Automated Test for Monthly Quality Assurance of Optical Surface Imaging Dynamic Localization Accuracy

被引:0
|
作者
Young, Kaleigh [1 ]
Wright, Eric A. [2 ]
机构
[1] Univ British Columbia, Fac Med, Okanagan Campus, Kelowna, BC, Canada
[2] Sunnybrook Hlth Sci Ctr, Odette Canc Ctr, Med Phys, Toronto, ON, Canada
关键词
optical imaging; radiation therapy technology; optical surface imaging; surface-guided radiotherapy; general radiation oncology;
D O I
10.7759/cureus.56242
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The American Association of Physicists in Medicine (AAPM) recently published the report of Task Group (TG) 302, which provides recommendations on acceptance, commissioning, and ongoing routine quality assurance (QA) for surface-guided radiation therapy (SGRT) systems. One of the recommended monthly QA tests is a dynamic localization accuracy test. This work aimed to develop an automated procedure for monthly SGRT dynamic localization QA. An anthropomorphic head phantom was rigidly attached to the 6-dof couch of a TrueBeam linac. TrueBeam Developer Mode was used to take an MV image of the phantom at the starting position, then automatically drive the couch through a series of translations and rotations, taking an MV image after each translation. The Identify SGRT system monitored the motion of the phantom surface from the starting position. Translations assessed on MV images were compared to translations reported in trajectory log files and Identify log files. Rotations were compared between trajectory log files and Identify log files. Three experiments were conducted. None of the translations or rotations from any experiment exceeded the tolerance values for stereotactic ablative body radiation therapy (SABR) recommended by AAPM TG-142. Maximum deviations from the expected translation values from MV imaging, trajectory log files, and Identify log files were -0.94mm, -0.11mm, and -0.78mm, respectively. Maximum deviations from the expected rotation values from trajectory log files and Identify log files were 0.01 and -0.2 degrees, respectively. The proposed method is a simple automated way to complete monthly dynamic localization QA of SGRT systems.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] GEOMETRIC ACCURACY OF IMAGING SYSTEMS ON TRILOGY MX USING AN AUTOMATED GEOMETRIC TEST TOOL
    Munro, P.
    Jeung, A.
    Graf, A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S433 - S434
  • [42] TEST-RESULTS FOR AUTOMATED OPTICAL-SURFACE FINISHING SYSTEM
    BESETSKII, VA
    VITRICHENKO, EA
    GROSHEV, KV
    KIRYAKOV, KR
    MAMONOV, SK
    NIKOLSKII, YV
    PROKHOROV, AM
    TRUSHIN, EV
    CHARIKOV, AV
    SOVIET JOURNAL OF OPTICAL TECHNOLOGY, 1982, 49 (02): : 103 - 105
  • [43] Geometric Accuracy of Imaging Systems on Trilogy MX Using an Automated Geometric Test Tool
    Jeung, A.
    Graf, A.
    Suri, R.
    Munro, P.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [44] QUANTITATIVE DCE-CT IMAGING QUALITY ASSURANCE WITH A NOVEL DYNAMIC FLOW PHANTOM
    Driscoll, B.
    Keller, H.
    Coolens, C.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S178 - S179
  • [45] Quantitative DCE-CT Imaging Quality Assurance with a Novel Dynamic Flow Phantom
    Driscoll, B.
    Coolens, C.
    Keller, H.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [46] Application of a video-optical beam imaging system for quality assurance of medical accelerators
    Ma, LJ
    Boyer, AL
    Findley, DO
    Geis, PB
    Mok, E
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12): : 3649 - 3659
  • [47] Quality assurance measurements of geometric accuracy for magnetic resonance imaging-based radiotherapy treatment planning
    Ranta, Iiro
    Kemppainen, Reko
    Keyrilainen, Jani
    Suilamo, Sami
    Heikkinen, Samuli
    Kapanen, Mika
    Saunavaara, Jani
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 62 : 47 - 52
  • [48] Rapid Prototyping of Vascular Trees for Quality Assurance of Dynamic Contrast Enhanced Perfusion Imaging and Analysis
    Thomas, A.
    Balter, J.
    Hollister, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [49] Dynamic-contrast-enhanced MR imaging: a phantom for offline parameter optimization and quality assurance
    Verhaart, R. F.
    Pellicer, R.
    Van der Lugt, A.
    Paulides, M. M.
    Kotek, G.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S32 - S33
  • [50] Quality assurance for dynamic multileaf collimator modulated fields using a fast beam imaging system
    Ma, LJ
    Geis, PB
    Boyer, AL
    MEDICAL PHYSICS, 1997, 24 (08) : 1213 - 1220