摘要
The authors have demonstrated a practical method to transform the densitometer readings to a quantity (apparent dose) that linearises the dose-response curve for the Kodak XV-2 film for all conditions used (different energies, irradiation geometries and processing). The determination of the constants for the pair film-densitometer has to be done only once, using a set of conditions chosen as reference. The linearisation method proposed permits relative film dosimetry, in the same film, without any calibration. Beam profiles (electrons or photons) obtained in several depths in a single film (parallel geometry), can be used to obtain the isodose chart, for a given field size. Only central axis depth dose data for the particular field is additionally required. Treatment plan systems can usually use these data as input to generate isodose charts.
引用
收藏
页码:565 / 569
页数:5
相关论文
共 50 条
  • [41] Dose-Response Mixed Models for Repeated Measures – a New Method for Assessment of Dose-Response
    Gustaf J. Wellhagen
    Bengt Hamrén
    Maria C. Kjellsson
    Magnus Åstrand
    Pharmaceutical Research, 2020, 37
  • [42] EVALUATION OF THE RELATIVE DOSE-RESPONSE TEST FOR VITAMIN-A NUTRITURE IN CIRRHOTICS
    MOBARHAN, S
    RUSSELL, RM
    UNDERWOOD, BA
    WALLINGFORD, J
    MATHIESON, RD
    ALMIDANI, H
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1981, 34 (10): : 2264 - 2270
  • [43] Dose-Response Mixed Models for Repeated Measures - a New Method for Assessment of Dose-Response
    Wellhagen, Gustaf J.
    Hamren, Bengt
    Kjellsson, Maria C.
    Astrand, Magnus
    PHARMACEUTICAL RESEARCH, 2020, 37 (08)
  • [44] The use of distribution mixtures for dose-response modeling in toxicological experiments
    Razzaghi, M
    ENVIRONMETRICS, 2002, 13 (5-6) : 657 - 667
  • [45] Errors introduced by dose scaling for relative dosimetry
    Watanabe, Yoichi
    Hayashi, Naoki
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05): : 269 - 281
  • [46] Errors Introduced by Dose Scaling for Relative Dosimetry
    Watanabe, Y.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [47] Dose Variations Using an X-Ray Cabinet to Establish in vitro Dose-Response Curves for Biological Dosimetry Assays
    Bucher, Martin
    Weiss, Tina
    Endesfelder, David
    Trompier, Francois
    Ristic, Yoann
    Kunert, Patrizia
    Schlattl, Helmut
    Giussani, Augusto
    Oestreicher, Ursula
    FRONTIERS IN PUBLIC HEALTH, 2022, 10
  • [48] Redrawing the dose-response curve
    Renner, R
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2004, 38 (05) : 90A - 95A
  • [49] MAC AND DOSE-RESPONSE CURVES
    EGER, EI
    ANESTHESIOLOGY, 1971, 34 (02) : 202 - &
  • [50] Screening for allelopathy with dose-response?
    Betz, R
    Duke, SO
    Hurle, K
    12TH EWRS (EUROPEAN WEED RESEARCH SOCIETY) SYMPOSIUM 2002, WAGENINGEN, PROCEEDINGS, 2002, : 256 - 257