A LABORATORY EXAFS SPECTROMETER IN TRANSMISSION DISPERSIVE MODE

被引:14
|
作者
LECANTE, P [1 ]
JAUD, J [1 ]
MOSSET, A [1 ]
GALY, J [1 ]
BURIAN, A [1 ]
机构
[1] POLISH ACAD SCI,PL-41800 ZABRZE,POLAND
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1994年 / 65卷 / 04期
关键词
D O I
10.1063/1.1144909
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A laboratory spectrometer designed for routine extended x-ray absorption fine structure (EXAFS) measurements in dispersive mode is described. Factors determining energy range and resolution are detailed. An empirical energy calibration procedure is proposed. EXAFS spectra obtained for copper and zirconium metal foil and molecular complex ZrOCl2 compare favorably with data obtained using a conventional synchrotron facility.
引用
收藏
页码:845 / 849
页数:5
相关论文
共 50 条
  • [31] Wavelength dispersive spectrometer and energy dispersive spectrometer automation: Past and future development
    McCarthy, JJ
    Friel, JJ
    MICROSCOPY AND MICROANALYSIS, 2001, 7 (02) : 150 - 158
  • [32] Development of an energy dispersive spectrometer for a transmission electron microscope utilizing a TES microcalorimeter array
    Tanaka, Keiichi
    Mitsuda, Kazuhisa
    Hara, Toni
    Maehata, Keisuke
    Yamasaki, Noriko Y.
    Odawara, Akikazu
    Nagata, Atsushi
    Watanabe, Katsuaki
    Takei, Yoh
    LOW TEMPERATURE DETECTORS LTD 13, 2009, 1185 : 715 - +
  • [33] STUDIES OF COXMG1-XO SOLID-SOLUTIONS USING LABORATORY EXAFS-SPECTROMETER
    KUZMIN, A
    PURANS, J
    SAZONOV, A
    MIRONOVA, N
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1993, 32 : 637 - 639
  • [34] PDA SYSTEM FOR ENERGY DISPERSIVE EXAFS.
    Allinson, N.M.
    Baker, G.
    Greaves, G.N.
    Nicoll, J.K.
    Nuclear instruments and methods in physics research, 1987, A266 (1-3): : 592 - 597
  • [35] DOUBLY MULTIPLEXING DISPERSIVE SPECTROMETER
    PHILLIPS, PG
    HARWIT, M
    APPLIED OPTICS, 1971, 10 (12): : 2780 - &
  • [36] Experience of 3 types of commercially available photodiode array in the daresbury laboratory energy dispersive EXAFS detector system
    Salvini, G
    Dent, AJ
    Farrow, RC
    Derbyshire, GE
    Bogg, D
    JOURNAL DE PHYSIQUE IV, 1997, 7 (C2): : 319 - 320
  • [37] Time-resolved extended x-ray absorption fine structure (EXAFS) studies by means of an energy dispersive spectrometer
    Oulianov, DA
    Tomov, IV
    Lin, SH
    Rentzepis, PM
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2001, 48 (02) : 127 - 132
  • [38] DEVELOPMENT OF A LABORATORY EXAFS FACILITY
    KNAPP, GS
    CHEN, H
    KLIPPERT, TE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1978, 49 (12): : 1658 - 1666
  • [39] LABORATORY EXAFS BY FLUORESCENCE DETECTION
    TOHJI, K
    MIZUSHIMA, T
    UDAGAWA, Y
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1990, 29 (10): : 2171 - 2175
  • [40] DEVELOPMENT OF A LABORATORY EXAFS FACILITY
    TOHJI, K
    UDAGAWA, Y
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1983, 22 (05): : 882 - 885