A Monte Carlo simulation of coincidence detection and imaging of gamma-ray cascades with a scintillation camera

被引:4
|
作者
Pahlka, R. Benton [1 ]
Kappadath, S. Cheenu [1 ]
Mawlawi, Osama R. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, UNIT 1472, POB 301402, Houston, TX 77230 USA
来源
关键词
gamma-ray cascades; coincidence detection; gamma camera;
D O I
10.1088/2057-1976/aad572
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Several radionuclei used in nuclear medicine decay through the emission of cascaded gamma-rays. These successively-emitted gamma-rays could potentially be detected in coincidence and then be used to create coincident event-based images. While several groups have considered various ways to leverage the gamma-ray cascade to produce images, the concept of producing images using cascaded gamma rays in coincidence with a dual-head gamma camera has not been previously explored in detail. Given that current dual-head gamma cameras cannot be configured in coincidence mode, the purpose of this work was to explore this concept with In-111 using a Monte Carlo simulation. First, GEANT4 was used to model a dual-head gamma camera and to simulate the decay of the nucleus. The model was then validated using planar data from a physical scanner. The validated model was then used to simulate a point source of In-111 located at the camera isocenter to evaluate the true coincidence efficiency with the detectors oriented at 90 degrees. The noise-equivalent-count-rate (NECR) as a function of timing window resolution was computed for different source activities. A simple reconstruction algorithm was then used to produce coincidence-based images from sources in various configurations. The coincident gamma-ray detection efficiency for a single point source in air with the detector heads oriented at 90 degrees was found to be 6 x 10(-8), independent of the source-to-detector distance. The time window resolution for the maximum NECR ranged between 200 and 500 ns, dependent on source activity. The algorithm could produce images over a range of timing window resolutions, with varying levels of reconstruction artifacts. Images produced from a small acrylic phantom showed that scattered events produced small degradations in resolution and contrast-to-noise ratios. This work described the conditions under which images based on coincident events from In-111 could be created using a dual head gamma camera, assuming appropriate electronics modifications.
引用
收藏
页数:18
相关论文
共 50 条
  • [31] CAPTURE GAMMA-RAY STUDIES USING A SCINTILLATION COINCIDENCE SPECTROMETER
    RECKSIEDLER, A
    HAMERMESH, B
    PHYSICAL REVIEW, 1954, 95 (02): : 650 - 650
  • [32] MONTE CARLO CALCULATION OF GAMMA-RAY DOSE
    BROWNELL, GL
    PHYSICS IN MEDICINE AND BIOLOGY, 1966, 11 (01): : 160 - &
  • [33] A double photon coincidence detection method for medical gamma-ray imaging
    Uenomachi, Mizuki
    Shimazoe, Kenji
    Takahashi, Hiroyuki
    BIO-ALGORITHMS AND MED-SYSTEMS, 2022, 18 (01) : 120 - 126
  • [34] MONTE CARLO CALCULATIONS OF GAMMA-RAY DOSE
    BROWNELL, GL
    REDDY, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1966, 11 (04): : 616 - &
  • [35] Simulation of Intrinsic Resolution of Scintillation Camera in Monte Carlo Environment
    Dong, Xianling
    Saad, Wira Hidayat Mohd
    Adnan, Wan Azizun Wan
    Hashim, Suhairul
    Hassan, Nor Pai'za Mohd.
    Nordin, Abd Jalil
    Saripan, M. Iqbal
    2013 IEEE INTERNATIONAL CONFERENCE ON SIGNAL AND IMAGE PROCESSING APPLICATIONS (IEEE ICSIPA 2013), 2013, : 11 - 14
  • [36] A MONTE-CARLO PROGRAM FOR THE SIMULATION OF SCINTILLATION CAMERA CHARACTERISTICS
    LJUNGBERG, M
    STRAND, SE
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1989, 29 (04) : 257 - 272
  • [37] Monte Carlo simulation of gamma-ray total counting efficiency for a Phoswich detector
    Yalcin, S.
    Gurler, O.
    Gundogdu, O.
    Kaynak, G.
    RADIATION MEASUREMENTS, 2009, 44 (01) : 80 - 85
  • [38] MONTE-CARLO SIMULATION OF GAMMA-RAY BACKSCATTER SOIL DENSITY GAUGES
    HOPKINS, WC
    GARDNER, RP
    NUCLEAR ENGINEERING AND DESIGN, 1973, 24 (03) : 332 - 343
  • [39] Monte Carlo simulation of induced radioactive background in gamma-ray detector materials
    Truscott, PR
    Evans, HE
    Dyer, CS
    Cosby, M
    Moss, CE
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 41 - 45
  • [40] Monte Carlo simulation of the Galactic 26Al gamma-ray map
    Lentz, EJ
    Branch, D
    Baron, E
    ASTROPHYSICAL JOURNAL, 1999, 512 (02): : 678 - 682