Intraprosthetic fixation techniques in the treatment of periprosthetic fractures: A biomechanical study

被引:4
|
作者
Brand, Stephan [1 ]
Klotz, Johannes [2 ]
Hassel, Thomas [2 ]
Petri, Maximilian [1 ]
Haasper, Carl [1 ]
Bach, Friedrich-Wilhelm [2 ]
Krettek, Christian [1 ]
Goesling, Thomas [1 ]
机构
[1] Med Sch Hannover MHH, Dept Trauma, Carl Neuberg St 1-OE 6230, D-30625 Hannover, Germany
[2] Leibniz Univ Hannover, Dept Mat Sci, D-30625 Hannover, Germany
来源
WORLD JOURNAL OF ORTHOPEDICS | 2012年 / 3卷 / 10期
关键词
Periprosthetic fracture; Less invasive stabilization system; Plate fixation; Intraprosthetic screw fixation; Material science; Biomechanical testing; Axial load-to-failure;
D O I
10.5312/wjo.v3.i10.162
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
AIM: To develop new fixation techniques for the treatment of periprosthetic fractures using intraprosthetic screw fixation with inserted threaded liners. METHODS: A Vancouver B1 periprosthetic fracture was simulated in femur prosthesis constructs using sawbones and cemented regular straight hip stems. Fixation was then performed with either unicortical locked-screw plating using the less invasive stabilization system-plate or with intraprosthetic screw fixation using inserted liners. Two experimental groups were formed using either prostheses made of titanium alloy or prostheses made of cobalt chrome alloy. Fixation stability was compared in an axial load-to-failure model. Drilling was performed using a specially invented prosthesis drill with constantly applied internal cooling. RESULTS: The intraprosthetic fixation model with tita-nium prostheses was superior to the unicortical lockeds-crew fixation in all tested devices. The intraprosthetic fixation model required 10 456 N +/- 1892 N for failure and the unicortical locked-screw plating required 7649 N +/- 653 N (P < 0.05). There was no significant difference between the second experimental group and the control group. CONCLUSION: Intraprosthetic screw anchorage with special threaded liners enhances the primary stability in treating periprosthetic fractures by internal fixation. (C) 2012 Baishideng. All rights reserved.
引用
收藏
页码:162 / 166
页数:5
相关论文
共 50 条
  • [31] A BIOMECHANICAL STUDY OF ODONTOID FRACTURES AND FRACTURE FIXATION
    DOHERTY, BJ
    HEGGENESS, MH
    ESSES, SI
    SPINE, 1993, 18 (02) : 178 - 184
  • [32] Fixation of segmental subtrochanteric fractures - A biomechanical study
    Kraemer, WJ
    Hearn, TC
    Powell, JN
    Mahomed, N
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (332) : 71 - 79
  • [33] Mennen plate fixation for the treatment of periprosthetic femoral fractures - Reply
    Noorda, RJP
    Wuisman, PIJM
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (11): : 2249 - 2249
  • [34] Treatment of Periprosthetic Humerus Fractures With Open Reduction and Internal Fixation
    Kurowicki, Jennifer
    Momoh, Enesi
    Levy, Jonathan C.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2016, 30 (11) : e369 - e374
  • [35] Biomechanical evaluation of four different posterior screw and rod fixation techniques for the treatment of the odontoid fractures
    Li, Lei
    Liu, Wen-Fei
    Jiang, Hong-Kun
    Li, Yun-Peng
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (06): : 8571 - 8580
  • [36] A biomechanical study comparing cortical onlay allograft struts and plates in the treatment of periprosthetic femoral fractures
    Wilson, D
    Frei, H
    Masri, BA
    Oxland, TR
    Duncan, CP
    CLINICAL BIOMECHANICS, 2005, 20 (01) : 70 - 76
  • [37] Biomechanical Evaluation of Osteoporotic Proximal Periprosthetic Femur Fractures With Proximal Bicortical Fixation and Allograft Struts
    O'Connell, Robert S.
    Owen, John R.
    Hansen, Erik J.
    Bashir, Azhar S.
    Wayne, Jennifer S.
    Satpathy, Jibanananda
    Kates, Stephen L.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (10) : 508 - 514
  • [38] Periprosthetic supracondylar femoral fractures following knee arthroplasty: a biomechanical comparison of four methods of fixation
    Tatu J. Mäkinen
    Herman S. Dhotar
    Simcha G. Fichman
    Matthew J. Gunton
    Mitchell Woodside
    Oleg Safir
    David Backstein
    Thomas L. Willett
    Paul R. T. Kuzyk
    International Orthopaedics, 2015, 39 : 1737 - 1742
  • [39] Biomechanical evaluation of periprosthetic femoral fracture fixation
    Zdero, Rad
    Walker, Richard
    Waddell, James P.
    Schemitsch, Emil H.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (05): : 1068 - 1077
  • [40] Tangential Bicortical Locked Fixation Improves Stability in Vancouver B1 Periprosthetic Femur Fractures: A Biomechanical Study
    Lewis, Gregory S.
    Caroom, Cyrus T.
    Wee, Hwabok
    Jurgensmeier, Darin
    Rothermel, Shane D.
    Bramer, Michelle A.
    Reid, John Spence
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (10) : e364 - e370