Effects of Eccentric Impeller Position on Radial Passive Stability in a Magnetically Levitated Centrifugal Blood Pump with a Double Volute

被引:9
|
作者
Shida, Shuya [1 ,3 ]
Masuzawa, Toru [2 ]
Osa, Masahiro [2 ]
Sato, Ryota [1 ]
机构
[1] Ibaraki Univ, Grad Sch Sci & Engn, Hitachi, Ibaraki, Japan
[2] Ibaraki Univ, Dept Mech Engn, Hitachi, Ibaraki, Japan
[3] 4-12-1 Nakanarusawa, Hitachi, Ibaraki 3168511, Japan
关键词
eccentric impeller position; double volute; magnetic suspension; hydraulic force; computational fluid dynamics (CFD);
D O I
10.14326/abe.7.63
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Evaluation of the impeller radial stability is important from the bioengineering point of view in the development of mechanical circulatory support devices (MCSDs) for safer use as bridge for several months or destination therapy for years. In this study, radial stability of a magnetically levitated impeller in a centrifugal blood pump with an axially magnetic suspension system was evaluated by investigating the effects of the eccentric impeller position on passive stability, aiming to propose a pump design guide for the development and safer clinical use of MCSDs. First, impeller displacements in the prototype pump were measured using a mock loop together with laser displacement sensors. Then, the radial hydraulic forces exerted on an eccentric impeller were calculated using computational fluid dynamics (CFD) analysis for four volute-casing geometries. In addition, hemocompatibility was assessed using CFD calculations of scalar shear stress exerted on blood. Measurement of impeller displacement showed that the displacement varied from 0.56 to 0.27 mm at a rotational speed of 1800 rpm as the flow rate increased from 0 to 6.5 L/min. In the CFD calculation, the radial hydraulic force increased linearly from 0.2 to 1.7 N as the impeller displacement increased from 0 to 0.5 mm for all the double volute geometries, under conditions of a rotational speed of 1800 rpm and flow rates of 3, 5 and 7 L/min. These results indicate that the impeller stability in the prototype pump is acceptable at the operation conditions of ventricular assist devices, because the magnetic bearing stiffness of radial component was 4.1 N/mm. In the pressure recovery analysis of eccentric impellers, a double volute was not effective because of the unbalanced pressure field generated by the unbalanced pressure recovery. Thus, the increase in radial hydraulic force associated with an eccentric impeller could not be avoided by changing the conventional double volute design. The CFD analysis of geometrical variation indicated that widening of the radial clearance is an effective approach to improve the radial stability as well as hemocompatibility, although the radial clearance should be designed based on trade-offs among impeller stability, hemocompatibility and pump performance.
引用
收藏
页码:63 / 71
页数:9
相关论文
共 50 条
  • [21] Disposable magnetically levitated centrifugal blood pump: Design and in vitro performance
    Hoshi, H
    Asama, J
    Shinshi, T
    Ohuchi, K
    Nakamura, M
    Mizuno, T
    Arai, H
    Shimokohbe, A
    Takatani, S
    ARTIFICIAL ORGANS, 2005, 29 (07) : 520 - 526
  • [22] HeartQuest ventricular assist device magnetically levitated centrifugal blood pump
    Bearnson, GB
    Jacobs, GB
    Kirk, J
    Khanwilkar, PS
    Nelson, KE
    Long, JW
    ARTIFICIAL ORGANS, 2006, 30 (05) : 339 - 346
  • [23] A new design for a compact centrifugal blood pump with a magnetically levitated rotor
    Asama, J
    Shinshi, T
    Hoshi, H
    Takatani, S
    Shimokohbe, A
    ASAIO JOURNAL, 2004, 50 (06) : 550 - 556
  • [24] Design Optimization of the Impeller and Volute of a Centrifugal Pump to Improve the Hydraulic Performance and Flow Stability
    Shim, Hyeon-Seok
    Kim, Kwang-Yong
    JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 2020, 142 (10):
  • [25] A magnetically levitated centrifugal blood pump with a simple-structured disposable pump head
    Hijikata, Wataru
    Shinshi, Tadahiko
    Asama, Junichi
    Li, Lichuan
    Hoshi, Hideo
    Takatani, Setsuo
    Shimokohbe, Akira
    ARTIFICIAL ORGANS, 2008, 32 (07) : 531 - 540
  • [26] Flow Characteristics in Volute of a Double-Suction Centrifugal Pump with Different Impeller Arrangements
    Song, Yu
    Fan, Honggang
    Zhang, Wei
    Xie, Zhifeng
    ENERGIES, 2019, 12 (04)
  • [27] The Effects of Ambulatory Accelerations on the Stability of a Magnetically Suspended Impeller for an Implantable Blood Pump
    Paul, Gordon
    Rezaienia, Mohammed Amin
    Rahideh, Akbar
    Munjiza, Ante
    Korakianitis, Theodosios
    ARTIFICIAL ORGANS, 2016, 40 (09) : 867 - 876
  • [28] Analysis of radial forces in centrifugal pump with single and double volute based on FLUENT
    Technical and Research Center of Fluid Machinery Engineering, Jiangsu University, Zhenjiang, Jiangsu 212013, China
    Paiguan Jixie, 2009, 2 (83-86):
  • [29] A cost-effective extracorporeal magnetically-levitated centrifugal blood pump employing a disposable magnet-free impeller
    Hijikata, W.
    Mamiya, T.
    Shinshi, T.
    Takatani, S.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H12) : 1149 - 1157
  • [30] NUMERICAL INVESTIGATION ON IMPELLER-VOLUTE INTERACTION IN THE CENTRIFUGAL PUMP WITH RADIAL GAP AND TONGUE PROFILE VARIATION
    Guo Pengcheng
    Luo Xingqi
    Liao Weili
    Zhu Guojun
    PROCEEDINGS OF THE ASME FLUIDS ENGINEERING DIVISION SUMMER CONFERENCE - 2008, VOL 2, 2009, : 217 - 224