Position Sensitive Detector Based on Microchannel Plates for Measuring the Characteristics of Fission Fragments at the MAVR Facility

被引:0
|
作者
Aznabaev, D. [1 ,2 ]
Isataev, T. [1 ,2 ]
Lukyanov, S. M. [1 ]
Smirnov, V. I. [1 ]
Stukalov, S. S. [1 ]
Solodov, A. N. [1 ]
机构
[1] Joint Inst Nucl Res, Dubna 141980, Moscow Oblast, Russia
[2] Inst Nucl Phys, Alma Ata 050032, Kazakhstan
关键词
Charged particles - Fission reactions - Microchannels - Nuclear reactions;
D O I
10.1134/S0020441224700027
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper presents the results of measurements of fission fragments using a time-of-flight system with a position-sensitive detector based on microchannel plates designed to measure the masses of nuclear reaction products at the MAVR facility. The advantage of such MCPs is their high positional sensitivity and high efficiency of registration of heavy charged particles with low energy. The paper presents a technical description of the fission fragments registration system and the results of measuring the parameters of a position-sensitive detector based on large-area microchannel plates (40 x 60 mm(2)) obtained by measuring Cf-252 spontaneous fission fragments. The coordinate resolution along the X and Y axes was determined when fission fragments were registered in the N-14+Au-192 reaction at the MAVR installation. The time resolution of the PSD was obtained by registering the time of flight of alpha particles emitted by the radioactive source Ra-226.
引用
收藏
页码:20 / 27
页数:8
相关论文
共 50 条
  • [31] HIGH-RESOLUTION POSITION-SENSITIVE DETECTORS USING MICROCHANNEL PLATES
    PARKES, W
    EVANS, KD
    MATHIESON, E
    NUCLEAR INSTRUMENTS & METHODS, 1974, 121 (01): : 151 - 159
  • [32] Hard x-ray detector based on microchannel plates
    A. F. Ioffe Physical-Technical Inst, St. Petersburg, Russia
    Rev Sci Instrum, 3 pt 1 (700-703):
  • [33] A novel digital mammography detector concept based on microchannel plates
    Shikhaliev, PM
    Xu, T
    Le, H
    Molloi, S
    MEDICAL PHYSICS, 2003, 30 (06) : 1455 - 1456
  • [34] Hard x-ray detector based on microchannel plates
    Shikhaliev, PM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (03): : 700 - 703
  • [35] NEW ALPHA-PARTICLE DETECTOR BASED ON MICROCHANNEL PLATES
    FROLOV, EA
    KONSTANTINOV, AA
    YUDIN, MF
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1983, 26 (04) : 813 - 814
  • [36] ELECTRON DETECTOR FOR MOSSBAUER-SPECTROSCOPY BASED ON MICROCHANNEL PLATES
    AMULYAVICHYUS, AP
    DAVIDONIS, RY
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1986, 29 (03) : 590 - 593
  • [37] ABSOLUTE AND ANGULAR EFFICIENCIES OF A MICROCHANNEL-PLATE POSITION-SENSITIVE DETECTOR
    GAO, RS
    GIBNER, PS
    NEWMAN, JH
    SMITH, KA
    STEBBINGS, RF
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (11): : 1756 - 1759
  • [38] RESPONSE OF POSITION-SENSITIVE DETECTORS TO FISSION FRAGMENTS AND OTHER HEAVY IONS
    KAUFMAN, SB
    WILKINS, BD
    FLUSS, MJ
    STEINBERG, EP
    NUCLEAR INSTRUMENTS & METHODS, 1970, 82 : 117 - +
  • [39] RESPONSE OF POSITION-SENSITIVE DETECTORS TO FISSION FRAGMENTS AND OTHER HEAVY IONS
    KAUFMAN, SB
    WILKINS, BD
    STEINBER.EP
    FLUSS, MJ
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1969, (SEP): : NU6 - &
  • [40] Hard X-ray beam detector based on microchannel plates
    Shikhaliev, PM
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 369 (01): : 147 - 150