CRITICAL ANATOMIC DIMENSIONS FOR INTRATHORACIC CIRCULATORY ASSIST DEVICES

被引:16
|
作者
MUSSIVAND, T [1 ]
MASTERS, RG [1 ]
HENDRY, PJ [1 ]
RAJAGOPALAN, K [1 ]
WALLEY, VM [1 ]
NAHON, D [1 ]
HICKS, A [1 ]
KEON, WJ [1 ]
机构
[1] UNIV OTTAWA,OTTAWA CIVIL HOSP,INST HEART,OTTAWA K1N 6N5,ONTARIO,CANADA
关键词
MECHANICAL CIRCULATORY SUPPORT DEVICES; CHEST CAVITY DIMENSIONS; THORAX DIMENSIONS;
D O I
10.1111/j.1525-1594.1992.tb00310.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The design of totally implantable circulatory support devices must account for the anatomical constraints of the body cavity. To aid development of an implantable electrohydraulic ventricular assist device (EVAD), thoracic anatomical measurements and fit trials were performed. Various dimensions of the thoracic cavity were measured in 19 preserved and fresh cadavers. The mean left ventricle apex to chest wall distance was 2.7 +/- 0.2 cm. The mean sagittal radius of curvature at the fifth rib was 11.1 +/- 0.4 cm, and the transverse radius at the fifth rib was 9.4 +/- 0.3 cm. The mean sternal length was 18.9 +/- 1.2 cm. Using these measurements a model measuring 17.5 x 10.5 x 3.8 cm was designed. This model was placed in the chest of 11 patients undergoing open heart surgery. In one patient there was significant lung compression and in another hypotension suggesting great vessel compression. Finally, 1 of 11 patients had a short sternum and a small thorax, and the device did not fit. In conclusion, thoracic measurements may be used in the design of circulatory support devices. The model fit adequately in 73% of patient trials.
引用
收藏
页码:281 / 285
页数:5
相关论文
共 50 条
  • [41] An Extended Computational Model of the Circulatory System for Designing Ventricular Assist Devices
    Hsu, Po-Lin
    Cheng, Sean J.
    Saumarez, Richard C.
    Dawes, William N.
    McMahon, Richard A.
    ASAIO JOURNAL, 2008, 54 (06) : 594 - 599
  • [42] FEASIBILITY OF IMPLANTABLE PLUTONIUM 238 POWER SOURCES FOR CIRCULATORY ASSIST DEVICES
    NORMAN, JC
    HARVEY, RJ
    COVELLI, VH
    ROBINSON, TC
    BERNHARD, WF
    CIRCULATION, 1967, 36 (4S2) : I200 - &
  • [43] DYNAMIC MECHANICAL-PROPERTIES OF ELASTOMERS FOR USE IN CIRCULATORY ASSIST DEVICES
    MURAYAMA, T
    MCMILLIN, CR
    JOURNAL OF APPLIED POLYMER SCIENCE, 1983, 28 (06) : 1871 - 1877
  • [44] A computer model of the pediatric circulatory system for testing pediatric assist devices
    Giridharan, Guruprasad A.
    Koenig, Steven C.
    Mitchell, Michael
    Gartner, Mark
    Pantalos, George M.
    ASAIO JOURNAL, 2007, 53 (01) : 74 - 81
  • [45] TRANSMISSION OF WATER THROUGH A BIOCOMPATIBLE POLYURETHANE - APPLICATION TO CIRCULATORY ASSIST DEVICES
    REID, JS
    ROSENBERG, G
    PIERCE, WS
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1985, 19 (09): : 1181 - 1202
  • [46] ANESTHETIC EFFECT IN MECHANICAL CIRCULATORY ASSIST DEVICES. EXPERIMENTAL STUDY
    Gomez-Tapia, B.
    Morillas-Sendin, P.
    Barranco, M.
    Del Canizo, J. F.
    Quintana-Villamandos, B.
    BRITISH JOURNAL OF SURGERY, 2016, 103 : 1 - 1
  • [48] SKIN COLOR CHANGE WITH CIRCULATORY ASSIST DEVICES: SUPERIOR VENA CAVA SYNDROME
    Mizuno, Atsushi
    Asano, Taku
    Niwa, Koichiro
    JOURNAL OF EMERGENCY MEDICINE, 2014, 47 (06): : E153 - E155
  • [49] Temporary Circulatory Support With Ventricular Assist Devices: Update on Surgical and Percutaneous Strategies
    Enrique Gongora
    Erik Orozco
    Charles Hoopes
    Current Heart Failure Reports, 2020, 17 : 350 - 356
  • [50] Right Ventricular Dysfunction: Lessons Learned From Mechanical Circulatory Assist Devices
    Hess, Michael L.
    Cooke, Richard H.
    Varma, Amit
    Shah, Keyur B.
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2014, 347 (01): : 80 - 83