Effects of 6LiF and 6LiCl codoping on gamma and neutron detection performance of NaI:Tl single crystal scintillators

被引:0
|
作者
Wang, Jingkang [1 ]
Wang, Chenger [2 ]
Tong, Yufeng [1 ]
Li, Huanyin [1 ]
Shi, Jian [1 ]
Ren, Guohao [1 ]
Sun, Xilei [3 ]
Wu, Yuntao [1 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Ceram, Shanghai 201899, Peoples R China
[2] Grirem Adv Mat Co Ltd, Natl Engn Res Ctr Rare Earth, Beijing 100088, Peoples R China
[3] Chinese Acad Sci, Inst High Energy Phys, Beijing 100043, Peoples R China
基金
中国国家自然科学基金;
关键词
NaI:Tl; Inorganic scintillator; Codoping; Energy resolution; n/gamma discrimination;
D O I
10.1016/j.radmeas.2024.107232
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
NaI:Tl single crystals codoped with (LiI)-Li-6 have been considered as promising neutron-gamma scintillators due to low cost, scalability, and high efficiency in detecting and discriminating neutrons and gamma-rays. This study aims to study the effects of (LiF)-Li-6 and (LiCl)-Li-6 codoping on the NaI:Tl single crystals grown by the multi-ampoule Bridgman method. The impact of (LiF)-Li-6 and (LiCl)-Li-6 on the optical properties, gamma spectroscopy, and pulse shape discrimination (PSD) of neutron and gamma detection performance were analyzed. Both (LiF)-Li-6 and (LiCl)-Li-6 caused a decrease in gamma-ray light yield and a deterioratrion of energy resolution with increasing codoping concentrations, similar to the scenarios of (LiI)-Li-6 codoping. Nonetheless, for NaI:Tl,(LiF)-Li-6, the light yield and PSD Figure-of-Merit (FoM) are almost unchanged about 37,500 photons/MeV and 4.6 when the Li content reaches 2 at%, which is better than the scenario of (LiI)-Li-6 codoping. Combing with the advantages of non-hygroscopic nature of LiF and a higher utilization yield of Li element in LiF comparing with LiI, NaI:Tl,(LiF)-Li-6 single crystals could be regarded as more cost-effective scintillators for neutron detection applications.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] 6LiF:ZnS(Ag) Neutron Detector Performance Optimized Using Waveform Recordings and ROC Curves
    Pritchard, K.
    Osovizky, A.
    Ziegler, J.
    Binkley, E.
    Tsai, P.
    Hadad, N.
    Jackson, M.
    Hurlbut, C.
    Baltic, G. M.
    Majkrzak, C. F.
    Maliszewskyj, N. C.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2020, 67 (01) : 414 - 421
  • [32] Eu-doped 6LiF-SrF2 eutectic scintillators for neutron detection
    Yanagida, Takayuki
    Fukuda, Kentaro
    Fujimoto, Yutaka
    Kawaguchi, Noriaki
    Kurosawa, Shunsuke
    Yamazaki, Atsushi
    Watanabe, Kenichi
    Futami, Yoshisuke
    Yokota, Yuui
    Pejchal, Jan
    Yoshikawa, Akira
    Uritani, Akira
    Iguchi, Tetsuo
    OPTICAL MATERIALS, 2012, 34 (05) : 868 - 871
  • [33] Research on Wide Energy Spectrum Neutron Detection Technology Based on 6LiF/ZnS(Ag) and Scintillating Fiber
    Xiong B.
    Wu Z.
    Wan B.
    Yang D.
    Li K.
    Li G.
    Zhang H.
    Hedongli Gongcheng/Nuclear Power Engineering, 2022, 43 (04): : 213 - 217
  • [34] Cold neutron radiation dose effects on a 6LiF:ZnS(Ag) neutron detector with wavelength shifting fibers and SiPM photodetector
    Pritchard, K.
    Hussey, D.
    Osovizky, A.
    Ziegler, J.
    Binkley, E.
    Tsai, P.
    Hadad, N.
    Jackson, M.
    Hurlbut, C.
    Baltic, G. M.
    Majkrzak, C. F.
    Maliszewskyj, N. C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 986
  • [35] Detection of thermal neutrons using ZnS(Ag):6LiF neutron scintillator read out with WLS fibers and SiPMs
    Hildebrandt, M.
    Stoykov, A.
    Mosset, J. -B.
    Greuter, U.
    Schlumpf, N.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 824 : 204 - 207
  • [36] Performance of a Prototype Large Area Neutron Detector Based on 6LiF:ZnS(Ag) with MPPC Read-out
    Vacheret, A.
    Scovell, P. R.
    Haigh, M.
    Shitov, Yu.
    Waldron, A.
    Weber, A.
    Giudicelli, A.
    Fierfort, A.
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [37] Efficiency and directional effects in the detection of low-energy recoil nuclei in a NaI(Tl) single crystal
    Graichen, J
    Maier, K
    Schüth, J
    Siepe, A
    von Witsch, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 485 (03): : 774 - 779
  • [38] Improving thermal-neutron detection efficiency of silicon neutron detectors using the combined layers of 10B4C on 6LiF
    Jiang, Yong
    Wu, Jian
    Yin, Yanpeng
    Lu, Yi
    Wu, Kunlin
    Fan, Xiaoqiang
    Lei, Jiarong
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 932 : 50 - 55
  • [39] COUNTING EFFICIENCY AND NEUTRON/GAMMA RATIO FOR KDP: TL+ AND UPS-923A SCINTILLATORS IN A SINGLE PHOTON DETECTION MODE
    Onyshchenko, Gennadiy
    Yakymenko, Ivan
    Grynyov, Borys
    Ryzhikov, Volodymyr
    Voronov, Olexiy
    Naydenov, Sergei
    EAST EUROPEAN JOURNAL OF PHYSICS, 2020, (03): : 54 - 61
  • [40] Evaluation of two thermal neutron detection units consisting of ZnS/6LiF scintillating layers with embedded WLS fibers read out with a SiPM
    Mosset, J. -B.
    Stoykov, A.
    Greuter, U.
    Hildebrandt, M.
    Schlumpf, N.
    Van Swygenhoven, H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 764 : 299 - 304