTION DOSIMETRY, 2019, 187 (01) : 115 - 128
  • [28] Tissue modeling schemes in low energy breast brachytherapy
    Afsharpour, Hossein
    Landry, Guillaume
    Reniers, Brigitte
    Pignol, Jean-Philippe
    Beaulieu, Luc
    Verhaegen, Frank
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22): : 7045 - 7060
  • [29] Dosimetric parameters for three low-energy brachytherapy sources using the Monte Carlo N-Particle code
    DeMarco, JJ
    Hugo, G
    Solberg, TD
    MEDICAL PHYSICS, 2002, 29 (05) : 662 - 668
  • [30] Simulation study on potential accuracy gains from dual energy CT tissue segmentation for low-energy brachytherapy Monte Carlo dose calculations
    Landry, Guillaume
    Granton, Patrick V.
    Reniers, Brigitte
    Ollers, Michel C.
    Beaulieu, Luc
    Wildberger, Joachim E.
    Verhaegen, Frank
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (19): : 6257 - 6278