Repeatability and Agreement of 4 Biometers Measuring Corneal Astigmatism in Eyes With Irregular Corneal Astigmatism Component

被引:0
|
作者
Wang, Lan [1 ]
Shao, Xu [1 ]
Fu, Jin-Ling [1 ]
Hu, Jing-Jing [1 ]
Zhou, Kai-Jing [1 ]
Li, Ran [1 ]
Yu, A-Yong [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Eye Hosp, Natl Clin Res Ctr Ocular Dis, Wenzhou, Peoples R China
[2] Wenzhou Med Univ, Eye Hosp, Natl Engn Res Ctr Ophthalmol & Optometry, Wenzhou, Peoples R China
[3] Wenzhou Med Univ, Eye Hosp, 270 Xueyuan West Rd, Wenzhou 325027, Zhejiang, Peoples R China
基金
国家重点研发计划;
关键词
SOURCE OPTICAL BIOMETRY; INTRAOCULAR-LENS; TOPOGRAPHY MEASUREMENTS; KERATOMETRY; POSTERIOR; COHERENCE; COMPARABILITY;
D O I
10.1016/j.ajo.2024.05.002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To investigate the repeatability and agreement of corneal astigmatism measurements in eyes with irregular corneal astigmatism component (ICAC) using four devices: IOLMaster 700 biometer, Lenstar 900 biometer, iTrace, and Pentacam.<br /> DESIGN: Prospective cross-sectional reliability analysis.<br /> METHODS: Sixty-four eyes (52 patients) with ICAC were examined three times using the four devices. The eye with ICAC in this study is defined as the cornea has a certain degree of irregular astigmatism (asymmetric and/or skewed bowtie pattern of corneal topography according to corneal topography classification), accompanied with total corneal higher-order aberrations in the 4 mm zone of 0.3 mu m or greater. Corneal astigmatism was evaluated using three categories: anterior corneal astigmatism (ACA), posterior corneal astigmatism, and total corneal astigmatism (TCA). The repeatability was determined using the Delta Ast (arithmetic mean of vector differences among three repeated corneal astigmatism measurements). Bland-Altman plots and astigmatism vector analyses were employed to assess agreement.<br /> RESULTS: The IOLMaster 700 ( Delta Ast = 0.27 +/- 0.20 D) showcased higher repeatability in ACA measurements compared to iTrace ( Delta Ast = 0.37 +/- 0.38 D, P = .040) and Pentacam ( Delta Ast = 0.50 +/- 0.22 D, P < .001), and paralleled the performance of Lenstar 900 ( Delta Ast = 0.31 +/- 0.26 D, P = .338). The Pentacam ( Delta Ast = 0.09 +/- 0.07 D, P < .001) demonstrated superior repeatability in posterior corneal astigmatism, whereas the IOLMaster 700 ( Delta Ast = 0.33 +/- 0.23 D, P < .001) excelled in TCA. The IOLMaster 700 exhibited good agreement with either Lenstar 900 or iTrace, characterized by narrow 95% limits of agreement and clinically acceptable vector differences. Conversely, vector differences between Pentacam and the other three devices in ACA and TCA measurements were clinically significant, exceeding 0.50 D (all P < .05).<br /> CONCLUSIONS: In terms of repeatability of corneal astigmatism measurements in eyes with ICAC, the IOLMaster 700 and Lenstar 900 outperformed iTrace and Pentacam. While the IOLMaster 700 can be used interchangeably with either Lenstar 900 or iTrace, the Pentacam is not interchangeable with the other three devices.
引用
收藏
页码:200 / 212
页数:13
相关论文
共 50 条
  • [21] Corneal Topographic Astigmatism (CorT) to Quantify Total Corneal Astigmatism
    Alpins, Noel
    Ong, James K. Y.
    Stamatelatos, George
    JOURNAL OF REFRACTIVE SURGERY, 2015, 31 (03) : 182 - U110
  • [22] Comparability and repeatability of different methods of corneal astigmatism assessment
    Ferreira, Tiago B.
    Ribeiro, Filomena J.
    CLINICAL OPHTHALMOLOGY, 2018, 12 : 29 - 34
  • [23] Evaluation of keratometric and total corneal astigmatism measurements from optical biometers and anterior segment tomographers and mapping to reconstructed corneal astigmatism vector components
    Langenbucher, Achim
    Szentmary, Nora
    Cayless, Alan
    Hoffmann, Peter
    Wendelstein, Jascha
    Pantanelli, Seth
    PLOS ONE, 2025, 20 (01):
  • [24] Repeatability and comparability of the Galilei-G4 and Cassini in measuring corneal power and astigmatism in normal and post-refractive surgery eyes
    Mohamed Mohamed
    Mahmood Khan
    Amritha Kanakamedala
    Isha Gupta
    Li Wang
    Douglas Koch
    Zaina Al-Mohtaseb
    Scientific Reports, 11
  • [25] Repeatability and comparability of the Galilei-G4 and Cassini in measuring corneal power and astigmatism in normal and post-refractive surgery eyes
    Mohamed, Mohamed
    Khan, Mahmood
    Kanakamedala, Amritha
    Gupta, Isha
    Li Wang
    Koch, Douglas
    Al-Mohtaseb, Zaina
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [26] Quantitative evaluation of corneal irregular astigmatism and wavefront aberrations
    Oshika, T
    CORNEA, 2000, 19 (06) : S165 - S172
  • [27] CONTACT-LENSES IN HIGH AND IRREGULAR CORNEAL ASTIGMATISM
    SCHMITT, H
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1973, 163 (03) : 338 - 343
  • [28] Effectiveness of toric intraocular lens implantation for correcting irregular corneal astigmatism in cataract eyes
    Chen, Xiteng
    Jiang, Yuanfeng
    Gao, Nan
    Gao, Yichen
    Yang, Jun
    Bu, Shaochong
    Tian, Fang
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [29] Measurements of Anterior and Posterior Corneal Curvatures with OCT and Scheimpflug Biometers in Patients with Low Total Corneal Astigmatism
    Muzyka-Wozniak, Maria
    Oleszko, Adam
    Grzybowski, Andrzej
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (23)
  • [30] Intrastromal corneal ring segments for the treatment of irregular astigmatism
    Kymionis, George D.
    Bouzoukis, Dimitrios I.
    Haft, Payman
    Siganos, Charambos S.
    Yoo, Sonia
    EXPERT REVIEW OF OPHTHALMOLOGY, 2008, 3 (01) : 9 - 15