Measurements of Arterial Occlusion Pressure Using Hand-Held Devices

被引:4
|
作者
Vehrs, Pat R. [1 ]
Richards, Shay [1 ]
Allen, Josh [1 ]
Barrett, Rachel [1 ]
Blazzard, Chase [1 ]
Burbank, Tyler [1 ]
Hart, Hannah [1 ]
Kasper, Nicole [1 ]
Lacey, Ryan [1 ]
Lopez, Daniela [1 ]
Fellingham, Gilbert W. [2 ]
机构
[1] Brigham Young Univ, Dept Exercise Sci, Provo, UT 84602 USA
[2] Brigham Young Univ, Dept Stat, Provo, UT USA
关键词
blood flow restriction; blood flow restriction exercise; pulse oximetry; Doppler; BLOOD-FLOW RESTRICTION; CUFF WIDTH; EXERCISE; HEMODYNAMICS; SEX;
D O I
10.1519/JSC.0000000000004716
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Arterial occlusion pressure (AOP) of the brachial artery was measured simultaneously using Doppler ultrasound (US), a hand-held Doppler (HHDOP), and a pulse oximeter (PO) in the dominant (DOM) and nondominant (NDOM) arms of males (n = 21) and females (n = 23) using continuous (CONT) and incremental (INCR) cuff inflation protocols. A mixed-model analysis of variance revealed significant (p < 0.05) overall main effects between AOP measured using a CONT (115.7 +/- 10.9) or INCR (115.0 +/- 11.5) cuff inflation protocol; between AOP measured using US (116.3 +/- 11.2), HHDOP (115.4 +/- 11.2), and PO (114.4 +/- 11.2); and between males (120.7 +/- 10.6) and females (110.5 +/- 9.4). The small overall difference (1.81 +/- 3.3) between US and PO measures of AOP was significant (p < 0.05), but the differences between US and HHDOP and between HHDOP and PO measures of AOP were not significant. There were no overall differences in AOP between the DOM and NDOM arms. Trial-to-trial variance in US measurements of AOP was not significant when using either cuff inflation protocol but was significant when using HHDOP and PO and a CONT cuff inflation protocol. Bland-Altman plots revealed reasonable limits of agreement for both HHDOP and PO measures of AOP. The small differences in US, HHDOP, and PO measurements of AOP when using CONT or INCR cuff inflation protocols are of minimal practical importance. The choice of cuff inflation protocol is one of personal preference. Hand-held Doppler of PO can be used to assess AOP before using blood flow restriction during exercise.
引用
收藏
页码:873 / 880
页数:8
相关论文
共 50 条
  • [21] Hand-held test, pressure calibration system
    不详
    CONTROL ENGINEERING, 2008, 55 (07) : 33 - 33
  • [22] Hand-Held Haptic Navigation Devices for Actual Walking
    Kappers, Astrid M. L.
    Oen, Max Fa Si
    Junggeburth, Tessa J. W.
    Plaisier, Myrthe A.
    IEEE TRANSACTIONS ON HAPTICS, 2022, 15 (04) : 655 - 666
  • [23] Distributed knowledge bases and hand-held computing devices
    Capraro, GT
    Berdan, GB
    Capraro, CT
    Spina, J
    IC-AI'2001: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON ARTIFICIAL INTELLIGENCE, VOLS I-III, 2001, : 1362 - 1368
  • [24] TOWARDS MULTISPECTRAL DATA ACQUISITION WITH HAND-HELD DEVICES
    Khan, Rahat
    van de Weijer, Joost
    Karatzas, Dimosthenis
    Muselet, Damien
    2013 20TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP 2013), 2013, : 2053 - 2057
  • [25] Frequency reconfigurable antenna for hand-held wireless devices
    Bhellar, Benazeer
    Tahir, Farooq A.
    IET MICROWAVES ANTENNAS & PROPAGATION, 2015, 9 (13) : 1412 - 1417
  • [26] Absorption Related to Hand-Held Devices in Data Mode
    Andersen, Jorgen Bach
    Nielsen, Jesper Odum
    Pedersen, Frolund
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2016, 58 (01) : 47 - 53
  • [27] Saliency Aware Resource Saving in Hand-Held Devices
    Yadav, P.
    Ramasubramanian, N.
    Joshi, A.
    Tamizharasan, P.
    PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON COMMUNICATION AND SIGNAL PROCESSING 2016 (ICCASP 2016), 2017, 137 : 499 - 505
  • [28] A review of hand-held Electrocardiogram (ECG) recording devices
    Collins, M. C.
    Elliott, E. E.
    Stain, N. S.
    Malcolmson, J. M.
    Monkhouse, C. M.
    EUROPEAN JOURNAL OF CARDIOVASCULAR NURSING, 2021, 20 : 157 - 157
  • [29] Ephemeral Data Storage for Networks of Hand-held Devices
    Silva, Joao A.
    Monteiro, Ricardo
    Paulino, Herve
    Lourenco, Joao M.
    2016 IEEE TRUSTCOM/BIGDATASE/ISPA, 2016, : 1106 - 1113
  • [30] A study of the use of hand-held devices in an emergency department
    Gururajan, R