Comparison of True and Virtual Non-Contrast Images of Liver Obtained with Single-Source Twin Beam and Dual-Source Dual-Energy CT

被引:0
|
作者
Lee, Jeong Sub [1 ]
Choi, Guk Myung [1 ]
Kim, Bong Soo [1 ]
Ko, Su Yeon [1 ]
Lee, Kyung Ryeol [1 ]
Kim, Jeong Jae [1 ]
Kim, Doo Ri [1 ]
机构
[1] Jeju Natl Univ, Coll Med, Jeju Natl Univ Hosp, Dept Radiol, 15 Aran 13 Gil, Jeju 63241, South Korea
来源
关键词
Radiography; Dual-Energy Scanned Projection; Tomography; X-Ray Computed; Feasibility Studies; COMPUTED-TOMOGRAPHY; ATTENUATION VALUES; NONCONTRAST IMAGES; UNENHANCED IMAGES; ENHANCED CT; ACCURACY; QUALITY;
D O I
10.3348/jksr.2021.0193
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To assess the magnitude of differences between attenuation values of the true non -contrast image (TNC) and virtual non -contrast image (VNC) derived from twin -beam dual -energy CT (tbDECT) and dual-source DECT (dsDECT).<br /> Materials and Methods This retrospective study included 62 patients who underwent liver dynamic DECT with tbDECT ( n = 32) or dsDECT ( n = 30). Arterial VNC (AVNC), portal VNC (PVNC), and delayed VNC (DVNC) were reconstructed using multiphasic DECT. Attenuation values of multiple intra-abdominal organs ( n = 11) on TNCs were subsequently compared to those on multiphasic VNCs. Further, we investigated the percentage of cases with an absolute difference between TNC and VNC of <= 10 Hounsfield units (HU).<br /> Results For the mean attenuation values of TNC and VNC, 33 items for each DECT were compared according to the multiphasic VNCs and organs. More than half of the comparison items for each DECT showed significant differences (tbDECT 17/33; dsDECT 19/33; Bonferroni correction p < 0.0167). The percentage of cases with an absolute difference <= 10 HU was 56.7%, 69.2%, and 78.6% in AVNC, PVNC, and DVNC in tbDECT, respectively, and 70.5%, 78%, and 78% in dsDECT, respectively.<br /> Conclusion VNCs derived from the two DECTs were insufficient to replace TNCs because of the considerable difference in attenuation values.
引用
收藏
页码:170 / 184
页数:15
相关论文
共 50 条
  • [21] Motion artifacts in kidney stone imaging using single-source and dual-source dual-energy CT scanners: a phantom study
    El-Sayed H. Ibrahim
    Joseph G. Cernigliaro
    Robert A. Pooley
    James C. Williams
    William E. Haley
    Abdominal Imaging, 2015, 40 : 3161 - 3167
  • [22] Motion artifacts in kidney stone imaging using single-source and dual-source dual-energy CT scanners: a phantom study
    Ibrahim, El-Sayed H.
    Cernigliaro, Joseph G.
    Pooley, Robert A.
    Williams, James C.
    Haley, William E.
    ABDOMINAL IMAGING, 2015, 40 (08): : 3161 - 3167
  • [23] Dual Energy CT Protocol Optimization for Increased Blood Detectability of Virtual Non-Contrast Images in a Single Source Dual Energy CT System: A Phantom Study
    Olguin, C.
    Leon, S.
    Barreto, I.
    De Jesus, R.
    Arreola, M.
    MEDICAL PHYSICS, 2020, 47 (06) : E529 - E530
  • [24] Dual-source CT coronary angiography in patients with atrial fibrillation: Comparison with single-source CT
    Wang, Yining
    Zhang, Zhuhua
    Kong, Lingyan
    Song, Lan
    Merges, Reto D.
    Chen, Jiuhong
    Jin, Zhengyu
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : 434 - 441
  • [25] Dual-Energy and Dual-Source CT: Is There a Role in the Abdomen and Pelvis?
    Fletcher, Joel G.
    Takahashi, Naoki
    Hartman, Robert
    Guimaraes, Luis
    Huprich, James E.
    Hough, David M.
    Yu, Lifeng
    McCollough, Cynthia H.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2009, 47 (01) : 41 - +
  • [26] Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography
    Alikhani, Babak
    Werner, Martin
    Jamali, Leila
    Wacker, Frank
    Werncke, Thomas
    ACADEMIC RADIOLOGY, 2019, 26 (08) : 1095 - 1101
  • [27] Comparison of Dual- and Single-Source Dual-Energy CT for Diagnosis of Acute Pulmonary Artery Embolism
    Petritsch, Bernhard
    Pannenbecker, Pauline
    Weng, Andreas Max
    Veldhoen, Simon
    Grunz, Jan-Peter
    Bley, Thorsten Alexander
    Kosmala, Aleksander
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2021, 193 (04): : 427 - 436
  • [28] Motion Artifacts in Kidney Stone Imaging Using Dual-Energy CT. A Phantom Study Comparing Single-Source and Dual-Source Scanners
    Ibrahim, El-Sayed
    Pooley, Robert
    Cernigliaro, Joseph
    Bridges, Mellena
    Williams, James
    Haley, William
    2015 IEEE 12th International Symposium on Biomedical Imaging (ISBI), 2015, : 1159 - 1162
  • [29] Virtual versus true non-contrast dual-energy CT imaging for the diagnosis of aortic intramural hematoma
    Salim Si-Mohamed
    Nicolas Dupuis
    Valérie Tatard-Leitman
    David Rotzinger
    Sara Boccalini
    Matthias Dion
    Alain Vlassenbroek
    Philippe Coulon
    Yoad Yagil
    Nadav Shapira
    Philippe Douek
    Loic Boussel
    European Radiology, 2019, 29 : 6762 - 6771
  • [30] Virtual Monochromatic Images from Dual-Energy Multidetector CT: Variance in CT Numbers from the Same Lesion between Single-Source Projection-based and Dual-Source Image-based Implementations
    Mileto, Achille
    Barina, Andrew
    Marin, Daniele
    Stinnett, Sandra S.
    Choudhury, Kingshuk Roy
    Wilson, Joshua M.
    Nelson, Rendon C.
    RADIOLOGY, 2016, 279 (01) : 269 - 277