Correlation of Cord Blood pH, Base Excess, and Lactate Concentration Measured With a Portable Device for Identifying Fetal Acidosis

被引:9
|
作者
Labrecque, Lyne [1 ]
Provencal, Mathieu [1 ]
Caqueret, Aurore [1 ]
Wo, Bi Lan [2 ]
Bujold, Emmanuel [3 ]
Lariviere, Francois [1 ]
Bedard, Marie-Josee [2 ]
机构
[1] Ctr Hosp Univ Montreal, Dept Biochem, Montreal, PQ, Canada
[2] Ctr Hosp Univ Montreal, Dept Obstet & Gynecol, Montreal, PQ, Canada
[3] Univ Laval, Fac Med, Dept Obstet & Gynecol, Quebec City, PQ, Canada
关键词
Lactate; fetal scalp; portable device; point-of-care testing;
D O I
10.1016/S1701-2163(15)30539-9
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective: To determine the effectiveness of portable lactate analyzers in identifying fetal acidosis by correlating arterial and venous lactate values from umbilical cord blood with lactate, pH, and base excess measurements from central laboratory analyzers. Methods: We performed a prospective study using arterial and venous cord blood from 52 women with a singleton fetus delivered at term. We evaluated the correlation between the cord blood lactate concentration measured using two of the same portable devices (Lactate Plus, Nova Biomedical) with the result from a central laboratory analyzer. Analyses of the correlation between arterial lactate concentration measured on the portable device with arterial pH and base excess were then performed. Results: We observed a median arterial pH of 7.24 (range 7.05 to 7.35) and a median arterial lactate concentration of 3.7 mmol/L (range 1.7 to 8.8 mmol/L). An excellent correlation was observed between lactate concentrations measured by the two portable devices (arterial R-2 = 0.98 and venous R-2 = 0.98), and between the portable device and the central laboratory analyzer (arterial R-2 = 0.94 and venous R-2 = 0.95). In our population, the optimal cut-offs to predict a pH < 7.20 or a base excess > -8.0 mmol/L were a lactate concentration of 4.9 mmol/L and 5.3 mmol/L, respectively, according to receiver operator characteristic analysis. With a lactate concentration > 4.9 mmol/L, the portable device had a sensitivity of 82% and a specificity of 90% to identify samples with an arterial pH < 7.20. Conclusion: Cord blood lactate concentration measured with a portable device is a good predictor of cord blood base excess and pH. Future studies should be designed to correlate scalp blood lactate measurements with clinical outcomes.
引用
收藏
页码:598 / 604
页数:7
相关论文
共 30 条
  • [21] Blood lactate, pH, base excess and pCO2 as prognostic indicators in caesarean-born kids from goats with pregnancy toxaemia
    I. M. Andrade
    P. B. A. Simões
    L. P. Lamas
    N. Carolino
    M. S. Lima
    Irish Veterinary Journal, 72
  • [22] Blood lactate, pH, base excess and pCO2 as prognostic indicators in caesarean-born kids from goats with pregnancy toxaemia
    Andrade, I. M.
    Simoes, P. B. A.
    Lamas, L. P.
    Carolino, N.
    Lima, M. S.
    IRISH VETERINARY JOURNAL, 2019, 72 (01)
  • [23] FETAL UMBILICAL CORD BLOOD ERYTHROPOIETIN, INTERLEUKIN-6, pH, pO2, pCO2, AND BASE EXCESS LEVELS IN HISTOLOGIC AND/OR CLINICAL CHORIOAMNIONITIS: IS THE RESPONSE ONLY INFLAMMATORY?
    Spellacy, William N.
    McCarthy, Joan M.
    Tsibris, John C. M.
    Gilbert-Barness, Enid
    Downes, Katheryne L.
    FETAL AND PEDIATRIC PATHOLOGY, 2009, 28 (05) : 239 - 246
  • [24] CORRELATION BETWEEN FETAL PH, CORD BLOOD-GLUCOSE LEVEL AND APGAR SCORE IN A FETAL INTENSIVE-CARE UNIT - PRELIMINARY-REPORT
    SOGBANMU, MO
    AJANI, B
    AFRICAN JOURNAL OF MEDICINE AND MEDICAL SCIENCES, 1978, 7 (01): : 17 - 21
  • [25] The correlation between total deceleration and acceleration surface areas on electronic fetal monitoring and neonatal cord blood pH in postdate pregnancies
    Talmor, Meital
    Rotem, Reut
    Wientraub, Adi Y. Y.
    Salem, Shimrit Yaniv
    INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2023, 161 (03) : 870 - 876
  • [26] Relation between umbilical cord blood pH, base deficit, lactate, 5-minute Apgar score and development of hypoxic ischemic encephalopathy
    Wiberg, Nana
    Kallen, Karin
    Herbst, Andreas
    Olofsson, Per
    ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 2010, 89 (10) : 1263 - 1269
  • [27] Correlation between total deceleration area and fetal cord blood pH in neonates complicated with meconium-stained amniotic fluid at term
    Loeub, Shirel Matmor
    Weintraub, Adi Y.
    Rotem, Reut
    Geva, Yael
    Salem, Shimrit Yaniv
    INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2022, 159 (03) : 974 - 978
  • [28] Identifying newborns with umbilical cord blood metabolic acidosis by intrapartum cardiotography combined with fetal ECG ST analysis (STAN): comparison of the new and old FIGO systems to classify cardiotocograms
    Olofsson, Per
    Noren, Hakan
    Carlsson, Ann
    Rosen, Karl G.
    JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2020, 33 (03): : 404 - 409
  • [29] The correlation between the concentration of selenium in the blood plasma of the women in labour with IUGR (Intrauterine growth retardation) and the newborns (Cord blood) and the flow parameters in the fetal-uteroplacental unit before the delivery (Doppler method)
    Bielanów, T
    Andrzejak, R
    Halama, P
    Janocha, A
    Chlosta, PL
    Maciejewski, M
    Fuchs, T
    PERINATOLOGY 2001, VOLS 1 AND 2, 2001, : 603 - 607
  • [30] Evaluation of plasma lactate concentration and base excess at the time of hospital admission as predictors of gastric necrosis and outcome and correlation between those variables in dogs with gastric dilatation-volvulus: 78 cases (2004-2009)
    Beer, Kari A. Santoro
    Syring, Rebecca S.
    Drobatz, Kenneth J.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2013, 242 (01): : 54 - 58