TOTALLY IMPLANTABLE TOTAL ARTIFICIAL-HEART AND VENTRICULAR ASSIST DEVICE WITH MULTIPURPOSE MINIATURE ELECTROMECHANICAL ENERGY SYSTEM

被引:13
|
作者
TAKATANI, S
ORIME, Y
TASAI, K
OHARA, Y
NAITO, K
MIZUGUCHI, K
MAKINOUCHI, K
DAMM, G
GLUECK, J
LING, J
NOON, G
NOSE, Y
机构
[1] Department of Surgery, Baylor College of Medicine, Houston, Texas
关键词
TOTALLY IMPLANTABLE TOTAL ARTIFICIAL HEART; TOTALLY IMPLANTABLE VENTRICULAR ASSIST DEVICE; MINIATURE ELECTROMECHANICAL ENERGY SYSTEM; HYBRID CONTROLLER; EMERGENCY DRIVE PORT;
D O I
10.1111/j.1525-1594.1994.tb03301.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A multipurpose miniature electromechanical energy system has been developed to yield a compact, efficient, durable, and biocompatible total artificial heart (TAH) and ventricular assist device (VAD). Associated controller-driver electronics were recently miniaturized and converted into hybrid circuits. The hybrid controller consists of a microprocessor and controller, motor driver, Hall sensor, and commutation circuit hybrids. The sizing study demonstrated that all these components can be incorporated in the pumping unit of the TAH and VAD, particularly in the centerpiece of the TAH and the motor housing of the VAD. Both TAH and VAD pumping units will start when their power line is connected to either the internal power pack or the external battery unit. As a redundant driving and diagnostic port, an emergency port was newly added and will be placed in subcutaneous location. In case of system failure, the skin will be cut down, and an external motor drive or a pneumatic driver will be connected to this port to run the TAH. This will minimize the circulatory arrest time. Overall efficiency of the TAH without the transcutaneous energy transmission system was 14-18% to deliver pump outputs of 4-9 L/min against the right and left afterload pressures of 25 and 100 mm Hg. The internal power requirement ranged from 6 to 13 W. The rechargeable batteries such as NiCd or NiMH with 1 AH capacity can run the TAH for 30-45 min. The external power requirement, when TETS efficiency of 75% was assumed, ranged from 8 to 18 W. The accelerated endurance test in the 42-degrees-C saline bath demonstrated stable performance over 4 months. Long-term endurance and chronic animal studies will continue toward a system with 5 years durability by the year 2000.
引用
收藏
页码:80 / 92
页数:13
相关论文
共 50 条
  • [21] STAGED CARDIAC TRANSPLANTATION - TOTAL ARTIFICIAL-HEART OR VENTRICULAR-ASSIST PUMP
    PAE, WE
    PIERCE, WS
    MYERS, JL
    WISMAN, CB
    CAMPBELL, DB
    WALDHAUSEN, JA
    CIRCULATION, 1988, 78 (05) : 66 - 72
  • [22] One piece ultracompact totally implantable electromechanical total artificial heart for permanent use
    Takatani, S
    Sakamoto, T
    Ohuchi, K
    Nakamura, M
    Mizuno, T
    Arai, H
    ASAIO JOURNAL, 2002, 48 (05) : 538 - 545
  • [23] ARTIFICIAL-HEART AND LEFT-VENTRICULAR ASSIST DEVICES
    KOLFF, J
    DEEB, GM
    SURGICAL CLINICS OF NORTH AMERICA, 1985, 65 (03) : 661 - 677
  • [24] LEFT-VENTRICULAR ASSIST PUMPS AND THE ARTIFICIAL-HEART
    PIERCE, WS
    MYERS, JL
    JOURNAL OF CARDIOVASCULAR MEDICINE, 1981, 6 (07): : 667 - &
  • [25] TOTAL ARTIFICIAL-HEART AND ASSIST DEVICES AS A BRIDGE TO TRANSPLANTATION
    ROKITANSKY, A
    WOLNER, E
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1989, 12 (02): : 77 - 84
  • [26] EXPERIMENTAL RESULTS FOR CHRONIC LEFT-VENTRICULAR ASSIST AND TOTAL ARTIFICIAL-HEART DEVELOPMENT
    NOSE, Y
    JACOBS, G
    KIRALY, RJ
    GOLDING, L
    HARASAKI, H
    TAKATANI, S
    MURABAYASHI, S
    SUKALAC, RW
    KAMBIC, H
    SNOW, J
    ARTIFICIAL ORGANS, 1983, 7 (01) : 55 - 63
  • [27] The totally implantable Novacor Left Ventricular Assist System
    Robbins, RC
    Kown, MH
    Portner, PM
    Oyer, PE
    ANNALS OF THORACIC SURGERY, 2001, 71 (03): : S162 - S165
  • [28] Gastrointestinal Bleeding in Ventricular Assist Device and Total Artificial Heart Implantations
    Lachar, Jatinder
    Christopher, Sanah
    Flattery, Maureen
    Shah, Keyur
    Tang, Daniel G.
    Smallfield, Melissa C.
    Cooke, Richard H.
    Kasirajan, Vigneshwar
    Cheng, Jianfeng
    Smallfield, George
    GASTROENTEROLOGY, 2014, 146 (05) : S322 - S322
  • [29] Totally Implantable total artificial heart for clinical application
    Nosé, Y
    ARTIFICIAL ORGANS, 2002, 26 (03) : 214 - 215
  • [30] Treatment of multiple organ failure by a totally implantable ventricular assist device
    Yambe T.
    Owada N.
    Kobayashi S.-I.
    Nanka S.-S.
    Tanaka A.
    Yoshizawa M.
    Abe K.-I.
    Tabayashi K.
    Takeda H.
    Hashimoto H.
    Nitta S.-I.
    Journal of Artificial Organs, 2001, 4 (3) : 252 - 256