Repeatability and Agreement of 4 Biometers Measuring Corneal Astigmatism in Eyes With Irregular Corneal Astigmatism Component

被引:0
|
作者
Wang, Lan [1 ]
Shao, Xu [1 ]
Fu, Jin-Ling [1 ]
Hu, Jing-Jing [1 ]
Zhou, Kai-Jing [1 ]
Li, Ran [1 ]
Yu, A-Yong [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Eye Hosp, Natl Clin Res Ctr Ocular Dis, Wenzhou, Peoples R China
[2] Wenzhou Med Univ, Eye Hosp, Natl Engn Res Ctr Ophthalmol & Optometry, Wenzhou, Peoples R China
[3] Wenzhou Med Univ, Eye Hosp, 270 Xueyuan West Rd, Wenzhou 325027, Zhejiang, Peoples R China
基金
国家重点研发计划;
关键词
SOURCE OPTICAL BIOMETRY; INTRAOCULAR-LENS; TOPOGRAPHY MEASUREMENTS; KERATOMETRY; POSTERIOR; COHERENCE; COMPARABILITY;
D O I
10.1016/j.ajo.2024.05.002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To investigate the repeatability and agreement of corneal astigmatism measurements in eyes with irregular corneal astigmatism component (ICAC) using four devices: IOLMaster 700 biometer, Lenstar 900 biometer, iTrace, and Pentacam.<br /> DESIGN: Prospective cross-sectional reliability analysis.<br /> METHODS: Sixty-four eyes (52 patients) with ICAC were examined three times using the four devices. The eye with ICAC in this study is defined as the cornea has a certain degree of irregular astigmatism (asymmetric and/or skewed bowtie pattern of corneal topography according to corneal topography classification), accompanied with total corneal higher-order aberrations in the 4 mm zone of 0.3 mu m or greater. Corneal astigmatism was evaluated using three categories: anterior corneal astigmatism (ACA), posterior corneal astigmatism, and total corneal astigmatism (TCA). The repeatability was determined using the Delta Ast (arithmetic mean of vector differences among three repeated corneal astigmatism measurements). Bland-Altman plots and astigmatism vector analyses were employed to assess agreement.<br /> RESULTS: The IOLMaster 700 ( Delta Ast = 0.27 +/- 0.20 D) showcased higher repeatability in ACA measurements compared to iTrace ( Delta Ast = 0.37 +/- 0.38 D, P = .040) and Pentacam ( Delta Ast = 0.50 +/- 0.22 D, P < .001), and paralleled the performance of Lenstar 900 ( Delta Ast = 0.31 +/- 0.26 D, P = .338). The Pentacam ( Delta Ast = 0.09 +/- 0.07 D, P < .001) demonstrated superior repeatability in posterior corneal astigmatism, whereas the IOLMaster 700 ( Delta Ast = 0.33 +/- 0.23 D, P < .001) excelled in TCA. The IOLMaster 700 exhibited good agreement with either Lenstar 900 or iTrace, characterized by narrow 95% limits of agreement and clinically acceptable vector differences. Conversely, vector differences between Pentacam and the other three devices in ACA and TCA measurements were clinically significant, exceeding 0.50 D (all P < .05).<br /> CONCLUSIONS: In terms of repeatability of corneal astigmatism measurements in eyes with ICAC, the IOLMaster 700 and Lenstar 900 outperformed iTrace and Pentacam. While the IOLMaster 700 can be used interchangeably with either Lenstar 900 or iTrace, the Pentacam is not interchangeable with the other three devices.
引用
收藏
页码:200 / 212
页数:13
相关论文
共 50 条
  • [1] Agreement Between Internal Astigmatism and Posterior Corneal Astigmatism in Pseudophakic Eyes
    Feizi, Sepehr
    Delfazayebaher, Siamak
    Javadi, Mohammad Ali
    JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (06) : 379 - +
  • [2] Comparison of corneal irregular astigmatism by the type of corneal regular astigmatism
    Yuta Ueno
    Risa Nomura
    Takahiro Hiraoka
    Katsuhito Kinoshita
    Mutsuko Ohara
    Tetsuro Oshika
    Scientific Reports, 11
  • [3] Comparison of corneal irregular astigmatism by the type of corneal regular astigmatism
    Ueno, Yuta
    Nomura, Risa
    Hiraoka, Takahiro
    Kinoshita, Katsuhito
    Ohara, Mutsuko
    Oshika, Tetsuro
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [4] Influence of posterior corneal astigmatism on total corneal astigmatism in eyes with moderate to high astigmatism
    Savini, Giacomo
    Versaci, Francesco
    Vestri, Gabriele
    Ducoli, Pietro
    Naeser, Kristian
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (10): : 1645 - 1653
  • [5] Agreement Between Refractive and Corneal Astigmatism in Pseudophakic Eyes
    Tejedor, Jaime
    Guirao, Antonio
    CORNEA, 2013, 32 (06) : 783 - 790
  • [6] Influence of Posterior Corneal Astigmatism on Total Corneal Astigmatism in Eyes With Keratoconus
    Savini, Giacomo
    Naeser, Kristian
    Schiano-Lomoriello, Domenico
    Mularoni, Alessandro
    CORNEA, 2016, 35 (11) : 1427 - 1433
  • [7] CORNEAL TOPOGRAPHY AND IRREGULAR ASTIGMATISM
    RAASCH, TW
    OPTOMETRY AND VISION SCIENCE, 1995, 72 (11) : 809 - 815
  • [8] Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices
    Feng, Yue
    Nitter, Tore
    Bertelsen, Geir
    Stojanovic, Aleksandar
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 52 (08): : 800 - 810
  • [9] Changes in corneal irregular astigmatism and visual acuity in eyes with keratoconus
    Suzuki, M
    Tanabe, T
    Amano, S
    Oshika, T
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [10] Evaluating keratometry and corneal astigmatism data from biometers and anterior segment tomographers and mapping to reconstructed corneal astigmatism
    Langenbucher, Achim
    Taroni, Leonardo
    Coutinho, Catarina P.
    Cayless, Alan
    Szentmary, Nora
    Hoffmann, Peter
    Wendelstein, Jascha
    Savini, Giacomo
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 52 (06): : 627 - 638