EMISSION CHARACTERISTICS OF A PULSED, RADIOFREQUENCY GLOW-DISCHARGE ATOMIC EMISSION DEVICE

被引:36
|
作者
WINCHESTER, MR [1 ]
MARCUS, RK [1 ]
机构
[1] CLEMSON UNIV,DEPT CHEM,HOWARD L HUNTER CHEM LABS,CLEMSON,SC 29634
关键词
D O I
10.1021/ac00042a008
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The instrumentation, initial observations, and operating characteristics of a pulsed radio-frequency glow discharge atomic emission source are described. Anomalies in the temporal emission intensity wave forms for some analyte transitions are also reported. These anomalles take the form of a well-defined intensity maximum located either near the beginning of the discharge pulse or just after power termination, depending on the particular transition. Still other analyte transitions demonstrate no such temporal irregularities. Selective, gated, detection of these emission anomalies suggests possible analytical advantages in terms of instantaneous emission intensities, which may translate into improved analytical sensitivities.