NEW DEVELOPMENT OF HGI2 X-RAY AND SOFT GAMMA-RAY DETECTOR FOR SYNCHROTRON RADIATION RESEARCH AND DOSIMETRY APPLICATIONS

被引:3
|
作者
GERVINO, G [1 ]
MONTICONE, E [1 ]
机构
[1] IST ELETTROTECNICO NAZL GALILEO FERRARIS,I-10127 TURIN,ITALY
关键词
D O I
10.1016/0924-4247(92)80028-2
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes recent progress in the development of HgI2 as an X-ray and soft gamma-ray detector for synchrotron radiation research and dosimetry application. The high efficiency of the detector depends on the photon absorption inside the crystal and on the free charge carrier collection at the electrodes under the influence of the applied bias. The dominant mechanisms determining the collection of charge carriers, produced by incident ionizing radiations, are the trapping-detrapping phenomena and the carrier drift mobility. The relative role of these two mechanisms, photon absorption and carrier collection, is related to the thickness of the HgI2 crystal. The analysis of trapping-detrapping phenomena, which are strongly related to the detector fabrication process, are discussed and the measurement of the carrier drift mobility, cutting the crystal parallel and perpendicular to the c-axis, is presented. The results of energy resolution measurements obtained exposing HgI2 detector prototypes of different thickness and with the electrodes realized by evaporizing different metals on the crystal surface, to X- and gamma-radiactive source, are presented and discussed. We have found, using a 900-mu-m-thick detector with 500 angstrom Pd electrodes evaporated on both surfaces (sandwich arrangement), an energy resolution equal to 600 eV at 5.9 keV X-ray (Mn-K-alpha) and 10.5 keV at 122 keV gamma-ray (Co-57).
引用
收藏
页码:455 / 463
页数:9
相关论文
共 50 条
  • [31] Textured α-HgI2 ceramics for sensitive X-ray detection
    Rault, Mathieu
    Binet, Laurent
    Gourier, Didier
    Wallez, Gilles
    Ponpon, Jean-Pierre
    Biava, Dominique
    Inglese, Jean-Marc
    Barboux, Philippe
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 808 : 35 - 40
  • [32] The long-term spectral stability of HgI2 gamma-ray detectors
    Vaccaro, FP
    van den Berg, L
    Szubart, LA
    Vigil, RD
    DeVito, RP
    Johnson, CR
    HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS III, 2001, 4507 : 108 - 118
  • [33] COSMIC X-RAY AND GAMMA-RAY LINE RADIATION
    FABIAN, A
    NATURE, 1977, 265 (5592) : 296 - 296
  • [34] Radiation Detectors for X-Ray and Gamma-Ray Spectroscopy
    G. F. Knoll
    Journal of Radioanalytical and Nuclear Chemistry, 2000, 243 : 125 - 131
  • [35] Radiation detectors for X-ray and gamma-ray spectroscopy
    Knoll, GF
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2000, 243 (01) : 125 - 131
  • [37] HIGH-EFFICIENCY CSI(TL)HGI2 GAMMA-RAY SPECTROMETERS
    WANG, YJ
    PATT, BE
    IWANCZYK, JS
    CHERRY, SR
    SHAO, Y
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) : 601 - 605
  • [38] X-RAY AND GAMMA-RAY STANDARDS FOR DETECTOR EFFICIENCY CALIBRATION
    NICHOLS, AL
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 286 (03): : 467 - 473
  • [39] THE EFFECT OF CHARGE TRAPPING ON THE SPECTROMETRIC PERFORMANCE OF HGI2 GAMMA-RAY DETECTORS
    IWANCZYK, JS
    SCHNEPPLE, WF
    MASTERSON, MJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03): : 421 - 426
  • [40] Soft X-ray diffractometer for synchrotron radiation
    Gau, TS
    Jean, YC
    Liu, KY
    Chung, CH
    Chen, CK
    Lai, SC
    Shu, CH
    Huang, YS
    Chao, CH
    Lee, YR
    Chen, CT
    Chang, SL
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 466 (03): : 569 - 575