Measurements of Arterial Occlusion Pressure Using Hand-Held Devices

被引:4
|
作者
Vehrs, Pat R. [1 ]
Richards, Shay [1 ]
Allen, Josh [1 ]
Barrett, Rachel [1 ]
Blazzard, Chase [1 ]
Burbank, Tyler [1 ]
Hart, Hannah [1 ]
Kasper, Nicole [1 ]
Lacey, Ryan [1 ]
Lopez, Daniela [1 ]
Fellingham, Gilbert W. [2 ]
机构
[1] Brigham Young Univ, Dept Exercise Sci, Provo, UT 84602 USA
[2] Brigham Young Univ, Dept Stat, Provo, UT USA
关键词
blood flow restriction; blood flow restriction exercise; pulse oximetry; Doppler; BLOOD-FLOW RESTRICTION; CUFF WIDTH; EXERCISE; HEMODYNAMICS; SEX;
D O I
10.1519/JSC.0000000000004716
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Arterial occlusion pressure (AOP) of the brachial artery was measured simultaneously using Doppler ultrasound (US), a hand-held Doppler (HHDOP), and a pulse oximeter (PO) in the dominant (DOM) and nondominant (NDOM) arms of males (n = 21) and females (n = 23) using continuous (CONT) and incremental (INCR) cuff inflation protocols. A mixed-model analysis of variance revealed significant (p < 0.05) overall main effects between AOP measured using a CONT (115.7 +/- 10.9) or INCR (115.0 +/- 11.5) cuff inflation protocol; between AOP measured using US (116.3 +/- 11.2), HHDOP (115.4 +/- 11.2), and PO (114.4 +/- 11.2); and between males (120.7 +/- 10.6) and females (110.5 +/- 9.4). The small overall difference (1.81 +/- 3.3) between US and PO measures of AOP was significant (p < 0.05), but the differences between US and HHDOP and between HHDOP and PO measures of AOP were not significant. There were no overall differences in AOP between the DOM and NDOM arms. Trial-to-trial variance in US measurements of AOP was not significant when using either cuff inflation protocol but was significant when using HHDOP and PO and a CONT cuff inflation protocol. Bland-Altman plots revealed reasonable limits of agreement for both HHDOP and PO measures of AOP. The small differences in US, HHDOP, and PO measurements of AOP when using CONT or INCR cuff inflation protocols are of minimal practical importance. The choice of cuff inflation protocol is one of personal preference. Hand-held Doppler of PO can be used to assess AOP before using blood flow restriction during exercise.
引用
收藏
页码:873 / 880
页数:8
相关论文
共 50 条
  • [1] Using Hand-Held Arrays for Automotive NVH Measurements
    Gade, Svend
    Gomes, Jesper
    Hald, Jorgen
    SOUND AND VIBRATION, 2014, 48 (03): : 12 - 16
  • [2] Build applications for hand-held devices
    不详
    BYTE, 1997, 22 (09): : 145 - 145
  • [3] Hand motion and image stabilization in hand-held devices
    Or, Etay Mar
    Pundik, Dmitry
    IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 2007, 53 (04) : 1508 - 1512
  • [4] Hand motion and image stabilization in hand-held devices
    Mar-Or, Etay
    Dmitry, Pundik
    IMAGE QUALITY AND SYSTEM PERFORMANCE V, 2008, 6808
  • [5] Remote robot control using intelligent hand-held devices
    Wu, SD
    Chen, YM
    FOURTH INTERNATIONAL CONFERENCE ON COMPUTER AND INFORMATION TECHNOLOGY, PROCEEDINGS, 2004, : 587 - 592
  • [6] Hand-held metal detectors and medical devices: Measurements and testing for electromagnetic compatibility
    Witters, DM
    Buzduga, V
    Seidman, S
    Kainz, W
    Casamento, J
    Ruggera, P
    39TH ANNUAL 2005 INTERNATIONAL CARNAHAN CONFERENCE ON SECURITY TECHNOLOGY, PROCEEDINGS, 2005, : 312 - 314
  • [7] The rules of sailing races for hand-held devices
    Turner, KJ
    Jennings, MA
    JOURNAL OF NAVIGATION, 2002, 55 (02): : 263 - 275
  • [8] Interactive topology optimization on hand-held devices
    Niels Aage
    Morten Nobel-Jørgensen
    Casper Schousboe Andreasen
    Ole Sigmund
    Structural and Multidisciplinary Optimization, 2013, 47 : 1 - 6
  • [9] A fresh look at hand-held communication devices
    Chalmers, M
    PHYSICS WORLD, 2003, 16 (05) : 27 - 28
  • [10] Interactive topology optimization on hand-held devices
    Aage, Niels
    Nobel-Jorgensen, Morten
    Andreasen, Casper Schousboe
    Sigmund, Ole
    STRUCTURAL AND MULTIDISCIPLINARY OPTIMIZATION, 2013, 47 (01) : 1 - 6